Device type · MVK
Wearable Automated External Defibrillator
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
MVK
Cardiovascular
Device class
Class III
FDA decisions since 2015
3
510(k), De Novo and PMA
Adverse event reports
12,477
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Wearable Automated External Defibrillator
- Product code
- MVK
- Device class
- Class III
- Review panel
- Cardiovascular
- Specialty
- Cardiology, Electrophysiology
- FDA record
- Classification database entry for MVK
Latest FDA decisions
P230022 · Decided Apr 30, 2025 · 667 days after receipt
Device type: Wearable automated external defibrillator (MVK) · Cardiology
Kestra Medical Technologies, Inc.
P200037 · Decided Jul 27, 2021 · 348 days after receipt
Device type: Wearable automated external defibrillator (MVK) · Cardiology
- LIFEVEST WEARABLE DEFIBRILLATORPMA supplement
Zoll Manufacturing Corporation
P010030-S056 · Decided Dec 17, 2015 · 1,031 days after receipt · Panel Track
Device type: Wearable automated external defibrillator (MVK) · Cardiology
Reports received
12,477
Received Nov 1, 2025 to Nov 19, 2026. Includes 2 of another or unstated type.
Malfunction reports
8,755
70% of reports
Injury reports
3,477
28% of reports
Death reports
243
1.9% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 632 | 241 | 28 | 0 |
| Jan 24 | 840 | 313 | 27 | 0 |
| Feb 24 | 889 | 289 | 25 | 0 |
| Mar 24 | 744 | 276 | 30 | 0 |
| Apr 24 | 681 | 296 | 22 | 0 |
| May 24 | 792 | 296 | 24 | 0 |
| Jun 24 | 820 | 268 | 30 | 0 |
| Jul 24 | 707 | 266 | 10 | 0 |
| Aug 24 | 800 | 236 | 19 | 0 |
| Sep 24 | 658 | 271 | 12 | 0 |
| Oct 24 | 698 | 277 | 28 | 0 |
| Nov 24 | 625 | 260 | 15 | 0 |
| Dec 24 | 732 | 292 | 17 | 0 |
| Jan 25 | 827 | 322 | 50 | 0 |
| Feb 25 | 630 | 305 | 25 | 0 |
| Mar 25 | 686 | 360 | 19 | 0 |
| Apr 25 | 722 | 340 | 19 | 0 |
| May 25 | 600 | 262 | 27 | 0 |
| Jun 25 | 630 | 288 | 19 | 0 |
| Jul 25 | 679 | 303 | 20 | 0 |
| Aug 25 | 670 | 273 | 13 | 0 |
| Sep 25 | 646 | 316 | 8 | 0 |
| Oct 25 | 838 | 276 | 23 | 0 |
| Nov 25 | 894 | 272 | 35 | 0 |
| Dec 25 | 593 | 286 | 16 | 0 |
| Jan 26 | 691 | 295 | 17 | 0 |
| Feb 26 | 709 | 274 | 10 | 0 |
| Mar 26 | 670 | 353 | 34 | 0 |
| Apr 26 | 765 | 308 | 11 | 0 |
| May 26 | 659 | 286 | 18 | 0 |
| Jun 26 | 757 | 299 | 19 | 0 |
| Jul 26 | 770 | 315 | 20 | 0 |
| Aug 26 | 750 | 247 | 27 | 1 |
| Sep 26 | 811 | 312 | 18 | 0 |
| Oct 26 | 686 | 230 | 18 | 1 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Zoll Manufacturing11,588
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Circuit Failure2,711
- Battery Problem2,027
- Patient Device Interaction Problem1,845
- Incorrect Interpretation of Signal1,157
- Charging Problem1,109
- Key or Button Unresponsive/not Working1,069
- Contamination /Decontamination Problem905
- Over-Sensing648
- Structural Problem500
- Electrical /Electronic Property Problem430
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions8,751
- Arrhythmia1,882
- Skin Inflammation/ Irritation1,839
- Insufficient Information28
- Asystole12
- Burn(s)7
- Abrasion6
- Fall6
- Bone Fracture(s)5
- Head Injury5
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-0353-2018 · Started Sep 12, 2017 · Posted by FDA Jan 14, 2018
Reason given: Incorrect service code for properly catching critical defects during self-check. Potential for defibrillation shock failure
Cause as FDA categorized it: Other
Z-0625-2011 · Started Sep 24, 2010 · Posted by FDA Dec 16, 2010
Reason given: The response button function may become intermittent. This can result in the device not progressing beyond the response button test during startup, which may prevent a conscious patient from delaying a treatment shock.
Cause as FDA categorized it: Other
Z-0883-2009 · Started Oct 21, 2008 · Posted by FDA Jan 28, 2009
Reason given: The charging circuit will not shut down properly.
Cause as FDA categorized it: Component design/selection
Z-0061-2008 · Started Sep 14, 2007 · Posted by FDA Oct 18, 2007
Reason given: Possible malfunction of the battery pack. A significant electrostatic discharge (ESD) can cause the battery pack to malfunction. The malfunction causes the battery pack to discharge more rapidly and prevents the pack from recharging.
Cause as FDA categorized it: Device Design
Z-0936-2007 · Started Apr 27, 2007 · Posted by FDA Jun 15, 2007
Reason given: The electrodes do not properly release conductive gel
Cause as FDA categorized it: Other
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.