MAUDE Monitor
Last checked: Nov 19, 2026Latest source record: Oct 30, 2026 (newest report received)openFDA last updated: Nov 11, 2026Methods & limitations
Filters
Filters are saved in the page address, so a filtered view can be bookmarked or shared.
With no dates set, the view covers the last 12 full months and the current month. Brand and manufacturer are matched as FDA recorded them, so try the short form of a name. Product code MVK: Wearable Automated External Defibrillator. Open this device type's page
Reports received
12,477
Received Nov 1, 2025 to Nov 19, 2026. Includes 2 of another or unstated type.
Malfunction reports
8,755
70% of reports
Injury reports
3,477
28% of reports
Death reports
243
1.9% of reports
Reports received per month
product code MVK. By the date FDA received each report, stacked by event type. The newest month is incomplete, and FDA adds reports for recent months as it processes them.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Nov 25 | 894 | 272 | 35 | 0 |
| Dec 25 | 593 | 286 | 16 | 0 |
| Jan 26 | 691 | 295 | 17 | 0 |
| Feb 26 | 709 | 274 | 10 | 0 |
| Mar 26 | 670 | 353 | 34 | 0 |
| Apr 26 | 765 | 308 | 11 | 0 |
| May 26 | 659 | 286 | 18 | 0 |
| Jun 26 | 757 | 299 | 19 | 0 |
| Jul 26 | 770 | 315 | 20 | 0 |
| Aug 26 | 750 | 247 | 27 | 1 |
| Sep 26 | 811 | 312 | 18 | 0 |
| Oct 26 | 686 | 230 | 18 | 1 |
| Nov 26 | 0 | 0 | 0 | 0 |
Device types with the most reports
By FDA product code. A report that names several devices counts once for each.
Manufacturers with the most reports
Bigger sellers get more reports. This is not a ranking of safety.
- Zoll Manufacturing11,588
Brand names with the most reports
As written on the reports. One product can appear under several spellings.
- LIFEVEST WCD 4000 SYSTEM10,459
- Jewel13
12,477 reports
Showing 1 to 25, newest first. Each title opens FDA's full record, which includes the reporter's description of the event.
- LifeVest WCD 5100 SystemInjury
ZOLL Manufacturing Corporation
3008642652-2026-10451 · Received Oct 30, 2026 · Event dated Oct 12, 2026
Device type: Wearable Automated External Defibrillator (MVK)
Device problem as coded: Patient Device Interaction Problem
Patient problem as coded: Skin Inflammation/ Irritation
- LifeVest WCD 4000 SystemMalfunction
ZOLL Manufacturing Corporation
3008642652-2026-10452 · Received Oct 30, 2026 · Event dated Oct 23, 2026
Device type: Wearable Automated External Defibrillator (MVK)
Device problem as coded: Circuit Failure; Contamination /Decontamination Problem
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- LifeVest WCD 4000 SystemMalfunction
ZOLL Manufacturing Corporation
3008642652-2026-10453 · Received Oct 30, 2026 · Event dated Oct 23, 2026
Device type: Wearable Automated External Defibrillator (MVK)
Device problem as coded: Circuit Failure
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- LifeVest WCD 5100 SystemMalfunction
ZOLL Manufacturing Corporation
3008642652-2026-10454 · Received Oct 30, 2026 · Event dated Oct 23, 2026
Device type: Wearable Automated External Defibrillator (MVK)
Device problem as coded: Circuit Failure
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- LifeVest WCD 4000 SystemInjury
ZOLL Manufacturing Corporation
3008642652-2026-10455 · Received Oct 30, 2026 · Event dated Oct 13, 2026
Device type: Wearable Automated External Defibrillator (MVK)
Device problem as coded: Patient Device Interaction Problem
Patient problem as coded: Skin Inflammation/ Irritation
- LifeVest WCD 4000 SystemMalfunction
ZOLL Manufacturing Corporation
3008642652-2026-10456 · Received Oct 30, 2026 · Event dated Oct 23, 2026
Device type: Wearable Automated External Defibrillator (MVK)
Device problem as coded: Circuit Failure
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- LifeVest WCD 5100 SystemMalfunction
ZOLL Manufacturing Corporation
3008642652-2026-10457 · Received Oct 30, 2026 · Event dated Oct 6, 2026
Device type: Wearable Automated External Defibrillator (MVK)
Device problem as coded: Key or Button Unresponsive/not Working
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- LifeVest WCD 4000 SystemMalfunction
ZOLL Manufacturing Corporation
3008642652-2026-10458 · Received Oct 30, 2026 · Event dated Oct 28, 2026
Device type: Wearable Automated External Defibrillator (MVK)
Device problem as coded: Circuit Failure
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- LifeVest WCD 4000 SystemMalfunction
ZOLL Manufacturing Corporation
3008642652-2026-10459 · Received Oct 30, 2026 · Event dated Oct 14, 2026
Device type: Wearable Automated External Defibrillator (MVK)
Device problem as coded: Circuit Failure
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- LifeVest WCD 4000 SystemMalfunction
ZOLL Manufacturing Corporation
3008642652-2026-10460 · Received Oct 30, 2026 · Event dated Oct 7, 2026
Device type: Wearable Automated External Defibrillator (MVK)
Device problem as coded: Circuit Failure
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- LifeVest WCD 5100 SystemInjury
ZOLL Manufacturing Corporation
3008642652-2026-10461 · Received Oct 30, 2026 · Event dated Oct 26, 2026
Device type: Wearable Automated External Defibrillator (MVK)
Device problem as coded: Patient Device Interaction Problem
Patient problem as coded: Skin Inflammation/ Irritation
- LifeVest WCD 4000 SystemInjury
ZOLL Manufacturing Corporation
3008642652-2026-10462 · Received Oct 30, 2026 · Event dated Oct 14, 2026
Device type: Wearable Automated External Defibrillator (MVK)
Device problem as coded: Patient Device Interaction Problem
Patient problem as coded: Skin Inflammation/ Irritation
- LifeVest WCD 5100 SystemInjury
ZOLL Manufacturing Corporation
3008642652-2026-10467 · Received Oct 30, 2026 · Event dated Oct 12, 2026
Device type: Wearable Automated External Defibrillator (MVK)
Device problem as coded: Patient Device Interaction Problem
Patient problem as coded: Skin Inflammation/ Irritation
- LifeVest WCD 5100 SystemInjury
ZOLL Manufacturing Corporation
3008642652-2026-10418 · Received Oct 29, 2026 · Event dated Oct 16, 2026
Device type: Wearable Automated External Defibrillator (MVK)
Device problem as coded: Patient Device Interaction Problem
Patient problem as coded: Skin Inflammation/ Irritation
- LifeVest WCD 4000 SystemMalfunction
ZOLL Manufacturing Corporation
3008642652-2026-10420 · Received Oct 29, 2026 · Event dated Oct 23, 2026
Device type: Wearable Automated External Defibrillator (MVK)
Device problem as coded: Key or Button Unresponsive/not Working
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- LifeVest WCD 4000 SystemMalfunction
ZOLL Manufacturing Corporation
3008642652-2026-10421 · Received Oct 29, 2026 · Event dated Oct 23, 2026
Device type: Wearable Automated External Defibrillator (MVK)
Device problem as coded: Key or Button Unresponsive/not Working
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- LifeVest WCD 5100 SystemInjury
ZOLL Manufacturing Corporation
3008642652-2026-10422 · Received Oct 29, 2026 · Event dated Oct 25, 2026
Device type: Wearable Automated External Defibrillator (MVK)
Device problem as coded: Adverse Event Without Identified Device or Use Problem
Patient problem as coded: Head Injury
- LifeVest WCD 4000 SystemMalfunction
ZOLL Manufacturing Corporation
3008642652-2026-10423 · Received Oct 29, 2026 · Event dated Oct 20, 2026
Device type: Wearable Automated External Defibrillator (MVK)
Device problem as coded: Key or Button Unresponsive/not Working
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- LifeVest WCD 4000 SystemMalfunction
ZOLL Manufacturing Corporation
3008642652-2026-10424 · Received Oct 29, 2026 · Event dated Oct 23, 2026
Device type: Wearable Automated External Defibrillator (MVK)
Device problem as coded: Key or Button Unresponsive/not Working
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- LifeVest WCD 5100 SystemMalfunction
ZOLL Manufacturing Corporation
3008642652-2026-10425 · Received Oct 29, 2026 · Event dated Sep 16, 2026
Device type: Wearable Automated External Defibrillator (MVK)
Device problem as coded: Charging Problem
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- LifeVest WCD 5100 SystemInjury
ZOLL Manufacturing Corporation
3008642652-2026-10426 · Received Oct 29, 2026 · Event dated Oct 13, 2026
Device type: Wearable Automated External Defibrillator (MVK)
Device problem as coded: Patient Device Interaction Problem
Patient problem as coded: Skin Inflammation/ Irritation
- LifeVest WCD 4000 SystemMalfunction
ZOLL Manufacturing Corporation
3008642652-2026-10427 · Received Oct 29, 2026 · Event dated Oct 13, 2026
Device type: Wearable Automated External Defibrillator (MVK)
Device problem as coded: Circuit Failure
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- LifeVest WCD 4000 SystemMalfunction
ZOLL Manufacturing Corporation
3008642652-2026-10428 · Received Oct 29, 2026 · Event dated Oct 20, 2026
Device type: Wearable Automated External Defibrillator (MVK)
Device problem as coded: Key or Button Unresponsive/not Working
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- LifeVest WCD 4000 SystemMalfunction
ZOLL Manufacturing Corporation
3008642652-2026-10429 · Received Oct 29, 2026 · Event dated Oct 12, 2026
Device type: Wearable Automated External Defibrillator (MVK)
Device problem as coded: Key or Button Unresponsive/not Working
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- LifeVest WCD 4000 SystemInjury
ZOLL Manufacturing Corporation
3008642652-2026-10431 · Received Oct 29, 2026 · Event dated Oct 23, 2026
Device type: Wearable Automated External Defibrillator (MVK)
Device problem as coded: Patient Device Interaction Problem
Patient problem as coded: Skin Inflammation/ Irritation
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.