Device type · HIF
Insufflator, Laparoscopic
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
HIF
Obstetrics & Gynecology
Device class
Class II
FDA decisions since 2015
38
510(k), De Novo and PMA
Adverse event reports
1,850
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Insufflator, Laparoscopic
- Product code
- HIF
- Device class
- Class II
- Review panel
- Obstetrics & Gynecology
- Regulation
- 21 CFR 884.1730
- Specialty
- Obstetrics & Gynecology
- FDA record
- Classification database entry for HIF
Latest FDA decisions
The 8 newest of 38 since 2015.
K260209 · Decided Aug 19, 2026 · 208 days after receipt
Device type: Insufflator, Laparoscopic (HIF) · Obstetrics & Gynecology
Weyo Surgical Technology , Ltd.
K252584 · Decided Jun 25, 2026 · 314 days after receipt
Device type: Insufflator, Laparoscopic (HIF) · Obstetrics & Gynecology
Guangdong Optomedic Technologies, Inc.
K252318 · Decided Apr 16, 2026 · 265 days after receipt
Device type: Insufflator, Laparoscopic (HIF) · Obstetrics & Gynecology
- ENDOFLATOR +510(k)
K250388 · Decided Sep 22, 2025 · 223 days after receipt
Device type: Insufflator, Laparoscopic (HIF) · Obstetrics & Gynecology
K242772 · Decided Apr 11, 2025 · 210 days after receipt
Device type: Insufflator, Laparoscopic (HIF) · Obstetrics & Gynecology
Olympus Medical Systems Corporation
K243527 · Decided Jan 13, 2025 · 60 days after receipt
Device type: Insufflator, Laparoscopic (HIF) · Obstetrics & Gynecology
K242536 · Decided Oct 15, 2024 · 50 days after receipt
Device type: Insufflator, Laparoscopic (HIF) · Obstetrics & Gynecology
K232464 · Decided May 2, 2024 · 261 days after receipt
Device type: Insufflator, Laparoscopic (HIF) · Obstetrics & Gynecology
Reports received
1,850
Received Nov 1, 2025 to Nov 19, 2026. Includes 2 of another or unstated type.
Malfunction reports
1,781
96% of reports
Injury reports
65
3.5% of reports
Death reports
2
0.1% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 88 | 11 | 2 | 0 |
| Jan 24 | 92 | 3 | 0 | 0 |
| Feb 24 | 57 | 11 | 0 | 0 |
| Mar 24 | 87 | 6 | 0 | 0 |
| Apr 24 | 68 | 4 | 0 | 0 |
| May 24 | 89 | 7 | 0 | 0 |
| Jun 24 | 109 | 6 | 2 | 0 |
| Jul 24 | 106 | 2 | 0 | 0 |
| Aug 24 | 126 | 8 | 2 | 0 |
| Sep 24 | 134 | 17 | 0 | 0 |
| Oct 24 | 174 | 107 | 0 | 0 |
| Nov 24 | 112 | 31 | 1 | 0 |
| Dec 24 | 96 | 25 | 1 | 0 |
| Jan 25 | 75 | 19 | 0 | 0 |
| Feb 25 | 119 | 7 | 0 | 0 |
| Mar 25 | 75 | 7 | 1 | 0 |
| Apr 25 | 78 | 11 | 1 | 0 |
| May 25 | 49 | 8 | 0 | 0 |
| Jun 25 | 55 | 7 | 0 | 0 |
| Jul 25 | 80 | 11 | 0 | 0 |
| Aug 25 | 91 | 17 | 0 | 0 |
| Sep 25 | 82 | 4 | 0 | 0 |
| Oct 25 | 418 | 4 | 0 | 0 |
| Nov 25 | 179 | 3 | 0 | 0 |
| Dec 25 | 111 | 5 | 2 | 0 |
| Jan 26 | 179 | 6 | 0 | 0 |
| Feb 26 | 135 | 6 | 0 | 0 |
| Mar 26 | 336 | 2 | 0 | 0 |
| Apr 26 | 172 | 3 | 0 | 0 |
| May 26 | 95 | 10 | 0 | 0 |
| Jun 26 | 118 | 4 | 0 | 0 |
| Jul 26 | 122 | 6 | 0 | 0 |
| Aug 26 | 95 | 7 | 0 | 0 |
| Sep 26 | 128 | 5 | 0 | 2 |
| Oct 26 | 111 | 8 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Shirakawa Olympus1,699
- CONMED67
- Stryker18
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Pressure Problem301
- Gas/Air Leak199
- Unexpected Shutdown104
- No Display/Image103
- Circuit Failure99
- Display or Visual Feedback Problem99
- Break83
- Prophylactic removal due to corrective action80
- Gas Output Problem69
- Increase in Pressure67
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions1,781
- Insufficient Information19
- Air Embolism18
- Cardiac Arrest7
- Abdominal Distention4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available4
- Pneumothorax4
- Stroke/CVA3
- Bowel Perforation2
- Bradycardia2
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
- Olympus High Flow Insufflation Unit - UHI-4. Model Number: UHI-4Open, Classified
Olympus Corporation of the Americas
Z-2776-2026 · Started Jun 16, 2026 · Posted by FDA Jul 22, 2026
Reason given: Potential for unexpected instances where the front panel display turns off unexpectedly, and air delivery (insufflation) ceases.
Cause as FDA categorized it: Nonconforming Material/Component
Z-2464-2026 · Started May 14, 2026 · Posted by FDA Jun 17, 2026
Reason given: Potential for overpressure alerts.
Cause as FDA categorized it: Process change control
Z-2465-2026 · Started May 14, 2026 · Posted by FDA Jun 17, 2026
Reason given: Potential for overpressure alerts.
Cause as FDA categorized it: Process change control
Z-2466-2026 · Started May 14, 2026 · Posted by FDA Jun 17, 2026
Reason given: Potential for overpressure alerts.
Cause as FDA categorized it: Process change control
Z-2467-2026 · Started May 14, 2026 · Posted by FDA Jun 17, 2026
Reason given: Potential for overpressure alerts.
Cause as FDA categorized it: Process change control
Z-2468-2026 · Started May 14, 2026 · Posted by FDA Jun 17, 2026
Reason given: Potential for overpressure alerts.
Cause as FDA categorized it: Process change control
Z-2469-2026 · Started May 14, 2026 · Posted by FDA Jun 17, 2026
Reason given: Potential for overpressure alerts.
Cause as FDA categorized it: Process change control
Z-2470-2026 · Started May 14, 2026 · Posted by FDA Jun 17, 2026
Reason given: Potential for overpressure alerts.
Cause as FDA categorized it: Process change control
Showing the 8 newest of 29 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.