With no dates set, the view covers the last 12 full months and the current month. Brand and manufacturer are matched as FDA recorded them, so try the short form of a name. Product code HIF: Insufflator, Laparoscopic. Open this device type's page
Reports received
1,850
Received Nov 1, 2025 to Nov 19, 2026. Includes 2 of another or unstated type.
Malfunction reports
1,781
96% of reports
Injury reports
65
3.5% of reports
Death reports
2
0.1% of reports
Reports received per month
product code HIF. By the date FDA received each report, stacked by event type. The newest month is incomplete, and FDA adds reports for recent months as it processes them.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Nov 25 | 179 | 3 | 0 | 0 |
| Dec 25 | 111 | 5 | 2 | 0 |
| Jan 26 | 179 | 6 | 0 | 0 |
| Feb 26 | 135 | 6 | 0 | 0 |
| Mar 26 | 336 | 2 | 0 | 0 |
| Apr 26 | 172 | 3 | 0 | 0 |
| May 26 | 95 | 10 | 0 | 0 |
| Jun 26 | 118 | 4 | 0 | 0 |
| Jul 26 | 122 | 6 | 0 | 0 |
| Aug 26 | 95 | 7 | 0 | 0 |
| Sep 26 | 128 | 5 | 0 | 2 |
| Oct 26 | 111 | 8 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Device types with the most reports
By FDA product code. A report that names several devices counts once for each.
Manufacturers with the most reports
Bigger sellers get more reports. This is not a ranking of safety.
- Shirakawa Olympus1,699
- CONMED67
- Stryker18
Brand names with the most reports
As written on the reports. One product can appear under several spellings.
- AIRSEAL68
1,850 reports
Showing 1 to 25, newest first. Each title opens FDA's full record, which includes the reporter's description of the event.
- CYLINDER HOSEMalfunction
3002808148-2026-37501 · Received Oct 29, 2026 · Event dated Jun 29, 2026
Device type: Insufflator, Laparoscopic (HIF)
Device problem as coded: Break
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- HIGH FLOW INSUFFLATION UNITMalfunction
3002808148-2026-37279 · Received Oct 28, 2026
Device type: Insufflator, Laparoscopic (HIF)
Device problem as coded: Gas/Air Leak
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- HIGH FLOW INSUFFLATION UNITMalfunction
3002808148-2026-37358 · Received Oct 28, 2026
Device type: Insufflator, Laparoscopic (HIF)
Device problem as coded: Increase in Pressure
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- ENDOFLATOR +Malfunction
9610617-2026-02562 · Received Oct 28, 2026
Device type: Insufflator, Laparoscopic (HIF)
Device problem as coded: No Display/Image
Patient problem as coded: Insufficient Information
- HIGH FLOW INSUFFLATION UNITMalfunction
3002808148-2026-37206 · Received Oct 27, 2026
Device type: Insufflator, Laparoscopic (HIF)
Device problem as coded: Decrease in Pressure
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- HIGH FLOW INSUFFLATION UNITMalfunction
3002808148-2026-37240 · Received Oct 27, 2026 · Event dated Oct 26, 2026
Device type: Insufflator, Laparoscopic (HIF)
Device problem as coded: Increase in Pressure; Improper Flow or Infusion
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- PKG, PNEUMO SURE XL HG FLW INSUFFLATORMalfunction
0002936485-2026-01441 · Received Oct 27, 2026 · Event dated Sep 9, 2026
Device type: Insufflator, Laparoscopic (HIF)
Device problem as coded: Decrease in Pressure
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- HIGH FLOW INSUFFLATION UNITMalfunction
3002808148-2026-37001 · Received Oct 26, 2026 · Event dated Sep 28, 2026
Device type: Insufflator, Laparoscopic (HIF)
Device problem as coded: Device Alarm System; Display or Visual Feedback Problem; Key or Button Unresponsive/not Working
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- HIGH FLOW INSUFFLATION UNITMalfunction
3002808148-2026-37002 · Received Oct 26, 2026 · Event dated Oct 13, 2026
Device type: Insufflator, Laparoscopic (HIF)
Device problem as coded: Improper Flow or Infusion
Patient problem as coded: Insufficient Information
- HIGH FLOW INSUFFLATION UNITMalfunction
3002808148-2026-37022 · Received Oct 26, 2026
Device type: Insufflator, Laparoscopic (HIF)
Device problem as coded: Gas/Air Leak
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- HIGH FLOW INSUFFLATION UNITMalfunction
3002808148-2026-37026 · Received Oct 26, 2026 · Event dated Oct 6, 2026
Device type: Insufflator, Laparoscopic (HIF)
Device problem as coded: Display or Visual Feedback Problem
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- HIGH FLOW INSUFFLATION UNITMalfunction
3002808148-2026-37033 · Received Oct 26, 2026 · Event dated Jun 12, 2025
Device type: Insufflator, Laparoscopic (HIF)
Device problem as coded: Corroded
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- CYLINDER HOSEMalfunction
3002808148-2026-37095 · Received Oct 26, 2026 · Event dated Oct 9, 2026
Device type: Insufflator, Laparoscopic (HIF)
Device problem as coded: Damaged Thread
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- PKG, PNEUMO SURE XL HG FLW INSUFFLATORMalfunction
0002936485-2026-01420 · Received Oct 26, 2026 · Event dated Sep 28, 2026
Device type: Insufflator, Laparoscopic (HIF)
Device problem as coded: Device Displays Incorrect Message
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- ENDOFLATOR +Malfunction
9610617-2026-02502 · Received Oct 24, 2026 · Event dated Sep 1, 2026
Device type: Insufflator, Laparoscopic (HIF)
Device problem as coded: Unexpected Shutdown
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- ENDOFLATOR +Malfunction
9610617-2026-02503 · Received Oct 24, 2026 · Event dated Sep 21, 2026
Device type: Insufflator, Laparoscopic (HIF)
Device problem as coded: Unexpected Shutdown
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- HIGH FLOW INSUFFLATION UNITMalfunction
3002808148-2026-36851 · Received Oct 23, 2026 · Event dated Sep 15, 2026
Device type: Insufflator, Laparoscopic (HIF)
Device problem as coded: Intermittent Loss of Power
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- HIGH FLOW INSUFFLATION UNITMalfunction
3002808148-2026-36868 · Received Oct 23, 2026 · Event dated Oct 22, 2026
Device type: Insufflator, Laparoscopic (HIF)
Device problem as coded: No Display/Image
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- HIGH FLOW INSUFFLATION UNITMalfunction
3002808148-2026-36867 · Received Oct 23, 2026
Device type: Insufflator, Laparoscopic (HIF)
Device problem as coded: No Display/Image
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- HIGH FLOW INSUFFLATION UNITMalfunction
3002808148-2026-36869 · Received Oct 23, 2026
Device type: Insufflator, Laparoscopic (HIF)
Device problem as coded: Erratic or Intermittent Display; Activation Failure
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- HIGH FLOW INSUFFLATION UNITMalfunction
3002808148-2026-36894 · Received Oct 23, 2026
Device type: Insufflator, Laparoscopic (HIF)
Device problem as coded: No Flow
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- HIGH FLOW INSUFFLATION UNITMalfunction
3002808148-2026-36703 · Received Oct 22, 2026
Device type: Insufflator, Laparoscopic (HIF)
Device problem as coded: Communication or Transmission Problem
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- HIGH FLOW INSUFFLATION UNITMalfunction
3002808148-2026-36704 · Received Oct 22, 2026 · Event dated Jul 2, 2025
Device type: Insufflator, Laparoscopic (HIF)
Device problem as coded: Key or Button Unresponsive/not Working
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- HIGH FLOW INSUFFLATION UNITMalfunction
3002808148-2026-36701 · Received Oct 22, 2026
Device type: Insufflator, Laparoscopic (HIF)
Device problem as coded: Low Audible Alarm; Decrease in Pressure; Unexpected Shutdown
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- HIGH FLOW INSUFFLATION UNITMalfunction
3002808148-2026-36768 · Received Oct 22, 2026
Device type: Insufflator, Laparoscopic (HIF)
Device problem as coded: Increase in Pressure
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.