FDA Clearance Tracker · P060037-S005
NEXGEN COMPLETE KNEE SOLUTION
Last checked: Nov 20, 2026Latest source record: Mar 26, 2009 (this decision)Source: openFDAMethods & limitations
Decided
Mar 26, 2009
Approved
Time to decision
178 days
Received Sep 29, 2008. Calendar days, including time FDA spent waiting on the applicant.
Pathway
PMA supplement
Normal 180 Day Track
Device class
Class III
The decision
- FDA number
- P060037-S005
- Device name
- NEXGEN COMPLETE KNEE SOLUTION
- Applicant
- Zimmer, Inc.
- Applicant location
- Warsaw, IN
- Received
- Sep 29, 2008
- Decided
- Mar 26, 2009
- Decision
- Approved
- Supplement type
- Normal 180 Day Track
- Reason for supplement
- Change Design/Components/Specifications/Material
- Original approval
- P060037
Device type
- Device type
- Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing (NJL)
- Device class
- Class III
FDA's approval statement
As written in FDA's record.
APPROVAL FOR CHANGES TO THE NEXGEN FLUTED STEM MOBILE TIBIAL COMPONENT,RESULTING IN THE NEXGEN MINIMALLY INVASIVE SOLUTIONS (MIS) LPS-MOBILE TIBIAL COMPONENT.CHANGES MADE TO THE NEXGEN FLUTED STEM MOBILE TIBIAL COMPONENT INCLUDE SHORTENING THETIBIAL BASEPLATE STEM AND ADDING BROAD PROXIMAL FINS, CHANGING THE OUTER PROFILE RADIUS OF THEBASEPLATE EDGE FROM A VARIABLE RADIUS TO A CONSTANT RADIUS, AND ADDING AN ADDITIONAL SIZE (4-PLUS)TO THE CURRENT RANGE OF SIZES (1 - 8).
Source documents
Other decisions for Zimmer, Inc.
Matched on the applicant name as FDA recorded it. A company that files under several names will have more.
K262955 · Decided Sep 24, 2026 · 31 days after receipt
Device type: Prosthesis, Total Anatomic Shoulder, Uncemented Metaphyseal Humeral Stem With No Diaphyseal Incursion, Semi-Constrained (PKC) · Orthopedic Surgery
- Z1 Hip System510(k)
K260831 · Decided Jun 16, 2026 · 95 days after receipt
Device type: Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented (LZO) · Orthopedic Surgery
K261683 · Decided Jun 16, 2026 · 26 days after receipt
Device type: Shoulder Prosthesis, Reverse Configuration (PHX) · Orthopedic Surgery
5 more FDA decisions with a free account
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.