Device type · PHX
Shoulder Prosthesis, Reverse Configuration
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
PHX
Orthopedic
Device class
Class II
FDA decisions since 2015
131
510(k), De Novo and PMA
Adverse event reports
2,954
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Shoulder Prosthesis, Reverse Configuration
- Product code
- PHX
- Device class
- Class II
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3660
- Specialty
- Orthopedic Surgery, Joint Reconstruction
- FDA record
- Classification database entry for PHX
Latest FDA decisions
The 8 newest of 131 since 2015.
K260417 · Decided Nov 5, 2026 · 269 days after receipt
Device type: Shoulder Prosthesis, Reverse Configuration (PHX) · Orthopedic Surgery
- Veritas Shoulder System510(k)
K263421 · Decided Oct 22, 2026 · 27 days after receipt
Device type: Shoulder Prosthesis, Reverse Configuration (PHX) · Orthopedic Surgery
K263086 · Decided Oct 2, 2026 · 30 days after receipt
Device type: Shoulder Prosthesis, Reverse Configuration (PHX) · Orthopedic Surgery
- Veritas Shoulder System510(k)
K260556 · Decided Jul 1, 2026 · 132 days after receipt
Device type: Shoulder Prosthesis, Reverse Configuration (PHX) · Orthopedic Surgery
K261683 · Decided Jun 16, 2026 · 26 days after receipt
Device type: Shoulder Prosthesis, Reverse Configuration (PHX) · Orthopedic Surgery
- Tornier Perform Reversed Glenoid and Tornier Perform Reversed Augmented Glenoid Shoulder System510(k)
Stryker Corporation (Tornier, Inc.)
K260102 · Decided Jun 12, 2026 · 150 days after receipt
Device type: Shoulder Prosthesis, Reverse Configuration (PHX) · Orthopedic Surgery
K261016 · Decided Jun 10, 2026 · 75 days after receipt
Device type: Shoulder Prosthesis, Reverse Configuration (PHX) · Orthopedic Surgery
K253992 · Decided May 7, 2026 · 146 days after receipt
Device type: Shoulder Prosthesis, Reverse Configuration (PHX) · Orthopedic Surgery
Reports received
2,954
Received Nov 1, 2025 to Nov 19, 2026. Includes 2 of another or unstated type.
Malfunction reports
341
12% of reports
Injury reports
2,610
88% of reports
Death reports
1
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 11 | 137 | 0 | 0 |
| Jan 24 | 19 | 111 | 0 | 0 |
| Feb 24 | 24 | 133 | 0 | 0 |
| Mar 24 | 49 | 172 | 0 | 0 |
| Apr 24 | 45 | 217 | 0 | 0 |
| May 24 | 32 | 134 | 0 | 0 |
| Jun 24 | 24 | 161 | 0 | 0 |
| Jul 24 | 16 | 133 | 0 | 0 |
| Aug 24 | 12 | 154 | 0 | 0 |
| Sep 24 | 21 | 155 | 0 | 0 |
| Oct 24 | 36 | 177 | 0 | 0 |
| Nov 24 | 18 | 138 | 0 | 0 |
| Dec 24 | 27 | 135 | 0 | 0 |
| Jan 25 | 29 | 127 | 0 | 0 |
| Feb 25 | 27 | 142 | 0 | 0 |
| Mar 25 | 33 | 177 | 0 | 0 |
| Apr 25 | 45 | 208 | 0 | 0 |
| May 25 | 32 | 210 | 0 | 0 |
| Jun 25 | 16 | 278 | 0 | 0 |
| Jul 25 | 33 | 272 | 9 | 0 |
| Aug 25 | 27 | 195 | 0 | 0 |
| Sep 25 | 34 | 204 | 0 | 0 |
| Oct 25 | 33 | 240 | 0 | 0 |
| Nov 25 | 39 | 200 | 0 | 0 |
| Dec 25 | 38 | 240 | 0 | 0 |
| Jan 26 | 32 | 253 | 0 | 0 |
| Feb 26 | 24 | 208 | 0 | 0 |
| Mar 26 | 29 | 249 | 0 | 0 |
| Apr 26 | 28 | 283 | 0 | 0 |
| May 26 | 15 | 225 | 0 | 0 |
| Jun 26 | 28 | 193 | 0 | 0 |
| Jul 26 | 23 | 171 | 0 | 0 |
| Aug 26 | 22 | 203 | 0 | 2 |
| Sep 26 | 40 | 213 | 0 | 0 |
| Oct 26 | 23 | 172 | 1 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Adverse Event Without Identified Device or Use Problem1,296
- Device Dislodged or Dislocated544
- Fracture274
- Unstable225
- Patient Device Interaction Problem149
- Loosening of Implant Not Related to Bone-Ingrowth118
- Detachment of Device or Device Component95
- Osseointegration Problem47
- Loss of Osseointegration45
- Naturally Worn41
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Joint Dislocation573
- Unspecified Infection488
- Joint Laxity432
- No Clinical Signs, Symptoms or Conditions340
- Insufficient Information302
- Pain270
- Bone Fracture(s)137
- Aseptic loosening110
- Medical device site infection101
- Osteolysis92
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
- Veritas Reverse Total Shoulder Replacement, Model/Catalog Number: 3500-140C3, 3500-140U3, 3500-140C1, 3500-140U1, size 1 -2, 40mm Polymer LinerOpen, Classified
Z-0348-2027 · Started Oct 1, 2026 · Posted by FDA Nov 3, 2026
Reason given: (Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components lack adequate retention strength, potentially resulting in dislocation, reduced range of motion, or...
Cause as FDA categorized it: Device Design
Z-2987-2026 · Started Jun 30, 2026 · Posted by FDA Aug 19, 2026
Reason given: Due to incorrect machining of peripheral screw hole resulting in potential for improper engagement of the peripheral screws.
Cause as FDA categorized it: Device Design
- Off-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 110040240Open, Classified
Z-2527-2026 · Started May 20, 2026 · Posted by FDA Jun 23, 2026
Reason given: Five complaints have been received identifying issues where the plastic guides were catching during reaming, leading to breakage.
Cause as FDA categorized it: Under Investigation by firm
- Off-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number: 110040241Open, Classified
Z-2528-2026 · Started May 20, 2026 · Posted by FDA Jun 23, 2026
Reason given: Five complaints have been received identifying issues where the plastic guides were catching during reaming, leading to breakage.
Cause as FDA categorized it: Under Investigation by firm
- Off-Axis Comprehensive Shoulder System, Large, Augment, Off-Axis, Reamer Guide Model/Catalog Number: 110040242Open, Classified
Z-2529-2026 · Started May 20, 2026 · Posted by FDA Jun 23, 2026
Reason given: Five complaints have been received identifying issues where the plastic guides were catching during reaming, leading to breakage.
Cause as FDA categorized it: Under Investigation by firm
- Off-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444Open, Classified
Z-2530-2026 · Started May 20, 2026 · Posted by FDA Jun 23, 2026
Reason given: Five complaints have been received identifying issues where the plastic guides were catching during reaming, leading to breakage.
Cause as FDA categorized it: Under Investigation by firm
- Off-Axis Alliance Glenoid, Right, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040476Open, Classified
Z-2531-2026 · Started May 20, 2026 · Posted by FDA Jun 23, 2026
Reason given: Five complaints have been received identifying issues where the plastic guides were catching during reaming, leading to breakage.
Cause as FDA categorized it: Under Investigation by firm
Z-2618-2026 · Started Apr 30, 2026 · Posted by FDA Jun 29, 2026
Reason given: Surgical instruments used during shoulder arthroplasty procedures does not progressively translate to replicate implant geometry, so the reaming guide and the stem reamer are positioned approximately 0.2 mm per size more laterally than intended, which may increase the risk of fracture in the medial region of the...
Cause as FDA categorized it: Device Design
Showing the 8 newest of 69 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.