Device type · LZO
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
LZO
Orthopedic
Device class
Class II
FDA decisions since 2015
169
510(k), De Novo and PMA
Adverse event reports
2,768
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented
- Product code
- LZO
- Device class
- Class II
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3353
- Specialty
- Orthopedic Surgery, Joint Reconstruction
- FDA record
- Classification database entry for LZO
Latest FDA decisions
The 8 newest of 169 since 2015.
- NEXUS® Hip Stem510(k)
K262473 · Decided Oct 1, 2026 · 76 days after receipt
Device type: Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented (LZO) · Orthopedic Surgery
K260173 · Decided Sep 17, 2026 · 239 days after receipt
Device type: Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented (LZO) · Orthopedic Surgery
K253948 · Decided Sep 2, 2026 · 266 days after receipt
Device type: Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented (LZO) · Orthopedic Surgery
K261009 · Decided Jul 6, 2026 · 102 days after receipt
Device type: Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented (LZO) · Orthopedic Surgery
- Z1 Hip System510(k)
K260831 · Decided Jun 16, 2026 · 95 days after receipt
Device type: Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented (LZO) · Orthopedic Surgery
K260182 · Decided Apr 27, 2026 · 96 days after receipt
Device type: Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented (LZO) · Orthopedic Surgery
K260037 · Decided Mar 30, 2026 · 83 days after receipt
Device type: Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented (LZO) · Orthopedic Surgery
- implaFit® short stems510(k)
K252401 · Decided Mar 12, 2026 · 224 days after receipt
Device type: Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented (LZO) · Orthopedic Surgery
Reports received
2,768
Received Nov 1, 2025 to Nov 19, 2026. Includes 3 of another or unstated type.
Malfunction reports
322
12% of reports
Injury reports
2,439
88% of reports
Death reports
4
0.1% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 24 | 171 | 0 | 0 |
| Jan 24 | 39 | 172 | 0 | 0 |
| Feb 24 | 15 | 156 | 0 | 0 |
| Mar 24 | 12 | 196 | 0 | 0 |
| Apr 24 | 20 | 191 | 1 | 0 |
| May 24 | 43 | 248 | 0 | 0 |
| Jun 24 | 25 | 217 | 3 | 0 |
| Jul 24 | 32 | 199 | 0 | 0 |
| Aug 24 | 25 | 192 | 0 | 0 |
| Sep 24 | 22 | 154 | 0 | 0 |
| Oct 24 | 16 | 181 | 0 | 0 |
| Nov 24 | 24 | 162 | 0 | 0 |
| Dec 24 | 33 | 208 | 0 | 0 |
| Jan 25 | 29 | 185 | 0 | 0 |
| Feb 25 | 29 | 144 | 0 | 0 |
| Mar 25 | 50 | 216 | 0 | 0 |
| Apr 25 | 68 | 319 | 0 | 0 |
| May 25 | 44 | 223 | 2 | 0 |
| Jun 25 | 56 | 247 | 0 | 0 |
| Jul 25 | 57 | 307 | 0 | 0 |
| Aug 25 | 27 | 229 | 0 | 0 |
| Sep 25 | 33 | 227 | 1 | 0 |
| Oct 25 | 39 | 256 | 0 | 0 |
| Nov 25 | 19 | 197 | 0 | 0 |
| Dec 25 | 29 | 256 | 0 | 0 |
| Jan 26 | 21 | 178 | 1 | 0 |
| Feb 26 | 33 | 140 | 0 | 0 |
| Mar 26 | 21 | 239 | 1 | 0 |
| Apr 26 | 22 | 228 | 1 | 0 |
| May 26 | 39 | 182 | 0 | 0 |
| Jun 26 | 17 | 179 | 0 | 0 |
| Jul 26 | 26 | 269 | 1 | 0 |
| Aug 26 | 38 | 172 | 0 | 0 |
| Sep 26 | 32 | 213 | 0 | 0 |
| Oct 26 | 25 | 186 | 0 | 3 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Adverse Event Without Identified Device or Use Problem1,206
- Device Dislodged or Dislocated456
- Fracture218
- Insufficient Device Problem Information127
- Osseointegration Problem116
- Tear, Rip or Hole in Device Packaging113
- Device Appears to Trigger Rejection98
- Loosening of Implant Not Related to Bone-Ingrowth80
- Material Erosion55
- Delivered as Unsterile Product47
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Unspecified Infection514
- Joint Dislocation461
- Insufficient Information351
- No Clinical Signs, Symptoms or Conditions325
- Pain314
- Joint Laxity175
- Medical device site infection154
- Aseptic loosening143
- Bone Fracture(s)109
- Inadequate Osseointegration96
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-0151-2026 · Started Sep 9, 2025 · Posted by FDA Oct 14, 2025
Reason given: Potential packaging issue that could compromise the sterile barrier. Defect may result in loss of sterility of the device, presenting potential risk of infection if device is implanted.
Cause as FDA categorized it: Package design/selection
Z-1645-2025 · Started Mar 24, 2025 · Posted by FDA Apr 21, 2025
Reason given: The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
Cause as FDA categorized it: Under Investigation by firm
Z-1646-2025 · Started Mar 24, 2025 · Posted by FDA Apr 21, 2025
Reason given: The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
Cause as FDA categorized it: Under Investigation by firm
Z-1647-2025 · Started Mar 24, 2025 · Posted by FDA Apr 21, 2025
Reason given: The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
Cause as FDA categorized it: Under Investigation by firm
Z-1648-2025 · Started Mar 24, 2025 · Posted by FDA Apr 21, 2025
Reason given: The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
Cause as FDA categorized it: Under Investigation by firm
Z-1649-2025 · Started Mar 24, 2025 · Posted by FDA Apr 21, 2025
Reason given: The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
Cause as FDA categorized it: Under Investigation by firm
- Z1 Femoral Hip System, Product Number 611777612Open, Classified
Z-1329-2025 · Started Feb 12, 2025 · Posted by FDA Mar 24, 2025
Reason given: There is a potential design issue with the Z1 Offset Stem Inserter leading to fracturing of the distal tip.
Cause as FDA categorized it: Under Investigation by firm
Waldemar Link GmbH & Co. KG (Mfg Site)
Z-0769-2025 · Started Dec 5, 2024 · Posted by FDA Dec 26, 2024
Reason given: Error on the main label and the patient label. The labels indicate a size XXL, this is a standard-size neck segment.. All further information on the labels is correct (Length, CCD and Offset).
Cause as FDA categorized it: Under Investigation by firm
Showing the 8 newest of 177 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.