All In MedTech

FDA Clearance Tracker · P010013-S058

NOVASURE IMPEDANCE CONTROLLED EDOMETRIAL ABLATION SYSTEM

A PMA supplement is a change to a device that already has premarket approval, such as a new indication, a design change or a manufacturing change.

Last checked: Nov 20, 2026Latest source record: Oct 28, 2014 (this decision)Source: openFDAMethods & limitations

PMA supplement

The decision

FDA number
P010013-S058
Device name
NOVASURE IMPEDANCE CONTROLLED EDOMETRIAL ABLATION SYSTEM
Applicant
Hologic, Inc.
Applicant location
Malborough, MA
Received
Jul 27, 2014
Decided
Oct 28, 2014
Decision
Approved
Supplement type
Normal 180 Day Track
Reason for supplement
Change Design/Components/Specifications/Material
Original approval
P010013

Device type

Device class
Class III

Every decision under product code MNB

FDA's approval statement

As written in FDA's record.

APPROVAL FOR DEVICE MODIFICATIONS, INCLUDING FLATTENING OF THE NVSRAM BATTERY CLIP AND ADDITION OF KAPTON TAPE TO THE NVSRAM BATTERY AND THE UI BOARD SHIELD TO OVERCOME A DEVICE ERROR OCCURRING IF CONTACT OF THESE COMPONENTS OCCURS.

Source documents

Other decisions for Hologic, Inc.

Matched on the applicant name as FDA recorded it. A company that files under several names will have more.

Company page

7 more FDA decisions with a free account

For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.