All In MedTech

Device type · MUE

Full Field Digital, System, X-Ray, Mammographic

FDA groups similar devices under a three-letter product code. This page gathers what FDA's public data says about MUE: how it is classified, what has been cleared or approved, and what has been reported and recalled.

Last checked: Nov 19, 2026Source: openFDAMethods & limitations

Product code

MUE

Radiology

Device class

Class II

FDA decisions since 2015

34

510(k), De Novo and PMA

Adverse event reports

66

Received Nov 1, 2025 to Nov 19, 2026

How FDA classifies it

Device type
Full Field Digital, System, X-Ray, Mammographic
Product code
MUE
Device class
Class II
Review panel
Radiology
Regulation
21 CFR 892.1715

Latest FDA decisions

The 8 newest of 34 since 2015.

All decisions for MUE

Reports received

66

Received Nov 1, 2025 to Nov 19, 2026. Includes 2 of another or unstated type.

Malfunction reports

61

92% of reports

Injury reports

3

4.5% of reports

Death reports

0

0.0% of reports

Adverse event reports received per month

By the date FDA received each report, stacked by event type. The newest month is incomplete.

Open in the MAUDE Monitor
  • Malfunction
  • Injury
  • Death
  • Other
View as table
MonthMalfunctionInjuryDeathOther
Dec 230000
Jan 240000
Feb 242000
Mar 242000
Apr 245000
May 242100
Jun 243100
Jul 241000
Aug 241000
Sep 245000
Oct 244000
Nov 242000
Dec 243100
Jan 252000
Feb 254000
Mar 253010
Apr 255000
May 254030
Jun 254000
Jul 250100
Aug 250001
Sep 251000
Oct 254001
Nov 250000
Dec 256000
Jan 261000
Feb 266001
Mar 2612100
Apr 263100
May 263001
Jun 265000
Jul 267000
Aug 263000
Sep 267100
Oct 268000
Nov 260000

Manufacturers named most often

Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.

Device problems as coded

Last 12 months. FDA's terms. A report can carry several.

  • Unintended System Motion22
  • Unintended Movement13
  • Break3
  • Installation-Related Problem3
  • Mechanics Altered3
  • Defective Device2
  • Device Handling Problem2
  • Device Operational Issue2
  • Mechanical Jam2
  • Mechanical Problem2

Patient problems as coded

Last 12 months. FDA's terms. A report can carry several.

  • No Clinical Signs, Symptoms or Conditions56
  • Insufficient Information2
  • Laceration(s)2
  • Bruise/Contusion1
  • Crushing Injury1
  • Deformity/ Disfigurement1
  • Eye Injury1
  • Foreign Body In Patient1
  • Head Injury1
  • Itching Sensation1

Recent recalls

From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.

  • Senographe PristinaOpen, Classified

    GE Medical Systems, LLC

    Z-0326-2027 · Started Aug 28, 2026 · Posted by FDA Nov 4, 2026

    Reason given: GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the...

    Cause as FDA categorized it: Radiation Control for Health and Safety Act

  • Senographe PristinaOpen, Classified

    GE Medical Systems, LLC

    Z-0327-2027 · Started Aug 28, 2026 · Posted by FDA Nov 4, 2026

    Reason given: GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the...

    Cause as FDA categorized it: Radiation Control for Health and Safety Act

  • Hologic, Inc

    Z-2167-2026 · Started Apr 8, 2026 · Posted by FDA May 12, 2026

    Reason given: Complaints have been received of systems developing loose, missing, or broken internal bolts over time.

    Cause as FDA categorized it: Device Design

  • GE Medical Systems, LLC

    Z-0875-2026 · Started Oct 24, 2025 · Posted by FDA Dec 9, 2025

    Reason given: Some Senographe Pristina systems either have a missing or an incorrect X-Ray Warning label, which is a nonconformance to FDA Regulation 21 CFR 1020.30(j).

    Cause as FDA categorized it: Radiation Control for Health and Safety Act

  • MAMMOMAT Revelation;Open, Classified

    Siemens Medical Solutions USA, Inc

    Z-0877-2026 · Started Oct 24, 2025 · Posted by FDA Dec 3, 2025

    Reason given: There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.

    Cause as FDA categorized it: Component design/selection

  • MAMMOMAT Inspiration;Open, Classified

    Siemens Medical Solutions USA, Inc

    Z-0878-2026 · Started Oct 24, 2025 · Posted by FDA Dec 3, 2025

    Reason given: There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.

    Cause as FDA categorized it: Component design/selection

  • MAMMOMAT Fusion;Open, Classified

    Siemens Medical Solutions USA, Inc

    Z-0879-2026 · Started Oct 24, 2025 · Posted by FDA Dec 3, 2025

    Reason given: There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.

    Cause as FDA categorized it: Component design/selection

  • Senographe PristinaOpen, Classified

    GE Medical Systems, SCS

    Z-0862-2024 · Started Dec 26, 2023 · Posted by FDA Feb 8, 2024

    Reason given: X-ray exposure termination audible signal on Senographe Pristina with Efficiency Suite 2 intermittently not emitted per 21CFR 1020.31(j) requirements.

    Cause as FDA categorized it: Radiation Control for Health and Safety Act

Showing the 8 newest of 43 recall records in FDA's data.

Back to all tools

For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.