Device type · MNB
Device, Thermal Ablation, Endometrial
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
MNB
Obstetrics & Gynecology
Device class
Class III
FDA decisions since 2015
4
510(k), De Novo and PMA
Adverse event reports
167
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Device, Thermal Ablation, Endometrial
- Product code
- MNB
- Device class
- Class III
- Review panel
- Obstetrics & Gynecology
- Specialty
- Obstetrics & Gynecology
- FDA record
- Classification database entry for MNB
Latest FDA decisions
P230002 · Decided Jul 28, 2023 · 179 days after receipt
Device type: Device, thermal ablation, endometrial (MNB) · Obstetrics & Gynecology
P180032 · Decided Mar 28, 2019 · 227 days after receipt
Device type: Device, thermal ablation, endometrial (MNB) · Obstetrics & Gynecology
P160047 · Decided Jun 14, 2017 · 232 days after receipt
Device type: Device, thermal ablation, endometrial (MNB) · Obstetrics & Gynecology
P140013 · Decided Jul 27, 2015 · 391 days after receipt
Device type: Device, thermal ablation, endometrial (MNB) · Obstetrics & Gynecology
Reports received
167
Received Nov 1, 2025 to Nov 19, 2026
Malfunction reports
76
46% of reports
Injury reports
91
54% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 4 | 12 | 0 | 0 |
| Jan 24 | 6 | 12 | 0 | 0 |
| Feb 24 | 20 | 5 | 0 | 0 |
| Mar 24 | 8 | 18 | 0 | 0 |
| Apr 24 | 7 | 5 | 0 | 0 |
| May 24 | 18 | 12 | 0 | 0 |
| Jun 24 | 9 | 9 | 0 | 0 |
| Jul 24 | 2 | 6 | 0 | 0 |
| Aug 24 | 6 | 2 | 0 | 0 |
| Sep 24 | 10 | 11 | 0 | 0 |
| Oct 24 | 6 | 7 | 0 | 0 |
| Nov 24 | 15 | 4 | 0 | 0 |
| Dec 24 | 6 | 9 | 0 | 0 |
| Jan 25 | 13 | 12 | 1 | 0 |
| Feb 25 | 5 | 6 | 0 | 0 |
| Mar 25 | 13 | 15 | 0 | 0 |
| Apr 25 | 9 | 24 | 0 | 0 |
| May 25 | 8 | 9 | 0 | 0 |
| Jun 25 | 12 | 13 | 0 | 0 |
| Jul 25 | 13 | 7 | 0 | 0 |
| Aug 25 | 16 | 11 | 1 | 0 |
| Sep 25 | 8 | 9 | 1 | 0 |
| Oct 25 | 7 | 11 | 0 | 0 |
| Nov 25 | 7 | 3 | 0 | 0 |
| Dec 25 | 6 | 11 | 0 | 0 |
| Jan 26 | 8 | 10 | 0 | 0 |
| Feb 26 | 7 | 13 | 0 | 0 |
| Mar 26 | 12 | 9 | 0 | 0 |
| Apr 26 | 5 | 6 | 0 | 0 |
| May 26 | 1 | 5 | 0 | 0 |
| Jun 26 | 9 | 10 | 0 | 0 |
| Jul 26 | 7 | 4 | 0 | 0 |
| Aug 26 | 2 | 5 | 0 | 0 |
| Sep 26 | 4 | 9 | 0 | 0 |
| Oct 26 | 8 | 6 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Hologic144
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Adverse Event Without Identified Device or Use Problem77
- Material Puncture/Hole35
- Detachment of Device or Device Component12
- Material Deformation11
- Use of Device Problem5
- Insufficient Device Problem Information4
- Material Disintegration3
- Melted3
- Activation Failure2
- Tear, Rip or Hole in Device Packaging2
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions70
- Uterine Perforation41
- Fever12
- Foreign Body In Patient9
- Bowel Perforation8
- Burn(s)8
- Sepsis8
- Abdominal Pain7
- Bowel Burn7
- Pain7
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-1584-2025 · Started Mar 20, 2025 · Posted by FDA Apr 16, 2025
Reason given: Due to evaporator gasket not sealing which can cause liquid nitrous oxide to drip/leak from the bottom of the device handle during treatment or after treatment during venting of the device
Cause as FDA categorized it: Under Investigation by firm
Z-1889-2024 · Started May 3, 2024 · Posted by FDA May 23, 2024
Reason given: Error code 003 was not listed in certain rows of the Device Instructions for Use Troubleshooting section (i.e., Table 22. Summary of LCD Messages and Error Codes).
Cause as FDA categorized it: Device Design
Z-1094-2014 · Started Jan 24, 2014 · Posted by FDA Feb 26, 2014
Reason given: NovaSure Radiofrequency Control Units may not meet a requirement of the electrical standard, IEC 60601-1,
Cause as FDA categorized it: Component design/selection
Z-2208-2009 · Started Jul 31, 2009 · Posted by FDA Sep 22, 2009
Reason given: Cracked procedure sheaths, incorrect care/use of device, and console malfunctions may result in fluid leaks and loss of a cervical seal that are responsible for burns.
Cause as FDA categorized it: Process design
Z-2209-2009 · Started Jul 31, 2009 · Posted by FDA Sep 22, 2009
Reason given: Cracked procedure sheaths, incorrect care/use of device, and console malfunctions may result in fluid leaks and loss of a cervical seal that are responsible for burns.
Cause as FDA categorized it: Process design
Z-0687-2007 · Started Jan 24, 2007 · Posted by FDA Mar 31, 2007
Reason given: Defective integrated circuit board could result in loss of the system pump and patient injury (hot fluid 90 degree C into uterus) if in use during the recirculation/heating phase of treatment
Cause as FDA categorized it: Other
Z-0688-2007 · Started Jan 24, 2007 · Posted by FDA Mar 31, 2007
Reason given: Defective integrated circuit board could result in loss of the system pump and patient injury (hot fluid 90 degree C into uterus) if in use during the recirculation/heating phase of treatment
Cause as FDA categorized it: Other
Z-0689-2007 · Started Jan 24, 2007 · Posted by FDA Mar 31, 2007
Reason given: Defective integrated circuit board could result in loss of the system pump and patient injury (hot fluid 90 degree C into uterus) if in use during the recirculation/heating phase of treatment
Cause as FDA categorized it: Other
Showing the 8 newest of 12 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.