Device type · QVV
Medial Knee Implanted Shock Absorber
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
QVV
Orthopedic
Device class
Class II
FDA decisions since 2015
1
510(k), De Novo and PMA
Adverse event reports
11
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Medial Knee Implanted Shock Absorber
- Product code
- QVV
- Device class
- Class II
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3610
- Specialty
- Orthopedic Surgery
- FDA record
- Classification database entry for QVV
Latest FDA decisions
- MISHA Knee SystemNew device typeBreakthroughDe Novo
DEN220033 · Decided Apr 10, 2023 · 308 days after receipt
Device type: Medial Knee Implanted Shock Absorber (QVV) · Orthopedic Surgery
Reports received
11
Received Nov 1, 2025 to Nov 19, 2026
Malfunction reports
2
18% of reports
Injury reports
9
82% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 0 | 0 | 0 | 0 |
| Jan 24 | 0 | 0 | 0 | 0 |
| Feb 24 | 0 | 0 | 0 | 0 |
| Mar 24 | 0 | 0 | 0 | 0 |
| Apr 24 | 0 | 1 | 0 | 0 |
| May 24 | 0 | 0 | 0 | 0 |
| Jun 24 | 0 | 0 | 0 | 0 |
| Jul 24 | 0 | 1 | 0 | 0 |
| Aug 24 | 0 | 0 | 0 | 0 |
| Sep 24 | 0 | 0 | 0 | 0 |
| Oct 24 | 0 | 0 | 0 | 0 |
| Nov 24 | 1 | 0 | 0 | 0 |
| Dec 24 | 0 | 0 | 0 | 0 |
| Jan 25 | 1 | 0 | 0 | 0 |
| Feb 25 | 1 | 0 | 0 | 0 |
| Mar 25 | 0 | 0 | 0 | 0 |
| Apr 25 | 1 | 0 | 0 | 0 |
| May 25 | 0 | 0 | 0 | 0 |
| Jun 25 | 0 | 0 | 0 | 0 |
| Jul 25 | 0 | 0 | 0 | 0 |
| Aug 25 | 0 | 0 | 0 | 0 |
| Sep 25 | 1 | 0 | 0 | 0 |
| Oct 25 | 0 | 0 | 0 | 0 |
| Nov 25 | 0 | 6 | 0 | 0 |
| Dec 25 | 0 | 0 | 0 | 0 |
| Jan 26 | 1 | 0 | 0 | 0 |
| Feb 26 | 0 | 0 | 0 | 0 |
| Mar 26 | 0 | 0 | 0 | 0 |
| Apr 26 | 0 | 0 | 0 | 0 |
| May 26 | 0 | 0 | 0 | 0 |
| Jun 26 | 1 | 0 | 0 | 0 |
| Jul 26 | 0 | 0 | 0 | 0 |
| Aug 26 | 0 | 0 | 0 | 0 |
| Sep 26 | 0 | 2 | 0 | 0 |
| Oct 26 | 0 | 1 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Moximed11
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Adverse Event Without Identified Device or Use Problem9
- Fracture1
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Post Operative Wound Infection5
- Loss of Range of Motion4
- No Clinical Signs, Symptoms or Conditions2
2 more in this ranking with a free account
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
- MISHA Knee System Implant Small, Left, REF: 2-1001. MISHA Knee System Implant Small, Right, REF: 2-1002.Open, Classified
Z-0762-2025 · Started Dec 2, 2024 · Posted by FDA Dec 23, 2024
Reason given: Knee implant may fracture due to a supplier-related issue with one component that has nonconforming geometry at the distal end of the absorber component that may result in a reduced wall thickness and microcracks, which may lead to new or worsening symptoms of discomfort, swelling / edema, pain or stiffness, which may...
Cause as FDA categorized it: Nonconforming Material/Component
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.