Device type · PID
Esophageal Dilator Balloon With Or Without Electrode Sensors
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
PID
Gastroenterology & Urology
Device class
Class II
FDA decisions since 2015
1
510(k), De Novo and PMA
Adverse event reports
16
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Esophageal Dilator Balloon With Or Without Electrode Sensors
- Product code
- PID
- Device class
- Class II
- Review panel
- Gastroenterology & Urology
- Regulation
- 21 CFR 876.5980
- Specialty
- GI, Urology & Nephrology, Gastroenterology
- FDA record
- Classification database entry for PID
Latest FDA decisions
K182895 · Decided Jan 25, 2019 · 102 days after receipt
Device type: Esophageal Dilator Balloon With Or Without Electrode Sensors (PID) · GI, Urology & Nephrology
Reports received
16
Received Nov 1, 2025 to Nov 19, 2026
Malfunction reports
8
50% of reports
Injury reports
8
50% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 0 | 0 | 0 | 0 |
| Jan 24 | 0 | 1 | 0 | 0 |
| Feb 24 | 0 | 0 | 0 | 0 |
| Mar 24 | 1 | 0 | 0 | 0 |
| Apr 24 | 0 | 0 | 0 | 0 |
| May 24 | 70 | 0 | 0 | 0 |
| Jun 24 | 4 | 3 | 0 | 0 |
| Jul 24 | 1 | 0 | 0 | 0 |
| Aug 24 | 1 | 0 | 0 | 0 |
| Sep 24 | 0 | 0 | 0 | 0 |
| Oct 24 | 0 | 0 | 0 | 0 |
| Nov 24 | 0 | 0 | 0 | 0 |
| Dec 24 | 2 | 0 | 0 | 0 |
| Jan 25 | 2 | 0 | 0 | 0 |
| Feb 25 | 3 | 0 | 0 | 0 |
| Mar 25 | 0 | 3 | 1 | 0 |
| Apr 25 | 0 | 0 | 0 | 0 |
| May 25 | 2 | 0 | 0 | 0 |
| Jun 25 | 0 | 0 | 0 | 0 |
| Jul 25 | 0 | 0 | 0 | 0 |
| Aug 25 | 0 | 0 | 0 | 0 |
| Sep 25 | 1 | 0 | 0 | 0 |
| Oct 25 | 0 | 0 | 0 | 0 |
| Nov 25 | 0 | 2 | 0 | 0 |
| Dec 25 | 4 | 1 | 0 | 0 |
| Jan 26 | 0 | 0 | 0 | 0 |
| Feb 26 | 0 | 0 | 0 | 0 |
| Mar 26 | 1 | 1 | 0 | 0 |
| Apr 26 | 1 | 0 | 0 | 0 |
| May 26 | 0 | 0 | 0 | 0 |
| Jun 26 | 2 | 0 | 0 | 0 |
| Jul 26 | 0 | 1 | 0 | 0 |
| Aug 26 | 0 | 0 | 0 | 0 |
| Sep 26 | 0 | 3 | 0 | 0 |
| Oct 26 | 0 | 0 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Adverse Event Without Identified Device or Use Problem7
- Contamination /Decontamination Problem4
- Device Difficult to Program or Calibrate2
- Insufficient Device Problem Information2
- Use of Device Problem2
- Burst Container or Vessel1
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions6
- Perforation of Esophagus5
- Insufficient Information2
- Perforation2
- Abscess1
- Pain1
- Unspecified Tissue Injury1
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
- EsoFLIP, 30mm BALLOON DILATION CATHETER, REF ES-330, Rx OnlyOpen, Classified
Z-2038-2024 · Started Apr 25, 2024 · Posted by FDA Jun 7, 2024
Reason given: Due to saline conductivity issues, their is a potential that dilation catheters may provide inaccurate esophageal diameter measurements that may lead to various harms.
Cause as FDA categorized it: Nonconforming Material/Component
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.