Device type · PFV
System, Endovascular Graft, Arteriovenous (Av) Dialysis Access Circuit Stenosis Treatment
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
PFV
Cardiovascular
Device class
Class III
FDA decisions since 2015
4
510(k), De Novo and PMA
Adverse event reports
276
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- System, Endovascular Graft, Arteriovenous (Av) Dialysis Access Circuit Stenosis Treatment
- Product code
- PFV
- Device class
- Class III
- Review panel
- Cardiovascular
- Specialty
- Cardiology, Interventional Cardiology
- FDA record
- Classification database entry for PFV
Latest FDA decisions
- WRAPSODY® Cell-Impermeable EndoprosthesisBreakthroughPMA
P240023 · Decided Dec 19, 2024 · 176 days after receipt
Device type: System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment (PFV) · Cardiology
- Covera Vascular Covered StentPMA supplement
P170042-S002 · Decided Mar 1, 2019 · 178 days after receipt · Panel Track
Device type: System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment (PFV) · Cardiology
P170042 · Decided Jul 30, 2018 · 230 days after receipt
Device type: System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment (PFV) · Cardiology
1 more FDA decisions with a free account
Reports received
276
Received Nov 1, 2025 to Nov 19, 2026
Malfunction reports
159
58% of reports
Injury reports
117
42% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 9 | 6 | 0 | 0 |
| Jan 24 | 8 | 3 | 0 | 0 |
| Feb 24 | 6 | 0 | 0 | 0 |
| Mar 24 | 14 | 1 | 0 | 0 |
| Apr 24 | 20 | 11 | 0 | 0 |
| May 24 | 12 | 2 | 0 | 0 |
| Jun 24 | 19 | 3 | 1 | 0 |
| Jul 24 | 26 | 4 | 0 | 0 |
| Aug 24 | 21 | 8 | 0 | 0 |
| Sep 24 | 6 | 3 | 0 | 0 |
| Oct 24 | 7 | 4 | 0 | 0 |
| Nov 24 | 13 | 2 | 0 | 0 |
| Dec 24 | 11 | 4 | 0 | 0 |
| Jan 25 | 8 | 2 | 0 | 0 |
| Feb 25 | 9 | 3 | 0 | 0 |
| Mar 25 | 16 | 4 | 0 | 0 |
| Apr 25 | 6 | 9 | 0 | 0 |
| May 25 | 12 | 1 | 0 | 0 |
| Jun 25 | 11 | 1 | 0 | 0 |
| Jul 25 | 11 | 3 | 0 | 0 |
| Aug 25 | 23 | 6 | 0 | 0 |
| Sep 25 | 17 | 6 | 0 | 0 |
| Oct 25 | 10 | 4 | 0 | 0 |
| Nov 25 | 10 | 7 | 0 | 0 |
| Dec 25 | 13 | 1 | 0 | 0 |
| Jan 26 | 7 | 0 | 0 | 0 |
| Feb 26 | 11 | 4 | 0 | 0 |
| Mar 26 | 17 | 1 | 0 | 0 |
| Apr 26 | 15 | 4 | 0 | 0 |
| May 26 | 18 | 11 | 0 | 0 |
| Jun 26 | 12 | 23 | 0 | 0 |
| Jul 26 | 20 | 26 | 0 | 0 |
| Aug 26 | 10 | 21 | 0 | 0 |
| Sep 26 | 16 | 9 | 0 | 0 |
| Oct 26 | 10 | 10 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
2 more in this ranking with a free account
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Adverse Event Without Identified Device or Use Problem71
- Misfire57
- Positioning Failure47
7 more in this ranking with a free account
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions160
- Thrombosis/Thrombus33
- Restenosis25
7 more in this ranking with a free account
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-1108-2023 · Started Dec 22, 2022 · Posted by FDA Feb 9, 2023
Reason given: An increase in the reported complaint rate for inability to deploy was identified and related to poor slide block bonding.
Cause as FDA categorized it: Unknown/Undetermined by firm
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.