Device type · PAH
Mesh, Surgical, Synthetic, Urogynecologic, For Stress Urinary Incontinence, Female, Mini-Sling
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
PAH
Gastroenterology & Urology
Device class
Class II
FDA decisions since 2015
3
510(k), De Novo and PMA
Adverse event reports
283
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Mesh, Surgical, Synthetic, Urogynecologic, For Stress Urinary Incontinence, Female, Mini-Sling
- Product code
- PAH
- Device class
- Class II
- Review panel
- Gastroenterology & Urology
- Regulation
- 21 CFR 878.3300
- Specialty
- GI, Urology & Nephrology, Urology
- FDA record
- Classification database entry for PAH
Latest FDA decisions
K242473 · Decided Oct 18, 2024 · 59 days after receipt
Device type: Mesh, Surgical, Synthetic, Urogynecologic, For Stress Urinary Incontinence, Female, Mini-Sling (PAH) · General & Plastic Surgery
K221874 · Decided Feb 15, 2023 · 232 days after receipt
Device type: Mesh, Surgical, Synthetic, Urogynecologic, For Stress Urinary Incontinence, Female, Mini-Sling (PAH) · General & Plastic Surgery
- Desara One Single Incision Sling System, 1 Pack ; Desara One Single Incision Sling System, 3 Pack510(k)
K191416 · Decided Feb 7, 2020 · 255 days after receipt
Device type: Mesh, Surgical, Synthetic, Urogynecologic, For Stress Urinary Incontinence, Female, Mini-Sling (PAH) · General & Plastic Surgery
Reports received
283
Received Nov 1, 2025 to Nov 19, 2026. Includes 2 of another or unstated type.
Malfunction reports
206
73% of reports
Injury reports
75
27% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 17 | 8 | 0 | 0 |
| Jan 24 | 16 | 7 | 0 | 0 |
| Feb 24 | 12 | 23 | 0 | 0 |
| Mar 24 | 14 | 3 | 0 | 0 |
| Apr 24 | 11 | 6 | 0 | 0 |
| May 24 | 20 | 5 | 0 | 0 |
| Jun 24 | 19 | 3 | 0 | 0 |
| Jul 24 | 18 | 8 | 0 | 0 |
| Aug 24 | 11 | 10 | 0 | 0 |
| Sep 24 | 14 | 10 | 0 | 0 |
| Oct 24 | 13 | 21 | 0 | 0 |
| Nov 24 | 12 | 7 | 0 | 0 |
| Dec 24 | 18 | 6 | 0 | 0 |
| Jan 25 | 11 | 2 | 0 | 0 |
| Feb 25 | 17 | 6 | 0 | 0 |
| Mar 25 | 15 | 3 | 0 | 0 |
| Apr 25 | 20 | 16 | 0 | 0 |
| May 25 | 16 | 10 | 0 | 0 |
| Jun 25 | 6 | 11 | 0 | 0 |
| Jul 25 | 17 | 4 | 0 | 0 |
| Aug 25 | 9 | 6 | 0 | 0 |
| Sep 25 | 21 | 4 | 0 | 0 |
| Oct 25 | 17 | 4 | 0 | 0 |
| Nov 25 | 5 | 6 | 0 | 0 |
| Dec 25 | 12 | 7 | 0 | 0 |
| Jan 26 | 21 | 14 | 0 | 0 |
| Feb 26 | 23 | 4 | 0 | 0 |
| Mar 26 | 19 | 2 | 0 | 0 |
| Apr 26 | 13 | 3 | 0 | 0 |
| May 26 | 18 | 6 | 0 | 0 |
| Jun 26 | 36 | 10 | 0 | 0 |
| Jul 26 | 21 | 11 | 0 | 0 |
| Aug 26 | 9 | 9 | 0 | 0 |
| Sep 26 | 21 | 2 | 0 | 1 |
| Oct 26 | 8 | 1 | 0 | 1 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Coloplast245
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Break161
- Adverse Event Without Identified Device or Use Problem47
- Material Fragmentation24
- Device Slipped11
- Device Appears to Trigger Rejection7
- Material Separation7
- Material Twisted/Bent6
- Material Deformation5
- Detachment of Device or Device Component4
- Difficult to Advance4
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions210
- Erosion37
- Pain29
- Dyspareunia28
- Urinary Incontinence28
- Pelvic Pain20
- Urinary Tract Infection19
- Abnormal Vaginal Discharge18
- Micturition Urgency17
- Urinary Frequency / Polyuria17
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-0279-2016 · Started Oct 26, 2015 · Posted by FDA Nov 16, 2015
Reason given: The lot number on the outer carton label did not match the lot number printed on the tray label or identification labels inside the box. The identification labels inside the box are placed on medical records and other documents associated with the procedure. Use of incorrectly identified labels may prevent direct...
Cause as FDA categorized it: Labeling mix-ups
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.