Device type · OZD
Temporary Non-Roller Type Left Heart Support Blood Pump
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
OZD
Cardiovascular
Device class
Class III
FDA decisions since 2015
4
510(k), De Novo and PMA
Adverse event reports
14,156
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Temporary Non-Roller Type Left Heart Support Blood Pump
- Product code
- OZD
- Device class
- Class III
- Review panel
- Cardiovascular
- Specialty
- Cardiology
- FDA record
- Classification database entry for OZD
Latest FDA decisions
- Impella Ventricular Support SystemsPMA supplement
P140003-S018 · Decided Feb 7, 2018 · 310 days after receipt · Panel Track
Device type: Temporary non-roller type left heart support blood pump (OZD) · Cardiology
- IMPELLA VENTRICULAR SUPPORT SYSTEMSPMA supplement
P140003-S004 · Decided Apr 7, 2016 · 239 days after receipt · Panel Track
Device type: Temporary non-roller type left heart support blood pump (OZD) · Cardiology
- IMPELLA LEFT VENTRICULAR SUPPORT SYSTEMPMA supplement
P140003-S005 · Decided Apr 7, 2016 · 238 days after receipt · Panel Track
Device type: Temporary non-roller type left heart support blood pump (OZD) · Cardiology
1 more FDA decisions with a free account
Reports received
14,156
Received Nov 1, 2025 to Nov 19, 2026. Includes 2 of another or unstated type.
Malfunction reports
5,130
36% of reports
Injury reports
6,785
48% of reports
Death reports
2,239
16% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 248 | 200 | 28 | 0 |
| Jan 24 | 418 | 337 | 36 | 0 |
| Feb 24 | 347 | 336 | 30 | 0 |
| Mar 24 | 576 | 1,130 | 96 | 0 |
| Apr 24 | 449 | 976 | 32 | 0 |
| May 24 | 354 | 857 | 52 | 0 |
| Jun 24 | 317 | 527 | 42 | 0 |
| Jul 24 | 268 | 464 | 25 | 0 |
| Aug 24 | 794 | 2,362 | 623 | 0 |
| Sep 24 | 500 | 416 | 43 | 0 |
| Oct 24 | 1,604 | 850 | 287 | 1 |
| Nov 24 | 1,495 | 723 | 476 | 0 |
| Dec 24 | 392 | 244 | 113 | 0 |
| Jan 25 | 248 | 283 | 110 | 0 |
| Feb 25 | 221 | 186 | 79 | 0 |
| Mar 25 | 265 | 258 | 126 | 0 |
| Apr 25 | 354 | 331 | 138 | 0 |
| May 25 | 289 | 420 | 141 | 0 |
| Jun 25 | 330 | 464 | 185 | 0 |
| Jul 25 | 337 | 446 | 162 | 0 |
| Aug 25 | 219 | 181 | 92 | 0 |
| Sep 25 | 744 | 472 | 236 | 0 |
| Oct 25 | 307 | 445 | 200 | 0 |
| Nov 25 | 176 | 360 | 124 | 0 |
| Dec 25 | 309 | 482 | 163 | 0 |
| Jan 26 | 517 | 915 | 340 | 0 |
| Feb 26 | 331 | 1,252 | 428 | 0 |
| Mar 26 | 400 | 940 | 312 | 0 |
| Apr 26 | 340 | 505 | 157 | 0 |
| May 26 | 320 | 375 | 169 | 0 |
| Jun 26 | 1,122 | 419 | 152 | 0 |
| Jul 26 | 487 | 363 | 122 | 0 |
| Aug 26 | 429 | 459 | 129 | 1 |
| Sep 26 | 461 | 477 | 102 | 1 |
| Oct 26 | 238 | 238 | 41 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
1 more in this ranking with a free account
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Adverse Event Without Identified Device or Use Problem3,998
- Patient Device Interaction Problem2,675
- Device Sensing Problem1,530
7 more in this ranking with a free account
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions6,202
- Hemorrhage/Blood Loss/Bleeding2,385
- Hemolysis1,354
7 more in this ranking with a free account
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
- Automated Impella Controller, Packaged. Primary user control interface for the Impella Catheter, catheter-based heart pump.Open, Classified
Z-0208-2027 · Started Sep 18, 2026 · Posted by FDA Oct 28, 2026
Reason given: Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. 2....
Cause as FDA categorized it: Under Investigation by firm
- Impella Optical Controller, Packaged. Primary user control interface for the Impella Catheter, a catheter-based heart pump.Open, Classified
Z-0209-2027 · Started Sep 18, 2026 · Posted by FDA Oct 28, 2026
Reason given: Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. 2....
Cause as FDA categorized it: Under Investigation by firm
- Impella Connect Optical AIC, Packaged. Primary user control interface for the Impella Catheter, a catheter-based heart pump.Open, Classified
Z-0210-2027 · Started Sep 18, 2026 · Posted by FDA Oct 28, 2026
Reason given: Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. 2....
Cause as FDA categorized it: Under Investigation by firm
47 more recall records with a free account
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.