Device type · OAS
X-Ray, Tomography, Computed, Dental
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
OAS
Radiology
Device class
Class II
FDA decisions since 2015
99
510(k), De Novo and PMA
Adverse event reports
1
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- X-Ray, Tomography, Computed, Dental
- Product code
- OAS
- Device class
- Class II
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1750
- Specialty
- Radiology & Imaging
- FDA record
- Classification database entry for OAS
Latest FDA decisions
The 8 newest of 99 since 2015.
K263033 · Decided Sep 30, 2026 · 30 days after receipt
Device type: X-Ray, Tomography, Computed, Dental (OAS) · Radiology & Imaging
Hefei Meyer Optoelectronic Technology, Inc.
K261748 · Decided Jul 23, 2026 · 56 days after receipt
Device type: X-Ray, Tomography, Computed, Dental (OAS) · Radiology & Imaging
- Dental X-ray System510(k)
Changzhou Boen Zhongding Medical Technology Co., Ltd.
K250839 · Decided Jun 11, 2026 · 448 days after receipt
Device type: X-Ray, Tomography, Computed, Dental (OAS) · Radiology & Imaging
Yofo Medical Technology Co., Ltd.
K260662 · Decided Jun 4, 2026 · 94 days after receipt
Device type: X-Ray, Tomography, Computed, Dental (OAS) · Radiology & Imaging
- medXion NEXUS510(k)
K253935 · Decided May 15, 2026 · 157 days after receipt
Device type: X-Ray, Tomography, Computed, Dental (OAS) · Radiology & Imaging
- RCT800510(k)
K260462 · Decided Apr 9, 2026 · 56 days after receipt
Device type: X-Ray, Tomography, Computed, Dental (OAS) · Radiology & Imaging
K260093 · Decided Mar 31, 2026 · 77 days after receipt
Device type: X-Ray, Tomography, Computed, Dental (OAS) · Radiology & Imaging
- myray ProXIma X6510(k)
K252353 · Decided Mar 10, 2026 · 224 days after receipt
Device type: X-Ray, Tomography, Computed, Dental (OAS) · Radiology & Imaging
Reports received
1
Received Nov 1, 2025 to Nov 19, 2026. Includes 1 of another or unstated type.
Malfunction reports
0
0.0% of reports
Injury reports
0
0.0% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 0 | 0 | 0 | 0 |
| Jan 24 | 0 | 0 | 0 | 0 |
| Feb 24 | 0 | 0 | 0 | 0 |
| Mar 24 | 0 | 0 | 0 | 0 |
| Apr 24 | 0 | 0 | 0 | 0 |
| May 24 | 0 | 0 | 0 | 0 |
| Jun 24 | 0 | 0 | 0 | 0 |
| Jul 24 | 0 | 0 | 0 | 0 |
| Aug 24 | 0 | 0 | 0 | 0 |
| Sep 24 | 0 | 1 | 0 | 0 |
| Oct 24 | 0 | 1 | 0 | 0 |
| Nov 24 | 0 | 0 | 0 | 0 |
| Dec 24 | 0 | 0 | 0 | 0 |
| Jan 25 | 0 | 0 | 0 | 0 |
| Feb 25 | 0 | 0 | 0 | 0 |
| Mar 25 | 0 | 0 | 0 | 0 |
| Apr 25 | 0 | 0 | 0 | 0 |
| May 25 | 0 | 0 | 0 | 0 |
| Jun 25 | 0 | 0 | 0 | 0 |
| Jul 25 | 0 | 0 | 0 | 0 |
| Aug 25 | 0 | 0 | 0 | 0 |
| Sep 25 | 0 | 1 | 0 | 0 |
| Oct 25 | 0 | 0 | 0 | 0 |
| Nov 25 | 0 | 0 | 0 | 0 |
| Dec 25 | 0 | 0 | 0 | 0 |
| Jan 26 | 0 | 0 | 0 | 0 |
| Feb 26 | 0 | 0 | 0 | 0 |
| Mar 26 | 0 | 0 | 0 | 0 |
| Apr 26 | 0 | 0 | 0 | 0 |
| May 26 | 0 | 0 | 0 | 0 |
| Jun 26 | 0 | 0 | 0 | 0 |
| Jul 26 | 0 | 0 | 0 | 0 |
| Aug 26 | 0 | 0 | 0 | 1 |
| Sep 26 | 0 | 0 | 0 | 0 |
| Oct 26 | 0 | 0 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
No manufacturers to rank.
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Mechanical Problem1
- Physical Resistance/Sticking1
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Physical Entrapment1
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-0313-2026 · Started Aug 26, 2025 · Posted by FDA Nov 5, 2025
Reason given: Osstem Implant Co. Ltd, is voluntarily initiating a medical device correction involving the current labeling for this electronic product may be missing a required Certification Statement under 21 CFR Subchapter J (Radiological Health Regulations). Specifically, the following statement may have been omitted:...
Cause as FDA categorized it: Radiation Control for Health and Safety Act
Z-0314-2026 · Started Aug 26, 2025 · Posted by FDA Nov 5, 2025
Reason given: Osstem Implant Co. Ltd, is voluntarily initiating a medical device correction involving the current labeling for this electronic product may be missing a required Certification Statement under 21 CFR Subchapter J (Radiological Health Regulations). Specifically, the following statement may have been omitted:...
Cause as FDA categorized it: Radiation Control for Health and Safety Act
Z-0315-2026 · Started Aug 26, 2025 · Posted by FDA Nov 5, 2025
Reason given: Osstem Implant Co. Ltd, is voluntarily initiating a medical device correction involving the current labeling for this electronic product may be missing a required Certification Statement under 21 CFR Subchapter J (Radiological Health Regulations). Specifically, the following statement may have been omitted:...
Cause as FDA categorized it: Radiation Control for Health and Safety Act
Z-2413-2024 · Started Oct 28, 2021 · Posted by FDA Aug 12, 2024
Reason given: It was noticed in production on 28.10.2021, that the final testing database gives a pass result for Pan ADC repeatability test even when the test result is not within the set acceptance criteria (repeatability within 4%).
Cause as FDA categorized it: Radiation Control for Health and Safety Act
Z-1627-2023 · Started Jul 21, 2021 · Posted by FDA May 27, 2023
Reason given: Device failed to function as intended and/or failed to conform to their design specifications.
Cause as FDA categorized it: Radiation Control for Health and Safety Act
- cone-beam CT systemOpen, Classified
Z-1101-2019 · Started Dec 31, 2018
Reason given: ORTHOPANTOMOGRAPH OP 3D device has a defect in the device firmware versions 2.1.0 and 2.1.1
Cause as FDA categorized it: Device Design
- Owandy Radiology I -MAX Touch 3D X-rayTerminated
Z-0129-2016 · Started Aug 5, 2015 · Posted by FDA Nov 6, 2015
Reason given: It has been discovered that due to lack of information accompanying documents the product fails to comply with 21CfR 1020.33(c),1020.33(d),1020.33(g),and 1020.330).
Cause as FDA categorized it: Other
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.