Device type · NWB
Endoscope, Accessories, Narrow Band Spectrum
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
NWB
Gastroenterology & Urology
Device class
Class II
FDA decisions since 2015
3
510(k), De Novo and PMA
Adverse event reports
3,182
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Endoscope, Accessories, Narrow Band Spectrum
- Product code
- NWB
- Device class
- Class II
- Review panel
- Gastroenterology & Urology
- Regulation
- 21 CFR 876.1500
- Specialty
- GI, Urology & Nephrology, Gastroenterology
- FDA record
- Classification database entry for NWB
Latest FDA decisions
K222020 · Decided Sep 8, 2022 · 62 days after receipt
Device type: Endoscope, Accessories, Narrow Band Spectrum (NWB) · GI, Urology & Nephrology
K211882 · Decided Apr 1, 2022 · 284 days after receipt
Device type: Endoscope, Accessories, Narrow Band Spectrum (NWB) · GI, Urology & Nephrology
K173921 · Decided Sep 5, 2018 · 253 days after receipt
Device type: Endoscope, Accessories, Narrow Band Spectrum (NWB) · GI, Urology & Nephrology
Reports received
3,182
Received Nov 1, 2025 to Nov 19, 2026. Includes 1 of another or unstated type.
Malfunction reports
3,176
100% of reports
Injury reports
5
0.2% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 191 | 4 | 0 | 0 |
| Jan 24 | 156 | 2 | 0 | 0 |
| Feb 24 | 193 | 3 | 0 | 0 |
| Mar 24 | 206 | 3 | 0 | 0 |
| Apr 24 | 231 | 0 | 0 | 0 |
| May 24 | 238 | 1 | 0 | 0 |
| Jun 24 | 202 | 3 | 0 | 0 |
| Jul 24 | 250 | 6 | 0 | 0 |
| Aug 24 | 273 | 1 | 0 | 0 |
| Sep 24 | 246 | 2 | 0 | 0 |
| Oct 24 | 263 | 4 | 0 | 0 |
| Nov 24 | 132 | 1 | 0 | 0 |
| Dec 24 | 232 | 1 | 0 | 0 |
| Jan 25 | 244 | 5 | 0 | 0 |
| Feb 25 | 313 | 1 | 0 | 0 |
| Mar 25 | 248 | 1 | 0 | 0 |
| Apr 25 | 244 | 0 | 0 | 0 |
| May 25 | 172 | 1 | 0 | 0 |
| Jun 25 | 168 | 1 | 0 | 0 |
| Jul 25 | 204 | 0 | 0 | 0 |
| Aug 25 | 221 | 1 | 0 | 0 |
| Sep 25 | 183 | 2 | 0 | 0 |
| Oct 25 | 377 | 1 | 0 | 0 |
| Nov 25 | 238 | 0 | 0 | 0 |
| Dec 25 | 263 | 1 | 0 | 0 |
| Jan 26 | 238 | 0 | 0 | 0 |
| Feb 26 | 261 | 3 | 0 | 0 |
| Mar 26 | 294 | 0 | 0 | 0 |
| Apr 26 | 359 | 0 | 0 | 0 |
| May 26 | 313 | 0 | 0 | 0 |
| Jun 26 | 234 | 0 | 0 | 0 |
| Jul 26 | 271 | 0 | 0 | 0 |
| Aug 26 | 250 | 1 | 0 | 0 |
| Sep 26 | 260 | 0 | 0 | 0 |
| Oct 26 | 195 | 0 | 0 | 1 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Shirakawa Olympus2,726
- Olympus453
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Communication or Transmission Problem697
- Poor Quality Image475
- No Display/Image413
- Output Problem322
- Activation Failure313
- Erratic or Intermittent Display146
- Break130
- Corroded128
- Connection Problem97
- Failure to Power Up86
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions3,171
- Insufficient Information6
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Laceration(s)2
- Unspecified Infection2
- Asystole1
- Cardiac Arrest1
- Cough1
- Discomfort1
- Foreign Body In Patient1
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Olympus Corporation of the Americas
Z-1911-2025 · Started May 5, 2025 · Posted by FDA Jun 5, 2025
Reason given: Recent post-market surveillance data suggests a possible association of higher microbial contamination levels from TJF duodenoscopes when manual cleaning was delayed beyond one hour and a presoak was performed, compared to those TJF duodenoscopes where manual cleaning began within one hour after patient procedure.
Cause as FDA categorized it: Labeling design
- Colonoscope, Model Number CF-H180AL.Open, Classified
Olympus Corporation of the Americas
Z-1205-2024 · Started Jan 11, 2024 · Posted by FDA Feb 29, 2024
Reason given: Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.
Cause as FDA categorized it: Nonconforming Material/Component
- Colonoscope, Model Number CF-Q180AL.Open, Classified
Olympus Corporation of the Americas
Z-1208-2024 · Started Jan 11, 2024 · Posted by FDA Feb 29, 2024
Reason given: Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.
Cause as FDA categorized it: Nonconforming Material/Component
- Colonoscope, Model Number PCF-H180AL.Open, Classified
Olympus Corporation of the Americas
Z-1209-2024 · Started Jan 11, 2024 · Posted by FDA Feb 29, 2024
Reason given: Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.
Cause as FDA categorized it: Nonconforming Material/Component
- Colonoscope, Model Number PCF-H190DL.Open, Classified
Olympus Corporation of the Americas
Z-1210-2024 · Started Jan 11, 2024 · Posted by FDA Feb 29, 2024
Reason given: Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.
Cause as FDA categorized it: Nonconforming Material/Component
- Colonoscope, Model Number PCF-H190TL.Open, Classified
Olympus Corporation of the Americas
Z-1212-2024 · Started Jan 11, 2024 · Posted by FDA Feb 29, 2024
Reason given: Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.
Cause as FDA categorized it: Nonconforming Material/Component
- Colonoscope, Model Number PCF-HQ190L.Open, Classified
Olympus Corporation of the Americas
Z-1213-2024 · Started Jan 11, 2024 · Posted by FDA Feb 29, 2024
Reason given: Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.
Cause as FDA categorized it: Nonconforming Material/Component
- Colonoscope, Model Number PCF-Q180AL.Open, Classified
Olympus Corporation of the Americas
Z-1214-2024 · Started Jan 11, 2024 · Posted by FDA Feb 29, 2024
Reason given: Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.
Cause as FDA categorized it: Nonconforming Material/Component
Showing the 8 newest of 19 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.