Device type · NSA
Over-The-Counter Automated External Defibrillator
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
NSA
Cardiovascular
Device class
Class III
FDA decisions since 2015
1
510(k), De Novo and PMA
Adverse event reports
685
Received Nov 1, 2025 to Nov 20, 2026
How FDA classifies it
- Device type
- Over-The-Counter Automated External Defibrillator
- Product code
- NSA
- Device class
- Class III
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.5310
- Specialty
- Cardiology, Electrophysiology
- FDA record
- Classification database entry for NSA
Latest FDA decisions
P160029 · Decided Jun 6, 2019 · 1,042 days after receipt
Device type: Over-the-counter automated external defibrillator (NSA) · Cardiology
Reports received
685
Received Nov 1, 2025 to Nov 20, 2026
Malfunction reports
664
97% of reports
Injury reports
8
1.2% of reports
Death reports
13
1.9% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 5 | 0 | 0 | 0 |
| Jan 24 | 5 | 0 | 0 | 0 |
| Feb 24 | 4 | 0 | 0 | 0 |
| Mar 24 | 3 | 0 | 0 | 0 |
| Apr 24 | 6 | 0 | 1 | 0 |
| May 24 | 5 | 0 | 0 | 0 |
| Jun 24 | 15 | 0 | 0 | 0 |
| Jul 24 | 11 | 0 | 0 | 0 |
| Aug 24 | 12 | 0 | 0 | 0 |
| Sep 24 | 3 | 0 | 1 | 0 |
| Oct 24 | 9 | 0 | 2 | 0 |
| Nov 24 | 6 | 0 | 0 | 0 |
| Dec 24 | 8 | 0 | 1 | 0 |
| Jan 25 | 8 | 0 | 0 | 0 |
| Feb 25 | 11 | 0 | 1 | 0 |
| Mar 25 | 6 | 0 | 0 | 0 |
| Apr 25 | 14 | 1 | 2 | 0 |
| May 25 | 29 | 0 | 0 | 0 |
| Jun 25 | 32 | 2 | 0 | 0 |
| Jul 25 | 55 | 2 | 0 | 0 |
| Aug 25 | 50 | 1 | 0 | 0 |
| Sep 25 | 94 | 1 | 1 | 0 |
| Oct 25 | 97 | 0 | 1 | 0 |
| Nov 25 | 45 | 0 | 1 | 0 |
| Dec 25 | 116 | 2 | 1 | 0 |
| Jan 26 | 49 | 1 | 2 | 0 |
| Feb 26 | 53 | 0 | 2 | 0 |
| Mar 26 | 72 | 2 | 2 | 0 |
| Apr 26 | 45 | 0 | 0 | 0 |
| May 26 | 44 | 1 | 1 | 0 |
| Jun 26 | 35 | 1 | 0 | 0 |
| Jul 26 | 61 | 0 | 1 | 0 |
| Aug 26 | 49 | 1 | 1 | 0 |
| Sep 26 | 46 | 0 | 1 | 0 |
| Oct 26 | 49 | 0 | 1 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Philips624
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Failure of Device to Self-Test365
- Failure to Power Up83
- Failure to Run on Battery54
7 more in this ranking with a free account
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions669
- Cardiac Arrest16
- Insufficient Information3
3 more in this ranking with a free account
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-1196-2010 · Started Nov 2, 2009 · Posted by FDA Mar 23, 2010
Reason given: Potential for failure of a capacitor during use which would prevent the AED from delivering effective defibrillation therapy when indicated.
Cause as FDA categorized it: Nonconforming Material/Component
Z-1781-2008 · Started Apr 30, 2008 · Posted by FDA Jul 9, 2008
Reason given: Buttons Sticking - The On/Off and Shock buttons on the front panel of the device may stick in place when pressed and fail to respond to the button press, causing a delay in the delivery of a defibrillation shock when the system advises a shock.
Cause as FDA categorized it: Component design/selection
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.