Device type · NGZ
Susceptibility Test Plate, Antifungal
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
NGZ
Microbiology
Device class
Class II
FDA decisions since 2015
10
510(k), De Novo and PMA
Adverse event reports
0
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Susceptibility Test Plate, Antifungal
- Product code
- NGZ
- Device class
- Class II
- Review panel
- Microbiology
- Regulation
- 21 CFR 866.1640
- Specialty
- Laboratory Medicine & Pathology
- FDA record
- Classification database entry for NGZ
Latest FDA decisions
The 8 newest of 10 since 2015.
K243738 · Decided Feb 27, 2025 · 85 days after receipt
Device type: Susceptibility Test Plate, Antifungal (NGZ) · Laboratory Medicine & Pathology
K232967 · Decided Jun 12, 2024 · 265 days after receipt
Device type: Susceptibility Test Plate, Antifungal (NGZ) · Laboratory Medicine & Pathology
K232963 · Decided Jun 7, 2024 · 260 days after receipt
Device type: Susceptibility Test Plate, Antifungal (NGZ) · Laboratory Medicine & Pathology
7 more FDA decisions with a free account
Reports received
0
Received Nov 1, 2025 to Nov 19, 2026
Malfunction reports
0
Injury reports
0
Death reports
0
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 0 | 0 | 0 | 0 |
| Jan 24 | 0 | 0 | 0 | 0 |
| Feb 24 | 1 | 0 | 0 | 0 |
| Mar 24 | 0 | 0 | 0 | 0 |
| Apr 24 | 0 | 0 | 0 | 0 |
| May 24 | 0 | 0 | 0 | 0 |
| Jun 24 | 0 | 0 | 0 | 0 |
| Jul 24 | 0 | 0 | 0 | 0 |
| Aug 24 | 0 | 0 | 0 | 0 |
| Sep 24 | 0 | 0 | 0 | 0 |
| Oct 24 | 0 | 0 | 0 | 0 |
| Nov 24 | 0 | 0 | 0 | 0 |
| Dec 24 | 0 | 0 | 0 | 0 |
| Jan 25 | 0 | 0 | 0 | 0 |
| Feb 25 | 0 | 0 | 0 | 0 |
| Mar 25 | 0 | 0 | 0 | 0 |
| Apr 25 | 0 | 0 | 0 | 0 |
| May 25 | 0 | 0 | 0 | 0 |
| Jun 25 | 0 | 0 | 0 | 0 |
| Jul 25 | 0 | 0 | 0 | 0 |
| Aug 25 | 0 | 0 | 0 | 0 |
| Sep 25 | 0 | 0 | 0 | 0 |
| Oct 25 | 0 | 0 | 0 | 0 |
| Nov 25 | 0 | 0 | 0 | 0 |
| Dec 25 | 0 | 0 | 0 | 0 |
| Jan 26 | 0 | 0 | 0 | 0 |
| Feb 26 | 0 | 0 | 0 | 0 |
| Mar 26 | 0 | 0 | 0 | 0 |
| Apr 26 | 0 | 0 | 0 | 0 |
| May 26 | 0 | 0 | 0 | 0 |
| Jun 26 | 0 | 0 | 0 | 0 |
| Jul 26 | 0 | 0 | 0 | 0 |
| Aug 26 | 0 | 0 | 0 | 0 |
| Sep 26 | 0 | 0 | 0 | 0 |
| Oct 26 | 0 | 0 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
No manufacturers to rank.
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
No device problems coded.
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
No patient problems coded.
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
- Yeastone Broth, 11ML, 10/BOX, used in antibiotic susceptibility testing of mycological clinical specimensOpen, Classified
Z-0355-2027 · Started Oct 19, 2026 · Posted by FDA Nov 3, 2026
Reason given: Affected lot did not meet specifications. Product may report false susceptibility to antifungal agents, specifically within the Azole family.
Cause as FDA categorized it: Under Investigation by firm
- Yeastone Broth, 11ML, 10/BOX YY3462Open, Classified
Z-1806-2026 · Started Mar 24, 2026 · Posted by FDA Apr 10, 2026
Reason given: Products may report incorrect AST results during quality control. Health consequences may include delayed response or the need to change to another antifungal agent if laboratory ignores Quality Control test result.
Cause as FDA categorized it: Under Investigation by firm
- VITEK 2 REAGENT AST-GP67 TEST KIT 20 CARDS, CATALOG 22226Open, Classified
Z-0792-2023 · Started Nov 9, 2022 · Posted by FDA Jan 4, 2023
Reason given: As the temperature and time out of range was exceeded, product performances cannot be guaranteed
Cause as FDA categorized it: Storage
2 more recall records with a free account
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.