Device type · MWJ
Electrocardiograph, Ambulatory (Without Analysis)
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
MWJ
Cardiovascular
Device class
Class II
FDA decisions since 2015
17
510(k), De Novo and PMA
Adverse event reports
3
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Electrocardiograph, Ambulatory (Without Analysis)
- Product code
- MWJ
- Device class
- Class II
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.2800
- Specialty
- Cardiology, Electrophysiology
- FDA record
- Classification database entry for MWJ
Latest FDA decisions
The 8 newest of 17 since 2015.
- iCOR Beat510(k)
K260383 · Decided Nov 2, 2026 · 270 days after receipt
Device type: Electrocardiograph, Ambulatory (Without Analysis) (MWJ) · Cardiology
Shenzhen Xinyan Medical Technology Co.,ltd.
K254119 · Decided Sep 11, 2026 · 266 days after receipt
Device type: Electrocardiograph, Ambulatory (Without Analysis) (MWJ) · Cardiology
K253529 · Decided Sep 4, 2026 · 295 days after receipt
Device type: Electrocardiograph, Ambulatory (Without Analysis) (MWJ) · Cardiology
- medilogFD510(k)
K253562 · Decided Aug 14, 2026 · 270 days after receipt
Device type: Electrocardiograph, Ambulatory (Without Analysis) (MWJ) · Cardiology
K253384 · Decided Jun 25, 2026 · 268 days after receipt
Device type: Electrocardiograph, Ambulatory (Without Analysis) (MWJ) · Cardiology
K252839 · Decided Jun 18, 2026 · 286 days after receipt
Device type: Electrocardiograph, Ambulatory (Without Analysis) (MWJ) · Cardiology
K231424 · Decided Dec 13, 2024 · 576 days after receipt
Device type: Electrocardiograph, Ambulatory (Without Analysis) (MWJ) · Cardiology
- ASSURE Wearable ECG510(k)
Kestra Medical Technologies, Inc.
K233864 · Decided May 7, 2024 · 153 days after receipt
Device type: Electrocardiograph, Ambulatory (Without Analysis) (MWJ) · Cardiology
Reports received
3
Received Nov 1, 2025 to Nov 19, 2026. Includes 1 of another or unstated type.
Malfunction reports
2
67% of reports
Injury reports
0
0.0% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 0 | 0 | 0 | 0 |
| Jan 24 | 0 | 0 | 0 | 0 |
| Feb 24 | 0 | 0 | 0 | 0 |
| Mar 24 | 0 | 0 | 0 | 0 |
| Apr 24 | 0 | 0 | 0 | 0 |
| May 24 | 1 | 0 | 0 | 0 |
| Jun 24 | 0 | 0 | 0 | 0 |
| Jul 24 | 0 | 0 | 0 | 0 |
| Aug 24 | 0 | 0 | 0 | 0 |
| Sep 24 | 0 | 1 | 0 | 0 |
| Oct 24 | 0 | 0 | 0 | 0 |
| Nov 24 | 1 | 0 | 0 | 0 |
| Dec 24 | 0 | 0 | 0 | 0 |
| Jan 25 | 0 | 1 | 0 | 0 |
| Feb 25 | 0 | 0 | 0 | 0 |
| Mar 25 | 0 | 0 | 0 | 0 |
| Apr 25 | 0 | 1 | 0 | 0 |
| May 25 | 0 | 1 | 0 | 0 |
| Jun 25 | 0 | 0 | 0 | 0 |
| Jul 25 | 1 | 0 | 0 | 0 |
| Aug 25 | 1 | 0 | 0 | 0 |
| Sep 25 | 0 | 0 | 0 | 0 |
| Oct 25 | 2 | 0 | 0 | 0 |
| Nov 25 | 0 | 0 | 0 | 0 |
| Dec 25 | 1 | 0 | 0 | 0 |
| Jan 26 | 0 | 0 | 0 | 0 |
| Feb 26 | 0 | 0 | 0 | 0 |
| Mar 26 | 0 | 0 | 0 | 1 |
| Apr 26 | 0 | 0 | 0 | 0 |
| May 26 | 0 | 0 | 0 | 0 |
| Jun 26 | 1 | 0 | 0 | 0 |
| Jul 26 | 0 | 0 | 0 | 0 |
| Aug 26 | 0 | 0 | 0 | 0 |
| Sep 26 | 0 | 0 | 0 | 0 |
| Oct 26 | 0 | 0 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Explosion1
- Overheating of Device1
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Burning Sensation1
- No Clinical Signs, Symptoms or Conditions1
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-0684-2025 · Started Nov 25, 2024 · Posted by FDA Dec 17, 2024
Reason given: Due to a manufacturing error the incorrect device serial number was potentially programmed into the holter recorder. The physical serial number label on the device might not match the serial number programmed into the device.
Cause as FDA categorized it: Process control
Z-0685-2025 · Started Nov 25, 2024 · Posted by FDA Dec 17, 2024
Reason given: Due to a manufacturing error the incorrect device serial number was potentially programmed into the holter recorder. The physical serial number label on the device might not match the serial number programmed into the device.
Cause as FDA categorized it: Process control
Z-1459-2020 · Started Jan 22, 2020
Reason given: An issue with the DL900 and DL950 Series Holter Recorder was recognized on January 1, 2020. This issue can prevent the recorder from operating normally. Braemar confirmed that beginning on January 1, 2020 if a AAA battery is inserted in the recorder and a user attempts to start it, the recorder will display Error: 602...
Cause as FDA categorized it: Software design
Z-1460-2020 · Started Jan 22, 2020
Reason given: An issue with the DL900 and DL950 Series Holter Recorder was recognized on January 1, 2020. This issue can prevent the recorder from operating normally. Braemar confirmed that beginning on January 1, 2020 if a AAA battery is inserted in the recorder and a user attempts to start it, the recorder will display Error: 602...
Cause as FDA categorized it: Software design
Z-1345-2020 · Started Jan 13, 2020
Reason given: if an AAA battery is inserted in the recorder and a user attempts to start it, or if the recorder is inserted in the docking station, the recorder will display Error: 602 and fail to function for a new patient study.
Cause as FDA categorized it: Device Design
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.