Device type · MOM
System, Hemodynamic, Implantable
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
MOM
Cardiovascular
Device class
Class III
FDA decisions since 2015
2
510(k), De Novo and PMA
Adverse event reports
2,176
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- System, Hemodynamic, Implantable
- Product code
- MOM
- Device class
- Class III
- Review panel
- Cardiovascular
- Specialty
- Cardiology
- FDA record
- Classification database entry for MOM
Latest FDA decisions
P230040 · Decided Jun 20, 2024 · 190 days after receipt
Device type: System, hemodynamic, implantable (MOM) · Cardiology
- CardioMEMs HF SystemPMA supplement
P100045-S056 · Decided Feb 18, 2022 · 232 days after receipt · Panel Track
Device type: System, hemodynamic, implantable (MOM) · Cardiology
Reports received
2,176
Received Nov 1, 2025 to Nov 19, 2026. Includes 4 of another or unstated type.
Malfunction reports
1,500
69% of reports
Injury reports
671
31% of reports
Death reports
1
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 92 | 56 | 0 | 0 |
| Jan 24 | 141 | 51 | 0 | 0 |
| Feb 24 | 95 | 53 | 1 | 0 |
| Mar 24 | 80 | 50 | 0 | 0 |
| Apr 24 | 73 | 45 | 1 | 0 |
| May 24 | 91 | 50 | 0 | 0 |
| Jun 24 | 92 | 50 | 0 | 0 |
| Jul 24 | 98 | 45 | 0 | 0 |
| Aug 24 | 80 | 59 | 2 | 0 |
| Sep 24 | 71 | 47 | 0 | 0 |
| Oct 24 | 236 | 55 | 0 | 0 |
| Nov 24 | 140 | 56 | 1 | 0 |
| Dec 24 | 93 | 60 | 0 | 0 |
| Jan 25 | 125 | 50 | 0 | 0 |
| Feb 25 | 175 | 43 | 0 | 0 |
| Mar 25 | 112 | 57 | 0 | 0 |
| Apr 25 | 252 | 50 | 0 | 0 |
| May 25 | 132 | 36 | 1 | 0 |
| Jun 25 | 143 | 62 | 1 | 0 |
| Jul 25 | 113 | 79 | 1 | 0 |
| Aug 25 | 93 | 55 | 1 | 0 |
| Sep 25 | 120 | 52 | 0 | 0 |
| Oct 25 | 98 | 60 | 4 | 0 |
| Nov 25 | 93 | 45 | 0 | 0 |
| Dec 25 | 124 | 59 | 0 | 0 |
| Jan 26 | 139 | 43 | 0 | 0 |
| Feb 26 | 113 | 57 | 0 | 0 |
| Mar 26 | 175 | 66 | 0 | 0 |
| Apr 26 | 168 | 72 | 1 | 0 |
| May 26 | 126 | 55 | 0 | 0 |
| Jun 26 | 121 | 52 | 0 | 0 |
| Jul 26 | 117 | 53 | 0 | 1 |
| Aug 26 | 113 | 60 | 0 | 2 |
| Sep 26 | 101 | 55 | 0 | 0 |
| Oct 26 | 110 | 54 | 0 | 1 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Material Frayed1,090
- Incorrect Measurement620
- Sparking246
- Application Program Problem141
- Inappropriate Waveform111
- Adverse Event Without Identified Device or Use Problem56
- Failure to Power Up53
- Failure to Read Input Signal26
- Migration26
- Intermittent Loss of Power23
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions2,080
- Dyspnea25
- Hemoptysis21
- Hypervolemia15
- Heart Failure/Congestive Heart Failure9
- Insufficient Information9
- Weight Changes9
- Great Vessel Perforation8
- Thrombosis/Thrombus7
- Kidney injury6
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-1427-2025 · Started Feb 3, 2025 · Posted by FDA Mar 21, 2025
Reason given: Due to issues associated with data migration of patient information and results to a Cloud service, the firm has received complaint associated with some patient profile information being duplicated while other patient information is missing.
Cause as FDA categorized it: Software design
Z-0987-2025 · Started Nov 6, 2024 · Posted by FDA Jan 17, 2025
Reason given: Due to a data migration from on-premises server to the cloud environment, some customers reported that they were not able to transmit readings.
Cause as FDA categorized it: Under Investigation by firm
Z-0988-2025 · Started Oct 5, 2024 · Posted by FDA Jan 17, 2025
Reason given: As a result of a cloud data migration of Patient and Hospital Electronic Systems (PES and HES) reverted to an outdated configuration causing incorrect/inaccurate pulmonary artery (PA) pressure readings.
Cause as FDA categorized it: Under Investigation by firm
Z-0989-2025 · Started Oct 5, 2024 · Posted by FDA Jan 17, 2025
Reason given: As a result of a cloud data migration of Patient and Hospital Electronic Systems (PES and HES) reverted to an outdated configuration causing incorrect/inaccurate pulmonary artery (PA) pressure readings.
Cause as FDA categorized it: Under Investigation by firm
Z-0218-2024 · Started Oct 4, 2023 · Posted by FDA Nov 3, 2023
Reason given: Potential for damaged and frayed power connector plug with repeated bending or manipulation beyond 90 degrees.
Cause as FDA categorized it: Device Design
Z-1218-2023 · Started Feb 7, 2023 · Posted by FDA Mar 3, 2023
Reason given: Radiated Radiofrequency (RF) emissions of the CardioMEMS Patient Electronics Systems (PES) Models CM1000, CM1010 and CM1100 and CardioMEMS Hospital Electronics Systems (HES) Model CM3000 exceed certain emissions limits and have the potential to cause interference with other medical devices, such as implantable...
Cause as FDA categorized it: Device Design
Z-1219-2023 · Started Feb 7, 2023 · Posted by FDA Mar 3, 2023
Reason given: Radiated Radiofrequency (RF) emissions of the CardioMEMS Patient Electronics Systems (PES) Models CM1000, CM1010 and CM1100 and CardioMEMS Hospital Electronics Systems (HES) Model CM3000 exceed certain emissions limits and have the potential to cause interference with other medical devices, such as implantable...
Cause as FDA categorized it: Device Design
Z-1220-2023 · Started Feb 7, 2023 · Posted by FDA Mar 3, 2023
Reason given: Radiated Radiofrequency (RF) emissions of the CardioMEMS Patient Electronics Systems (PES) Models CM1000, CM1010 and CM1100 and CardioMEMS Hospital Electronics Systems (HES) Model CM3000 exceed certain emissions limits and have the potential to cause interference with other medical devices, such as implantable...
Cause as FDA categorized it: Device Design
Showing the 8 newest of 13 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.