Device type · MBI
Fastener, Fixation, Nondegradable, Soft Tissue
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
MBI
Orthopedic
Device class
Class II
FDA decisions since 2015
308
510(k), De Novo and PMA
Adverse event reports
4,343
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Fastener, Fixation, Nondegradable, Soft Tissue
- Product code
- MBI
- Device class
- Class II
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3040
- Specialty
- Orthopedic Surgery
- FDA record
- Classification database entry for MBI
Latest FDA decisions
The 8 newest of 308 since 2015.
- Arthrex FiberButton510(k)
K263477 · Decided Oct 27, 2026 · 28 days after receipt
Device type: Fastener, Fixation, Nondegradable, Soft Tissue (MBI) · Orthopedic Surgery
K263348 · Decided Oct 23, 2026 · 30 days after receipt
Device type: Fastener, Fixation, Nondegradable, Soft Tissue (MBI) · Orthopedic Surgery
- TeKBrace Knotless Anchor510(k)
K263352 · Decided Oct 21, 2026 · 28 days after receipt
Device type: Fastener, Fixation, Nondegradable, Soft Tissue (MBI) · Orthopedic Surgery
K261869 · Decided Sep 17, 2026 · 105 days after receipt
Device type: Fastener, Fixation, Nondegradable, Soft Tissue (MBI) · Orthopedic Surgery
K253065 · Decided Sep 2, 2026 · 345 days after receipt
Device type: Fastener, Fixation, Nondegradable, Soft Tissue (MBI) · Orthopedic Surgery
K262809 · Decided Sep 2, 2026 · 26 days after receipt
Device type: Fastener, Fixation, Nondegradable, Soft Tissue (MBI) · Orthopedic Surgery
- Iconix Knotless T510(k)
K262677 · Decided Aug 27, 2026 · 28 days after receipt
Device type: Fastener, Fixation, Nondegradable, Soft Tissue (MBI) · Orthopedic Surgery
K262047 · Decided Aug 3, 2026 · 46 days after receipt
Device type: Fastener, Fixation, Nondegradable, Soft Tissue (MBI) · Orthopedic Surgery
Reports received
4,343
Received Nov 1, 2025 to Nov 19, 2026. Includes 13 of another or unstated type.
Malfunction reports
3,996
92% of reports
Injury reports
334
7.7% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 318 | 22 | 0 | 0 |
| Jan 24 | 277 | 17 | 0 | 0 |
| Feb 24 | 218 | 26 | 0 | 0 |
| Mar 24 | 229 | 37 | 0 | 0 |
| Apr 24 | 298 | 19 | 0 | 0 |
| May 24 | 561 | 30 | 0 | 0 |
| Jun 24 | 492 | 19 | 0 | 0 |
| Jul 24 | 195 | 36 | 0 | 0 |
| Aug 24 | 171 | 24 | 0 | 0 |
| Sep 24 | 260 | 73 | 0 | 0 |
| Oct 24 | 263 | 22 | 0 | 0 |
| Nov 24 | 186 | 23 | 0 | 0 |
| Dec 24 | 287 | 14 | 0 | 1 |
| Jan 25 | 243 | 12 | 0 | 0 |
| Feb 25 | 274 | 12 | 0 | 0 |
| Mar 25 | 259 | 16 | 0 | 0 |
| Apr 25 | 374 | 30 | 0 | 0 |
| May 25 | 295 | 8 | 0 | 0 |
| Jun 25 | 312 | 20 | 0 | 0 |
| Jul 25 | 287 | 37 | 0 | 0 |
| Aug 25 | 310 | 11 | 0 | 0 |
| Sep 25 | 328 | 21 | 0 | 0 |
| Oct 25 | 352 | 20 | 0 | 0 |
| Nov 25 | 219 | 23 | 0 | 1 |
| Dec 25 | 381 | 30 | 0 | 0 |
| Jan 26 | 290 | 33 | 0 | 0 |
| Feb 26 | 354 | 46 | 0 | 0 |
| Mar 26 | 343 | 16 | 0 | 0 |
| Apr 26 | 318 | 25 | 0 | 0 |
| May 26 | 376 | 16 | 0 | 0 |
| Jun 26 | 362 | 38 | 0 | 0 |
| Jul 26 | 384 | 22 | 0 | 0 |
| Aug 26 | 288 | 21 | 0 | 3 |
| Sep 26 | 368 | 39 | 0 | 6 |
| Oct 26 | 313 | 25 | 0 | 3 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Arthrex1,501
- Medos Sã Rl940
- Smith+Nephew747
- Arthrocare195
- Medos168
- Stryker53
- CONMED48
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Break1,386
- Device-Device Incompatibility455
- Device Dislodged or Dislocated364
- Premature Activation274
- Fracture249
- Difficult to Advance212
- Material Split, Cut or Torn183
- Detachment of Device or Device Component149
- Positioning Failure141
- Migration122
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions3,480
- Foreign Body In Patient259
- Insufficient Information233
- Perforation194
- Device Embedded In Tissue or Plaque81
- Pain45
- Unspecified Infection42
- Failure of Implant38
- Joint Laxity19
- Unspecified Tissue Injury14
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
- BIOMET ZipTight, Acute AC Joint Implant, Single Ziploop , Model/Catalog Number: 904834; Fixation System for AC Joint RepairOpen, Classified
Z-0212-2026 · Started Sep 23, 2025 · Posted by FDA Oct 17, 2025
Reason given: There is a potential for a missing slotted button assembly in one lot of ZipTight, Acute AC Joint Implant, Single Ziploop
Cause as FDA categorized it: Under Investigation by firm
- Juggerknot Mini Soft Anchors, Model/Catalog Number: 912080Open, Classified
Z-2619-2025 · Started Sep 9, 2025 · Posted by FDA Sep 23, 2025
Reason given: The outer carton box label is for the JuggerKnot 1.0mm Mini products. However, the products within the carton are JuggerKnot 1.4mm Short devices.
Cause as FDA categorized it: Under Investigation by firm
- Stryker CinchLock Flex Knotless Anchor with Inserter and Suture, REF CAT02643, Sterile EO, Rx Only MR Safe CE 0197Open, Classified
Z-1328-2025 · Started Jan 29, 2025 · Posted by FDA Mar 12, 2025
Reason given: Increase in complaints concerning "pullwire" breakage that may result in the broken pullwire remaining in the anchor implant body after deployment.
Cause as FDA categorized it: Under Investigation by firm
Z-0581-2025 · Started Oct 29, 2024 · Posted by FDA Nov 29, 2024
Reason given: Sterile barrier breach due to inadequate packaging design that could not hold the geometry of the device.
Cause as FDA categorized it: Package design/selection
Z-0518-2024 · Started Oct 25, 2023 · Posted by FDA Dec 8, 2023
Reason given: Use of a 1.4 drill in hard bone conditions caused increased insertion forces leading to subsequent failures of the inserter, when the inserters bend, it may also move off the drill hole location leading to damage of the suture construct during insertion. Suture damage may lead to suture breaks during use.
Cause as FDA categorized it: Component design/selection
- Cinchlock SS Knotless Anchor Inserter, Part Number CAT02462Open, Classified
Z-2645-2023 · Started Aug 23, 2023 · Posted by FDA Sep 28, 2023
Reason given: Expired Product distributed
Cause as FDA categorized it: Unknown/Undetermined by firm
T.A.G. MEDICAL PRODUCTS CORPORATION, LTD.
Z-2060-2023 · Started May 4, 2023 · Posted by FDA Jul 3, 2023
Reason given: Drill guides might bend when axial forces are applied by surgeon during procedure and result in potential metal shavings due to friction with the drill bit.Risk to patients includes potential of injury.
Cause as FDA categorized it: Device Design
T.A.G. MEDICAL PRODUCTS CORPORATION, LTD.
Z-2061-2023 · Started May 4, 2023 · Posted by FDA Jul 3, 2023
Reason given: Drill guides might bend when axial forces are applied by surgeon during procedure and result in potential metal shavings due to friction with the drill bit.Risk to patients includes potential of injury.
Cause as FDA categorized it: Device Design
Showing the 8 newest of 113 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.