Device type · LYJ
Stimulator, Autonomic Nerve, Implanted For Epilepsy
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
LYJ
Neurology
Device class
Class III
FDA decisions since 2015
1
510(k), De Novo and PMA
Adverse event reports
1,495
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Stimulator, Autonomic Nerve, Implanted For Epilepsy
- Product code
- LYJ
- Device class
- Class III
- Review panel
- Neurology
- Specialty
- Neurology & Neurosurgery, Neuromodulation
- FDA record
- Classification database entry for LYJ
Latest FDA decisions
- VNS THERAPY SYSTEMPMA supplement
P970003-S207 · Decided Jun 23, 2017 · 218 days after receipt · Panel Track
Device type: Stimulator, autonomic nerve, implanted for epilepsy (LYJ) · Neurology & Neurosurgery
Reports received
1,495
Received Nov 1, 2025 to Nov 19, 2026. Includes 2 of another or unstated type.
Malfunction reports
796
53% of reports
Injury reports
630
42% of reports
Death reports
67
4.5% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 95 | 96 | 5 | 0 |
| Jan 24 | 50 | 29 | 5 | 0 |
| Feb 24 | 91 | 41 | 10 | 0 |
| Mar 24 | 71 | 46 | 5 | 0 |
| Apr 24 | 62 | 60 | 5 | 0 |
| May 24 | 83 | 48 | 6 | 0 |
| Jun 24 | 94 | 47 | 9 | 0 |
| Jul 24 | 89 | 54 | 11 | 0 |
| Aug 24 | 87 | 41 | 3 | 0 |
| Sep 24 | 53 | 48 | 10 | 1 |
| Oct 24 | 69 | 61 | 3 | 0 |
| Nov 24 | 64 | 57 | 4 | 0 |
| Dec 24 | 66 | 37 | 4 | 0 |
| Jan 25 | 53 | 43 | 3 | 0 |
| Feb 25 | 68 | 48 | 9 | 0 |
| Mar 25 | 71 | 52 | 5 | 0 |
| Apr 25 | 80 | 51 | 12 | 0 |
| May 25 | 67 | 56 | 8 | 0 |
| Jun 25 | 66 | 42 | 6 | 0 |
| Jul 25 | 84 | 44 | 1 | 0 |
| Aug 25 | 73 | 44 | 7 | 0 |
| Sep 25 | 52 | 37 | 10 | 0 |
| Oct 25 | 64 | 47 | 6 | 0 |
| Nov 25 | 68 | 41 | 6 | 0 |
| Dec 25 | 76 | 47 | 8 | 0 |
| Jan 26 | 79 | 81 | 2 | 0 |
| Feb 26 | 57 | 35 | 12 | 0 |
| Mar 26 | 66 | 54 | 5 | 0 |
| Apr 26 | 71 | 58 | 3 | 0 |
| May 26 | 68 | 55 | 8 | 0 |
| Jun 26 | 60 | 47 | 4 | 0 |
| Jul 26 | 69 | 52 | 6 | 0 |
| Aug 26 | 60 | 43 | 3 | 1 |
| Sep 26 | 57 | 65 | 8 | 0 |
| Oct 26 | 65 | 52 | 2 | 1 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Adverse Event Without Identified Device or Use Problem838
- Fracture299
- High impedance197
7 more in this ranking with a free account
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions661
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available463
- Convulsion, Clonic353
7 more in this ranking with a free account
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-1067-2025 · Started Jan 3, 2025 · Posted by FDA Jan 30, 2025
Reason given: Their is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue.
Cause as FDA categorized it: Device Design
Z-1068-2025 · Started Jan 3, 2025 · Posted by FDA Jan 30, 2025
Reason given: Their is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue.
Cause as FDA categorized it: Device Design
- SenTiva, REF: 1000; SenTiva Duo, REF: 1000-D; Generators Part of the VNS Therapy System. Used for Vagus Nerve Stimulation (VNS).Open, Classified
Z-1072-2025 · Started Dec 12, 2024 · Posted by FDA Jan 31, 2025
Reason given: Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo...
Cause as FDA categorized it: Device Design
34 more recall records with a free account
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.