Device type · LOX
Catheters, Transluminal Coronary Angioplasty, Percutaneous
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
LOX
Cardiovascular
Device class
Class II
FDA decisions since 2015
53
510(k), De Novo and PMA
Adverse event reports
1,478
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Catheters, Transluminal Coronary Angioplasty, Percutaneous
- Product code
- LOX
- Device class
- Class II
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.5100
- Specialty
- Cardiology, Interventional Cardiology
- FDA record
- Classification database entry for LOX
Latest FDA decisions
The 8 newest of 53 since 2015.
K261428 · Decided Aug 14, 2026 · 106 days after receipt
Device type: Catheters, Transluminal Coronary Angioplasty, Percutaneous (LOX) · Cardiology
K261759 · Decided Jul 27, 2026 · 60 days after receipt
Device type: Catheters, Transluminal Coronary Angioplasty, Percutaneous (LOX) · Cardiology
Shanghai MicroPort Medical (Group) Co., Ltd.
K252116 · Decided Apr 7, 2026 · 274 days after receipt
Device type: Catheters, Transluminal Coronary Angioplasty, Percutaneous (LOX) · Cardiology
50 more FDA decisions with a free account
Reports received
1,478
Received Nov 1, 2025 to Nov 19, 2026. Includes 4 of another or unstated type.
Malfunction reports
1,283
87% of reports
Injury reports
176
12% of reports
Death reports
15
1.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 113 | 13 | 1 | 0 |
| Jan 24 | 134 | 12 | 0 | 0 |
| Feb 24 | 135 | 13 | 4 | 0 |
| Mar 24 | 117 | 23 | 6 | 0 |
| Apr 24 | 126 | 25 | 0 | 0 |
| May 24 | 141 | 17 | 2 | 0 |
| Jun 24 | 133 | 15 | 0 | 0 |
| Jul 24 | 176 | 18 | 4 | 0 |
| Aug 24 | 151 | 19 | 12 | 0 |
| Sep 24 | 81 | 13 | 0 | 0 |
| Oct 24 | 104 | 16 | 0 | 0 |
| Nov 24 | 134 | 20 | 1 | 0 |
| Dec 24 | 143 | 9 | 0 | 0 |
| Jan 25 | 108 | 17 | 0 | 0 |
| Feb 25 | 87 | 123 | 38 | 0 |
| Mar 25 | 113 | 17 | 3 | 0 |
| Apr 25 | 117 | 11 | 0 | 0 |
| May 25 | 92 | 224 | 27 | 0 |
| Jun 25 | 99 | 23 | 4 | 0 |
| Jul 25 | 104 | 10 | 0 | 0 |
| Aug 25 | 113 | 19 | 9 | 0 |
| Sep 25 | 135 | 10 | 3 | 0 |
| Oct 25 | 120 | 13 | 0 | 0 |
| Nov 25 | 119 | 30 | 0 | 0 |
| Dec 25 | 123 | 11 | 0 | 0 |
| Jan 26 | 105 | 8 | 3 | 0 |
| Feb 26 | 116 | 14 | 1 | 0 |
| Mar 26 | 125 | 16 | 0 | 0 |
| Apr 26 | 108 | 8 | 0 | 0 |
| May 26 | 76 | 17 | 1 | 0 |
| Jun 26 | 93 | 16 | 1 | 0 |
| Jul 26 | 108 | 14 | 3 | 0 |
| Aug 26 | 98 | 7 | 2 | 1 |
| Sep 26 | 89 | 18 | 2 | 1 |
| Oct 26 | 123 | 17 | 2 | 2 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
6 more in this ranking with a free account
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Material Rupture899
- Difficult to Advance287
- Nonstandard Device156
7 more in this ranking with a free account
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions1,322
- Foreign Body In Patient33
- Vascular Dissection27
7 more in this ranking with a free account
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-1483-2022 · Started Jun 24, 2022
Reason given: Luer has balloon diameter printed as "2.25mmx12mm" instead of the correct diameter, "2.5mmx12mm" as reflected on the product outer packaging.
Cause as FDA categorized it: Labeling Change Control
Z-1128-2020 · Started Jan 29, 2020 · Posted by FDA Feb 14, 2020
Reason given: Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
Cause as FDA categorized it: Process change control
Z-1129-2020 · Started Jan 29, 2020 · Posted by FDA Feb 14, 2020
Reason given: Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
Cause as FDA categorized it: Process change control
118 more recall records with a free account
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.