Device type · LGW
Stimulator, Spinal-Cord, Totally Implanted For Pain Relief
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
LGW
Neurology
Device class
Class III
FDA decisions since 2015
13
510(k), De Novo and PMA
Adverse event reports
31,557
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Stimulator, Spinal-Cord, Totally Implanted For Pain Relief
- Product code
- LGW
- Device class
- Class III
- Review panel
- Neurology
- Specialty
- Neurology & Neurosurgery, Neuromodulation
- FDA record
- Classification database entry for LGW
Latest FDA decisions
The 8 newest of 13 since 2015.
- Saluda Medical Evoke® SCS SystemPMA supplement
P190002-S019 · Decided Mar 20, 2024 · 180 days after receipt · Panel Track
Device type: Stimulator, spinal-cord, totally implanted for pain relief (LGW) · Neurology & Neurosurgery
- Spectra WaveWriter SCS System, WaveWriter Alpha SCS System, WaveWriter Alpha Prime SCS SystemPMA supplement
P030017-S363 · Decided Feb 5, 2024 · 251 days after receipt · Panel Track
Device type: Stimulator, spinal-cord, totally implanted for pain relief (LGW) · Neurology & Neurosurgery
- Precision, Precision Spectra, Precision Novi, Precision Montage, Precision Montage MRI, Spectra WaveWriter, WaveWriter APMA supplement
P030017-S357 · Decided Oct 5, 2023 · 281 days after receipt · Panel Track
Device type: Stimulator, spinal-cord, totally implanted for pain relief (LGW) · Neurology & Neurosurgery
- Prodigy, Proclaim XR, Proclaim Plus, and Externa Spinal Cord Stimulation (SCS) SystemsPMA supplement
P010032-S191 · Decided May 10, 2023 · 250 days after receipt · Panel Track
Device type: Stimulator, spinal-cord, totally implanted for pain relief (LGW) · Neurology & Neurosurgery
P210037 · Decided Mar 31, 2023 · 484 days after receipt
Device type: Stimulator, spinal-cord, totally implanted for pain relief (LGW) · Neurology & Neurosurgery
P010032-S189 · Decided Jan 24, 2023 · 193 days after receipt · Panel Track
Device type: Stimulator, spinal-cord, totally implanted for pain relief (LGW) · Neurology & Neurosurgery
P190002 · Decided Feb 28, 2022 · 1,133 days after receipt
Device type: Stimulator, spinal-cord, totally implanted for pain relief (LGW) · Neurology & Neurosurgery
- Restore, Itrel, Synergy, Intellis, and Vanta Spinal Cord Stimulation Systems, Pisces, Specify and Vectris Spinal Cord StPMA supplement
P840001-S469 · Decided Jan 21, 2022 · 164 days after receipt · Panel Track
Device type: Stimulator, spinal-cord, totally implanted for pain relief (LGW) · Neurology & Neurosurgery
Reports received
31,557
Received Nov 1, 2025 to Nov 19, 2026. Includes 90 of another or unstated type.
Malfunction reports
5,801
18% of reports
Injury reports
25,448
81% of reports
Death reports
218
0.7% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 907 | 1,837 | 20 | 0 |
| Jan 24 | 1,730 | 2,062 | 21 | 4 |
| Feb 24 | 2,097 | 1,960 | 24 | 0 |
| Mar 24 | 919 | 1,990 | 29 | 0 |
| Apr 24 | 613 | 2,150 | 16 | 0 |
| May 24 | 689 | 2,201 | 23 | 0 |
| Jun 24 | 639 | 1,968 | 30 | 1 |
| Jul 24 | 687 | 2,007 | 32 | 0 |
| Aug 24 | 667 | 2,062 | 32 | 0 |
| Sep 24 | 526 | 1,838 | 15 | 0 |
| Oct 24 | 556 | 2,226 | 2 | 1 |
| Nov 24 | 502 | 2,039 | 1 | 0 |
| Dec 24 | 482 | 2,234 | 2 | 0 |
| Jan 25 | 627 | 1,892 | 0 | 0 |
| Feb 25 | 526 | 1,943 | 2 | 0 |
| Mar 25 | 539 | 2,127 | 1 | 0 |
| Apr 25 | 606 | 2,190 | 3 | 0 |
| May 25 | 493 | 2,188 | 2 | 0 |
| Jun 25 | 513 | 2,158 | 1 | 0 |
| Jul 25 | 545 | 2,128 | 2 | 1 |
| Aug 25 | 485 | 2,019 | 1 | 0 |
| Sep 25 | 554 | 1,889 | 3 | 0 |
| Oct 25 | 572 | 2,371 | 6 | 0 |
| Nov 25 | 445 | 1,908 | 6 | 1 |
| Dec 25 | 602 | 2,095 | 1 | 0 |
| Jan 26 | 462 | 1,773 | 3 | 0 |
| Feb 26 | 503 | 1,641 | 3 | 0 |
| Mar 26 | 538 | 2,189 | 2 | 0 |
| Apr 26 | 559 | 2,167 | 9 | 0 |
| May 26 | 570 | 2,060 | 29 | 2 |
| Jun 26 | 592 | 2,176 | 30 | 1 |
| Jul 26 | 519 | 2,202 | 28 | 4 |
| Aug 26 | 357 | 2,235 | 26 | 26 |
| Sep 26 | 313 | 2,426 | 49 | 38 |
| Oct 26 | 341 | 2,576 | 32 | 18 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Adverse Event Without Identified Device or Use Problem7,667
- High impedance4,216
- Charging Problem4,007
- Migration3,147
- Insufficient Device Problem Information2,742
- Unexpected Therapeutic Results2,637
- Battery Problem1,731
- Inappropriate/Inadequate Shock/Stimulation1,670
- Wireless Communication Problem1,642
- Therapeutic or Diagnostic Output Failure1,594
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions10,551
- Inadequate Pain Relief7,963
- Pain5,415
- Insufficient Information2,462
- Implant Pain1,936
- Unspecified Infection1,331
- Discomfort1,000
- Electric Shock887
- Burning Sensation740
- Inappropriate Device Stimulation of Tissue610
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Boston Scientific Neuromodulation Corporation
Z-2879-2026 · Started Jun 17, 2026 · Posted by FDA Aug 5, 2026
Reason given: Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, including lead explant and replacement.
Cause as FDA categorized it: Device Design
- Boston Scientific Model SC-4318, Clik X Anchor REF: M365SC43180Open, Classified
Boston Scientific Neuromodulation Corporation
Z-2404-2026 · Started Apr 29, 2026 · Posted by FDA Jun 11, 2026
Reason given: Sterile anchors my lack sterility assurance.
Cause as FDA categorized it: Packaging process control
Z-1043-2026 · Started Dec 3, 2025 · Posted by FDA Jan 9, 2026
Reason given: Complaints received that Vanta A71200 CP App does not function as intended during use. There is a potential for prolonged or postponed surgical procedure.
Cause as FDA categorized it: Software design
Z-0996-2026 · Started Nov 20, 2025 · Posted by FDA Dec 23, 2025
Reason given: Software issue where a Device Reset message displayed on the app was unable to be cleared. In rare cases, this resulted in an inability to resume therapy, and the patient experienced a recurrence of underlying pain symptoms. Resolution would require surgical replacement of the INS.
Cause as FDA categorized it: Software design
- Eterna SCS IPG (Implantable Pulse Generator), Model Number 32400Open, Classified
Z-0458-2026 · Started Oct 6, 2025 · Posted by FDA Nov 7, 2025
Reason given: There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse Generators (IPGs), Model 62400 . These devicess may experience a loss of communication with the Clinician Programmer (CP) and/or the...
Cause as FDA categorized it: Process control
Z-1543-2025 · Started Mar 4, 2025 · Posted by FDA Apr 8, 2025
Reason given: There is a software issue that can permanently disable communication with an implantable neurostimulator (INS) when a specific intraoperative programming sequence is used.
Cause as FDA categorized it: Under Investigation by firm
Z-0372-2025 · Started Sep 18, 2024 · Posted by FDA Nov 13, 2024
Reason given: The Wireless Rechargers in the kits may be unable to enter a recharging session upon first use.
Cause as FDA categorized it: Under Investigation by firm
- Vanta Implantable Neurostimulator (INS), Model 977006Open, Classified
Z-0132-2025 · Started Aug 21, 2024 · Posted by FDA Oct 17, 2024
Reason given: Model 977006 Vanta INS batterty life has depleted faster than HCPs or patients expected leading to outpatient reprogramming or earlier than anticipated surgical replacements.
Cause as FDA categorized it: Under Investigation by firm
Showing the 8 newest of 96 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.