Device type · KOC
Accessories, Blood Circuit, Hemodialysis
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
KOC
Gastroenterology & Urology
Device class
Class II
FDA decisions since 2015
9
510(k), De Novo and PMA
Adverse event reports
34
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Accessories, Blood Circuit, Hemodialysis
- Product code
- KOC
- Device class
- Class II
- Review panel
- Gastroenterology & Urology
- Regulation
- 21 CFR 876.5820
- Specialty
- GI, Urology & Nephrology, Nephrology
- FDA record
- Classification database entry for KOC
Latest FDA decisions
The 8 newest of 9 since 2015.
K242176 · Decided May 15, 2025 · 295 days after receipt
Device type: Accessories, Blood Circuit, Hemodialysis (KOC) · GI, Urology & Nephrology
K242479 · Decided May 15, 2025 · 267 days after receipt
Device type: Accessories, Blood Circuit, Hemodialysis (KOC) · GI, Urology & Nephrology
K231589 · Decided Jul 28, 2023 · 57 days after receipt
Device type: Accessories, Blood Circuit, Hemodialysis (KOC) · GI, Urology & Nephrology
K230514 · Decided Jun 16, 2023 · 112 days after receipt
Device type: Accessories, Blood Circuit, Hemodialysis (KOC) · GI, Urology & Nephrology
Fresenius Medical Care Renal Therapies Group, LLC
K201207 · Decided Sep 18, 2020 · 136 days after receipt
Device type: Accessories, Blood Circuit, Hemodialysis (KOC) · GI, Urology & Nephrology
Fresenius Medical Care Renal Therapies Group, LLC
K172238 · Decided Dec 14, 2017 · 141 days after receipt
Device type: Accessories, Blood Circuit, Hemodialysis (KOC) · GI, Urology & Nephrology
Fresenius Medical Care Renal Therapies Group, LLC
K160749 · Decided Apr 15, 2016 · 28 days after receipt
Device type: Accessories, Blood Circuit, Hemodialysis (KOC) · GI, Urology & Nephrology
Fresenius Medical Care North America
K152953 · Decided Nov 5, 2015 · 29 days after receipt
Device type: Accessories, Blood Circuit, Hemodialysis (KOC) · GI, Urology & Nephrology
Reports received
34
Received Nov 1, 2025 to Nov 19, 2026
Malfunction reports
34
100% of reports
Injury reports
0
0.0% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 0 | 0 | 0 | 0 |
| Jan 24 | 3 | 0 | 0 | 0 |
| Feb 24 | 1 | 0 | 0 | 0 |
| Mar 24 | 0 | 0 | 0 | 0 |
| Apr 24 | 1 | 0 | 0 | 0 |
| May 24 | 5 | 0 | 0 | 0 |
| Jun 24 | 2 | 0 | 0 | 0 |
| Jul 24 | 1 | 0 | 0 | 0 |
| Aug 24 | 1 | 0 | 0 | 0 |
| Sep 24 | 0 | 0 | 0 | 0 |
| Oct 24 | 0 | 0 | 0 | 0 |
| Nov 24 | 1 | 0 | 0 | 0 |
| Dec 24 | 3 | 0 | 0 | 0 |
| Jan 25 | 0 | 0 | 0 | 0 |
| Feb 25 | 2 | 0 | 0 | 0 |
| Mar 25 | 3 | 0 | 0 | 0 |
| Apr 25 | 1 | 0 | 1 | 0 |
| May 25 | 1 | 0 | 1 | 0 |
| Jun 25 | 2 | 0 | 0 | 0 |
| Jul 25 | 2 | 1 | 0 | 0 |
| Aug 25 | 14 | 0 | 0 | 0 |
| Sep 25 | 1 | 0 | 0 | 0 |
| Oct 25 | 13 | 0 | 0 | 0 |
| Nov 25 | 5 | 0 | 0 | 0 |
| Dec 25 | 3 | 0 | 0 | 0 |
| Jan 26 | 4 | 0 | 0 | 0 |
| Feb 26 | 4 | 0 | 0 | 0 |
| Mar 26 | 6 | 0 | 0 | 0 |
| Apr 26 | 1 | 0 | 0 | 0 |
| May 26 | 2 | 0 | 0 | 0 |
| Jun 26 | 7 | 0 | 0 | 0 |
| Jul 26 | 1 | 0 | 0 | 0 |
| Aug 26 | 1 | 0 | 0 | 0 |
| Sep 26 | 0 | 0 | 0 | 0 |
| Oct 26 | 0 | 0 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Fluid/Blood Leak32
- Crack5
- Fire2
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Hemorrhage/Blood Loss/Bleeding27
- No Clinical Signs, Symptoms or Conditions7
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Fresenius Medical Care Renal Therapies Group, LLC
Z-3084-2026 · Started Aug 10, 2026 · Posted by FDA Sep 2, 2026
Reason given: Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.
Cause as FDA categorized it: Component change control
- 5008X HD/HDF Standard Blood Tubing Set Part Number 03-5300-3Open, Classified
Fresenius Medical Care Renal Therapies Group, LLC
Z-3085-2026 · Started Aug 10, 2026 · Posted by FDA Sep 2, 2026
Reason given: Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.
Cause as FDA categorized it: Component change control
- 5008X HD/HDF with CLiC Blood Tubing Set Part number 03-5300-3COpen, Classified
Fresenius Medical Care Renal Therapies Group, LLC
Z-3086-2026 · Started Aug 10, 2026 · Posted by FDA Sep 2, 2026
Reason given: Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.
Cause as FDA categorized it: Component change control
- 5008X HD/HDF with Twister Blood Tubing Set Part Number 03-5350-8Open, Classified
Fresenius Medical Care Renal Therapies Group, LLC
Z-3087-2026 · Started Aug 10, 2026 · Posted by FDA Sep 2, 2026
Reason given: Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.
Cause as FDA categorized it: Component change control
Z-0759-2019 · Started Sep 6, 2018
Reason given: PRISMAFLO II and PRISMAFLO II S Heating Profile only and not the Control Unit, damage may result in a hotspot on the Heating Profile, may cause low level skin burn
Cause as FDA categorized it: Device Design
Z-0760-2019 · Started Sep 6, 2018
Reason given: PRISMAFLO II and PRISMAFLO II S Heating Profile only and not the Control Unit, damage may result in a hotspot on the Heating Profile, may cause low level skin burn
Cause as FDA categorized it: Device Design
Z-0761-2019 · Started Sep 6, 2018
Reason given: PRISMAFLO II and PRISMAFLO II S Heating Profile only and not the Control Unit, damage may result in a hotspot on the Heating Profile, may cause low level skin burn
Cause as FDA categorized it: Device Design
Fresenius Medical Care Renal Therapies Group, LLC
Z-2044-2016 · Started May 24, 2016
Reason given: Use of the Blood Pressure Graph located on the Crit-Line screen of the 2008T Hemodialysis Machine may result in a frozen screen, possibly causing the blood pump to stop.
Cause as FDA categorized it: Software design
Showing the 8 newest of 17 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.