Device type · KCT
Sterilization Wrap Containers, Trays, Cassettes & Other Accessories
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
KCT
General Hospital
Device class
Class II
FDA decisions since 2015
151
510(k), De Novo and PMA
Adverse event reports
27
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Sterilization Wrap Containers, Trays, Cassettes & Other Accessories
- Product code
- KCT
- Device class
- Class II
- Review panel
- General Hospital
- Regulation
- 21 CFR 880.6850
- Specialty
- Hospital Medicine & Nursing
- FDA record
- Classification database entry for KCT
Latest FDA decisions
The 8 newest of 151 since 2015.
- S.I.N. Instruments Kits510(k)
K262084 · Decided Oct 22, 2026 · 122 days after receipt
Device type: Sterilization Wrap Containers, Trays, Cassettes & Other Accessories (KCT) · Hospital Medicine & Nursing
Alliedstar Medical Equipment Co., Ltd.
K262042 · Decided Oct 14, 2026 · 119 days after receipt
Device type: Sterilization Wrap Containers, Trays, Cassettes & Other Accessories (KCT) · Hospital Medicine & Nursing
- LiteSet Trays510(k)
K253988 · Decided Sep 8, 2026 · 270 days after receipt
Device type: Sterilization Wrap Containers, Trays, Cassettes & Other Accessories (KCT) · Hospital Medicine & Nursing
K261935 · Decided Sep 4, 2026 · 87 days after receipt
Device type: Sterilization Wrap Containers, Trays, Cassettes & Other Accessories (KCT) · Hospital Medicine & Nursing
K253187 · Decided Jun 18, 2026 · 265 days after receipt
Device type: Sterilization Wrap Containers, Trays, Cassettes & Other Accessories (KCT) · Hospital Medicine & Nursing
K252854 · Decided Jun 3, 2026 · 268 days after receipt
Device type: Sterilization Wrap Containers, Trays, Cassettes & Other Accessories (KCT) · Hospital Medicine & Nursing
K260726 · Decided Jun 3, 2026 · 90 days after receipt
Device type: Sterilization Wrap Containers, Trays, Cassettes & Other Accessories (KCT) · Hospital Medicine & Nursing
K253791 · Decided Feb 26, 2026 · 90 days after receipt
Device type: Sterilization Wrap Containers, Trays, Cassettes & Other Accessories (KCT) · Hospital Medicine & Nursing
Reports received
27
Received Nov 1, 2025 to Nov 19, 2026
Malfunction reports
25
93% of reports
Injury reports
2
7.4% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 11 | 0 | 0 | 0 |
| Jan 24 | 1 | 0 | 0 | 0 |
| Feb 24 | 2 | 0 | 0 | 0 |
| Mar 24 | 0 | 1 | 0 | 0 |
| Apr 24 | 0 | 0 | 0 | 0 |
| May 24 | 1 | 0 | 0 | 0 |
| Jun 24 | 3 | 1 | 0 | 0 |
| Jul 24 | 1 | 0 | 0 | 0 |
| Aug 24 | 0 | 0 | 0 | 0 |
| Sep 24 | 2 | 0 | 0 | 0 |
| Oct 24 | 4 | 0 | 0 | 0 |
| Nov 24 | 1 | 0 | 0 | 0 |
| Dec 24 | 0 | 1 | 0 | 0 |
| Jan 25 | 8 | 0 | 0 | 0 |
| Feb 25 | 1 | 0 | 0 | 0 |
| Mar 25 | 2 | 2 | 0 | 0 |
| Apr 25 | 3 | 0 | 0 | 0 |
| May 25 | 4 | 0 | 0 | 0 |
| Jun 25 | 15 | 0 | 0 | 0 |
| Jul 25 | 0 | 0 | 0 | 0 |
| Aug 25 | 6 | 0 | 0 | 0 |
| Sep 25 | 0 | 0 | 0 | 0 |
| Oct 25 | 6 | 0 | 0 | 0 |
| Nov 25 | 2 | 0 | 0 | 0 |
| Dec 25 | 0 | 1 | 0 | 0 |
| Jan 26 | 3 | 0 | 0 | 0 |
| Feb 26 | 4 | 0 | 0 | 0 |
| Mar 26 | 0 | 0 | 0 | 0 |
| Apr 26 | 2 | 0 | 0 | 0 |
| May 26 | 2 | 0 | 0 | 0 |
| Jun 26 | 5 | 0 | 0 | 0 |
| Jul 26 | 2 | 0 | 0 | 0 |
| Aug 26 | 0 | 1 | 0 | 0 |
| Sep 26 | 4 | 0 | 0 | 0 |
| Oct 26 | 1 | 0 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Aesculap13
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Material Puncture/Hole12
- Extra Components3
- Material Discolored2
- Material Split, Cut or Torn2
- Adverse Event Without Identified Device or Use Problem1
- Device-Device Incompatibility1
- Material Rupture1
- No Apparent Adverse Event1
- Positioning Problem1
- Tear, Rip or Hole in Device Packaging1
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions25
- Ambulation or Postural Difficulties1
- Discomfort1
- Laceration(s)1
- Swelling/ Edema1
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
- Jewel Precision Reusable Rigid Sterilization Container System. Model Number: JP-24-8.Open, Classified
Jewel Precision Sheet Metal & Machining Co, Inc.
Z-2033-2025 · Started May 23, 2025 · Posted by FDA Jun 26, 2025
Reason given: The 8-inch model of the Jewel Precision Reusable Rigid Sterilization Container System (i.e., model number: JP-24-8) was not cleared by the FDA. The letter to file justifying the change in size was insufficient.
Cause as FDA categorized it: No Marketing Application
- Jewel Precision Reusable Rigid Sterilization Container System. Model/Catalog Number: JP-24-6.Open, Classified
Jewel Precision Sheet Metal & Machining Co, Inc.
Z-2026-2025 · Started May 22, 2025 · Posted by FDA Jun 25, 2025
Reason given: The Instructions for Use do not match the 510(k) FDA cleared indications for use or sterility shelf life.
Cause as FDA categorized it: No Marketing Application
Implant Direct Sybron Manufacturing LLC
Z-2172-2025 · Started May 1, 2025 · Posted by FDA Jul 25, 2025
Reason given: The Drill stop kit contains incorrect components.
Cause as FDA categorized it: Under Investigation by firm
Z-1991-2023 · Started Apr 6, 2023 · Posted by FDA Jun 21, 2023
Reason given: Metal sterilization trays used to encase and protect medical devices during sterilization have a new 510(k) with instructions for use that is for the first time specific to the tray models, that also includes information regarding the sterilization modality.
Cause as FDA categorized it: No Marketing Application
- KARL STORZ Radel Sterilization Trays, Models: 27717A, 27717B, 39301A, 39301C, 39231XA, 39301C1, 39311A, 39317AOpen, Classified
Z-1981-2023 · Started Apr 5, 2023 · Posted by FDA Jun 21, 2023
Reason given: Plastic sterilization trays used to encase and protect medical devices during sterilization have a new 510(k) with instructions for use that is for the first time specific to the tray models, that also includes information regarding the sterilization modality.
Cause as FDA categorized it: No Marketing Application
Z-1236-2023 · Started Jan 30, 2023 · Posted by FDA Mar 9, 2023
Reason given: Product did not meet shelf-life testing requirements resulting in a breach of the sterility of the Genesis container holding surgical equipment.
Cause as FDA categorized it: Packaging change control
- CareFusion V.Mueller GENESIS MINI CONTAINER 10-1/2 X 7-1/2" (26.7X19.0CM) 3" (7.6CM) DEEP STERRAD COMPATIBLE, REF CD0-3STOpen, Classified
Z-0203-2023 · Started Sep 26, 2022 · Posted by FDA Nov 11, 2022
Reason given: Genesis STERRAD Reusable Rigid Sterilization Containers have not consistently met the requirements for aerosol challenge testing.
Cause as FDA categorized it: Under Investigation by firm
- CareFusion V.Mueller GENESIS QTR-LENGTH CONTAINER 12 X 8" (30.3X20.3CM) 4" (10.2CM) DEEP STERRAD COMPATIBLE, REF CD0-4STOpen, Classified
Z-0204-2023 · Started Sep 26, 2022 · Posted by FDA Nov 11, 2022
Reason given: Genesis STERRAD Reusable Rigid Sterilization Containers have not consistently met the requirements for aerosol challenge testing.
Cause as FDA categorized it: Under Investigation by firm
Showing the 8 newest of 38 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.