Device type · JQC
Centrifuges (Micro, Ultra, Refrigerated) For Clinical Use
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
JQC
Clinical Chemistry
Device class
Class I
FDA decisions since 2015
1
510(k), De Novo and PMA
Adverse event reports
33
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Centrifuges (Micro, Ultra, Refrigerated) For Clinical Use
- Product code
- JQC
- Device class
- Class I
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.2050
- Specialty
- Laboratory Medicine & Pathology
- FDA record
- Classification database entry for JQC
Latest FDA decisions
K183205 · Decided May 30, 2019 · 192 days after receipt
Device type: Centrifuges (Micro, Ultra, Refrigerated) For Clinical Use (JQC) · Laboratory Medicine & Pathology
Reports received
33
Received Nov 1, 2025 to Nov 19, 2026
Malfunction reports
33
100% of reports
Injury reports
0
0.0% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 0 | 0 | 0 | 0 |
| Jan 24 | 0 | 0 | 0 | 0 |
| Feb 24 | 0 | 1 | 0 | 0 |
| Mar 24 | 0 | 0 | 0 | 0 |
| Apr 24 | 0 | 1 | 0 | 0 |
| May 24 | 0 | 0 | 0 | 0 |
| Jun 24 | 0 | 0 | 0 | 0 |
| Jul 24 | 0 | 0 | 0 | 0 |
| Aug 24 | 0 | 0 | 0 | 0 |
| Sep 24 | 0 | 0 | 0 | 0 |
| Oct 24 | 0 | 0 | 0 | 0 |
| Nov 24 | 0 | 0 | 0 | 0 |
| Dec 24 | 0 | 0 | 0 | 0 |
| Jan 25 | 1 | 0 | 0 | 1 |
| Feb 25 | 1 | 0 | 0 | 0 |
| Mar 25 | 0 | 0 | 0 | 0 |
| Apr 25 | 0 | 0 | 0 | 0 |
| May 25 | 1 | 0 | 0 | 0 |
| Jun 25 | 1 | 0 | 0 | 0 |
| Jul 25 | 0 | 0 | 0 | 0 |
| Aug 25 | 0 | 0 | 0 | 0 |
| Sep 25 | 0 | 0 | 0 | 0 |
| Oct 25 | 0 | 0 | 0 | 0 |
| Nov 25 | 0 | 0 | 0 | 0 |
| Dec 25 | 1 | 0 | 0 | 0 |
| Jan 26 | 0 | 0 | 0 | 0 |
| Feb 26 | 2 | 0 | 0 | 0 |
| Mar 26 | 1 | 0 | 0 | 0 |
| Apr 26 | 1 | 0 | 0 | 0 |
| May 26 | 0 | 0 | 0 | 0 |
| Jun 26 | 1 | 0 | 0 | 0 |
| Jul 26 | 1 | 0 | 0 | 0 |
| Aug 26 | 1 | 0 | 0 | 0 |
| Sep 26 | 22 | 0 | 0 | 0 |
| Oct 26 | 3 | 0 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
1 more in this ranking with a free account
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Break7
- Electrical /Electronic Property Problem6
- Detachment of Device or Device Component4
7 more in this ranking with a free account
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions30
- Insufficient Information3
- Exposure to Body Fluids1
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-0201-2027 · Started Sep 21, 2026 · Posted by FDA Oct 21, 2026
Reason given: Optima XE XPN ultracentrifuge rotors may contain a microstructure defect that can cause rotors to break into two or more pieces (bifurcate) during a centrifugation run, resulting in destruction of samples being processed. When used for in vitro diagnostic (IVD) sample preparation, rotor breakage may delay patient test...
Cause as FDA categorized it: Nonconforming Material/Component
Z-0202-2027 · Started Sep 21, 2026 · Posted by FDA Oct 21, 2026
Reason given: Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use. Rotors that break into multiple pieces during use may result in significant damage to materials (i.e., samples) being processed in the centrifuge and to the centrifuge instrument.
Cause as FDA categorized it: Nonconforming Material/Component
Z-1048-2019 · Started Feb 6, 2019
Reason given: The clear, plastic, polycarbonate hematocrit rotor may break apart during use and be ejected through the plastic housing, posing a safety hazard to the technician and potentially damaging the surrounding area.
Cause as FDA categorized it: Device Design
15 more recall records with a free account
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.