Device type · JOY
Device, Automated Cell-Locating
Last checked: Nov 20, 2026Source: openFDAMethods & limitations
Product code
JOY
Hematology
Device class
Class II
FDA decisions since 2015
7
510(k), De Novo and PMA
Adverse event reports
3
Received Nov 1, 2025 to Nov 20, 2026
How FDA classifies it
- Device type
- Device, Automated Cell-Locating
- Product code
- JOY
- Device class
- Class II
- Review panel
- Hematology
- Regulation
- 21 CFR 864.5260
- Specialty
- Laboratory Medicine & Pathology
- FDA record
- Classification database entry for JOY
Latest FDA decisions
K243144 · Decided Jun 27, 2025 · 270 days after receipt
Device type: Device, Automated Cell-Locating (JOY) · Laboratory Medicine & Pathology
- AI100 with Shonit510(k)
Sigtuple Technologies Pvt. , Ltd.
K221309 · Decided Sep 19, 2023 · 502 days after receipt
Device type: Device, Automated Cell-Locating (JOY) · Laboratory Medicine & Pathology
K220013 · Decided May 3, 2022 · 119 days after receipt
Device type: Device, Automated Cell-Locating (JOY) · Laboratory Medicine & Pathology
4 more FDA decisions with a free account
Reports received
3
Received Nov 1, 2025 to Nov 20, 2026
Malfunction reports
3
100% of reports
Injury reports
0
0.0% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 0 | 0 | 0 | 0 |
| Jan 24 | 0 | 0 | 0 | 0 |
| Feb 24 | 0 | 2 | 0 | 0 |
| Mar 24 | 0 | 1 | 0 | 0 |
| Apr 24 | 0 | 1 | 0 | 0 |
| May 24 | 0 | 0 | 0 | 0 |
| Jun 24 | 0 | 0 | 0 | 0 |
| Jul 24 | 0 | 0 | 0 | 0 |
| Aug 24 | 0 | 0 | 0 | 0 |
| Sep 24 | 0 | 0 | 0 | 0 |
| Oct 24 | 0 | 0 | 0 | 0 |
| Nov 24 | 1 | 0 | 0 | 0 |
| Dec 24 | 0 | 0 | 0 | 0 |
| Jan 25 | 0 | 0 | 0 | 0 |
| Feb 25 | 0 | 0 | 0 | 0 |
| Mar 25 | 0 | 0 | 0 | 0 |
| Apr 25 | 0 | 0 | 0 | 0 |
| May 25 | 0 | 0 | 0 | 0 |
| Jun 25 | 0 | 0 | 0 | 0 |
| Jul 25 | 0 | 0 | 0 | 0 |
| Aug 25 | 0 | 0 | 0 | 0 |
| Sep 25 | 0 | 0 | 0 | 0 |
| Oct 25 | 1 | 0 | 0 | 0 |
| Nov 25 | 0 | 0 | 0 | 0 |
| Dec 25 | 0 | 0 | 0 | 0 |
| Jan 26 | 0 | 0 | 0 | 0 |
| Feb 26 | 0 | 0 | 0 | 0 |
| Mar 26 | 0 | 0 | 0 | 0 |
| Apr 26 | 0 | 0 | 0 | 0 |
| May 26 | 0 | 0 | 0 | 0 |
| Jun 26 | 1 | 0 | 0 | 0 |
| Jul 26 | 2 | 0 | 0 | 0 |
| Aug 26 | 0 | 0 | 0 | 0 |
| Sep 26 | 0 | 0 | 0 | 0 |
| Oct 26 | 0 | 0 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- False Negative Result2
- Low Test Results1
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions3
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-0542-2026 · Started Oct 8, 2025 · Posted by FDA Nov 18, 2025
Reason given: Automated cell-locating device barcode reader may read the barcode of the previously processed slide resulting in a misattribution of diagnostic results.
Cause as FDA categorized it: Software design
Z-2557-2018 · Started May 7, 2018
Reason given: Discrepant results have been reported in patients with severe microcytic anemia (e.g.iron deficiency, thalassemia) and the following parameters are affected: RBC (red blood cell count), HGB (hemoglobin concentration), MCH (mean corpuscular hemoglobin), HCT (hematocrit), and MCV (mean corpuscular volume).
Cause as FDA categorized it: Software design
Z-2184-2018 · Started Dec 4, 2017
Reason given: A software malfunction was found where WBC, RBC and PLT comments added after a slide is signed, are not sent to the LIS. This can only occur where customers process multiple slides per blood sample.
Cause as FDA categorized it: Software design
1 more recall records with a free account
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.