MedTech Resources

Device type · IZL

System, X-Ray, Mobile

FDA groups similar devices under a three-letter product code. This page gathers what FDA's public data says about IZL: how it is classified, what has been cleared or approved, and what has been reported and recalled.

Last checked: Nov 19, 2026Source: openFDAMethods & limitations

Product code

IZL

Radiology

Device class

Class II

FDA decisions since 2015

102

510(k), De Novo and PMA

Adverse event reports

31

Received Nov 1, 2025 to Nov 19, 2026

How FDA classifies it

Device type
System, X-Ray, Mobile
Product code
IZL
Device class
Class II
Review panel
Radiology
Regulation
21 CFR 892.1720

Latest FDA decisions

The 8 newest of 102 since 2015.

All decisions for IZL

Reports received

31

Received Nov 1, 2025 to Nov 19, 2026. Includes 1 of another or unstated type.

Malfunction reports

24

77% of reports

Injury reports

6

19% of reports

Death reports

0

0.0% of reports

Adverse event reports received per month

By the date FDA received each report, stacked by event type. The newest month is incomplete.

Open in the MAUDE Monitor
  • Malfunction
  • Injury
  • Death
  • Other
View as table
MonthMalfunctionInjuryDeathOther
Dec 233200
Jan 244000
Feb 243000
Mar 243100
Apr 241100
May 244100
Jun 243000
Jul 245000
Aug 243000
Sep 241000
Oct 242103
Nov 243000
Dec 240000
Jan 252100
Feb 250100
Mar 252100
Apr 253000
May 252001
Jun 250100
Jul 251101
Aug 250001
Sep 250100
Oct 251100
Nov 254100
Dec 251000
Jan 262300
Feb 262000
Mar 261000
Apr 260000
May 263100
Jun 264000
Jul 260000
Aug 263000
Sep 262101
Oct 262000
Nov 260000

Manufacturers named most often

Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.

Device problems as coded

Last 12 months. FDA's terms. A report can carry several.

  • Smoking4
  • Unintended System Motion4
  • Sharp Edges3
  • Break2
  • Component Missing2
  • Detachment of Device or Device Component2
  • Insufficient Device Problem Information2
  • Key or Button Unresponsive/not Working2
  • Mechanical Jam2
  • Appropriate Device Problem Term/Code Not Available1

Patient problems as coded

Last 12 months. FDA's terms. A report can carry several.

  • No Clinical Signs, Symptoms or Conditions21
  • Aspiration/Inhalation3
  • Abrasion1
  • Cardiac Arrest1
  • Ecchymosis1
  • Electric Shock1
  • Insufficient Information1
  • Laceration(s)1
  • Pain1
  • Swelling/ Edema1

Recent recalls

From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.

  • Model Number: 712010.Open, Classified

    Philips North America

    Z-0288-2027 · Started Sep 18, 2026 · Posted by FDA Oct 26, 2026

    Reason given: Potential for the device to fail immediately after an exposure, causing the system to remain stationary.

    Cause as FDA categorized it: Software design

  • GE Medical Systems, LLC

    Z-2726-2026 · Started Jun 5, 2026 · Posted by FDA Jul 13, 2026

    Reason given: GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work,...

    Cause as FDA categorized it: Process control

  • GE Medical Systems, LLC

    Z-2727-2026 · Started Jun 5, 2026 · Posted by FDA Jul 13, 2026

    Reason given: GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work,...

    Cause as FDA categorized it: Process control

  • GE Medical Systems, LLC

    Z-2728-2026 · Started Jun 5, 2026 · Posted by FDA Jul 13, 2026

    Reason given: GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work,...

    Cause as FDA categorized it: Process control

  • Micro-X Ltd.

    Z-1918-2026 · Started Mar 27, 2026 · Posted by FDA Apr 22, 2026

    Reason given: X-Ray generator may malfunction resulting in the X-Ray being inoperable.

    Cause as FDA categorized it: Component design/selection

  • NeuroLogica Corporation

    Z-0480-2026 · Started Oct 7, 2025 · Posted by FDA Nov 7, 2025

    Reason given: The four (4) M5 bolts that secure the Tube Head Unit to the end of the arm may loosen or fracture due to metal fatigue.

    Cause as FDA categorized it: Device Design

  • SEDECAL SA

    Z-0242-2026 · Started Jun 10, 2025 · Posted by FDA Oct 23, 2025

    Reason given: Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

    Cause as FDA categorized it: Device Design

  • SEDECAL SA

    Z-0243-2026 · Started Jun 10, 2025 · Posted by FDA Oct 23, 2025

    Reason given: Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

    Cause as FDA categorized it: Device Design

Showing the 8 newest of 171 recall records in FDA's data.

Back to all tools

For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.