Device type · IYN
System, Imaging, Pulsed Doppler, Ultrasonic
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
IYN
Radiology
Device class
Class II
FDA decisions since 2015
635
510(k), De Novo and PMA
Adverse event reports
587
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- System, Imaging, Pulsed Doppler, Ultrasonic
- Product code
- IYN
- Device class
- Class II
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1550
- Specialty
- Radiology & Imaging
- FDA record
- Classification database entry for IYN
Latest FDA decisions
The 8 newest of 635 since 2015.
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
K261585 · Decided Oct 16, 2026 · 156 days after receipt
Device type: System, Imaging, Pulsed Doppler, Ultrasonic (IYN) · Radiology & Imaging
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
K261617 · Decided Oct 8, 2026 · 146 days after receipt
Device type: System, Imaging, Pulsed Doppler, Ultrasonic (IYN) · Radiology & Imaging
K263150 · Decided Oct 8, 2026 · 30 days after receipt
Device type: System, Imaging, Pulsed Doppler, Ultrasonic (IYN) · Radiology & Imaging
GE Medical Systems Ultrasound and Primary care Diagnostics, LLC
K261933 · Decided Oct 2, 2026 · 115 days after receipt
Device type: System, Imaging, Pulsed Doppler, Ultrasonic (IYN) · Radiology & Imaging
- EPIQ Inspira Series Diagnostic Ultrasound System; Affiniti Series Diagnostic Ultrasound System510(k)
K262548 · Decided Sep 25, 2026 · 64 days after receipt
Device type: System, Imaging, Pulsed Doppler, Ultrasonic (IYN) · Radiology & Imaging
K263182 · Decided Sep 25, 2026 · 14 days after receipt
Device type: System, Imaging, Pulsed Doppler, Ultrasonic (IYN) · Radiology & Imaging
GE Medical Systems Ultrasound and Primary Care Diagnostics
K261313 · Decided Sep 2, 2026 · 134 days after receipt
Device type: System, Imaging, Pulsed Doppler, Ultrasonic (IYN) · Radiology & Imaging
K254027 · Decided Aug 24, 2026 · 251 days after receipt
Device type: System, Imaging, Pulsed Doppler, Ultrasonic (IYN) · Radiology & Imaging
Reports received
587
Received Nov 1, 2025 to Nov 19, 2026. Includes 2 of another or unstated type.
Malfunction reports
519
88% of reports
Injury reports
60
10% of reports
Death reports
6
1.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 13 | 5 | 0 | 0 |
| Jan 24 | 17 | 0 | 0 | 0 |
| Feb 24 | 21 | 2 | 0 | 0 |
| Mar 24 | 19 | 3 | 1 | 0 |
| Apr 24 | 20 | 0 | 0 | 0 |
| May 24 | 17 | 3 | 0 | 0 |
| Jun 24 | 17 | 1 | 1 | 0 |
| Jul 24 | 38 | 5 | 0 | 0 |
| Aug 24 | 16 | 0 | 0 | 0 |
| Sep 24 | 18 | 3 | 0 | 0 |
| Oct 24 | 18 | 4 | 1 | 0 |
| Nov 24 | 11 | 1 | 0 | 1 |
| Dec 24 | 26 | 2 | 0 | 0 |
| Jan 25 | 26 | 3 | 0 | 0 |
| Feb 25 | 49 | 6 | 0 | 0 |
| Mar 25 | 35 | 2 | 1 | 0 |
| Apr 25 | 129 | 2 | 1 | 0 |
| May 25 | 33 | 3 | 3 | 0 |
| Jun 25 | 30 | 4 | 1 | 1 |
| Jul 25 | 28 | 0 | 0 | 0 |
| Aug 25 | 35 | 5 | 1 | 0 |
| Sep 25 | 46 | 2 | 0 | 0 |
| Oct 25 | 65 | 4 | 0 | 0 |
| Nov 25 | 38 | 5 | 2 | 0 |
| Dec 25 | 50 | 10 | 1 | 0 |
| Jan 26 | 33 | 6 | 0 | 2 |
| Feb 26 | 42 | 1 | 1 | 0 |
| Mar 26 | 60 | 2 | 1 | 0 |
| Apr 26 | 58 | 4 | 0 | 0 |
| May 26 | 47 | 5 | 0 | 0 |
| Jun 26 | 42 | 7 | 1 | 0 |
| Jul 26 | 16 | 2 | 0 | 0 |
| Aug 26 | 29 | 3 | 0 | 0 |
| Sep 26 | 42 | 15 | 0 | 0 |
| Oct 26 | 62 | 0 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Philips448
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Unintended System Motion125
- Fracture58
- Material Integrity Problem55
- Application Program Freezes or Fails to Launch50
- Computer Software Problem26
- Adverse Event Without Identified Device or Use Problem20
- Material Split, Cut or Torn18
- Unintended Application Program Shut Down17
- Poor Quality Image16
- Unexpected Shutdown15
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions516
- Insufficient Information32
- Perforation11
- Electric Shock10
- Laceration(s)8
- Discomfort5
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available4
- Pain4
- Unspecified Respiratory Problem3
- Dyspnea2
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
- Philips Lumify Diagnostic Ultrasound SystemOpen, Classified
Z-0152-2027 · Started Sep 24, 2026 · Posted by FDA Oct 16, 2026
Reason given: Ultrasound system may: 1) Not recognize previous transducer registration, prevent scanning, require re-registration with Wi-Fi/cellular access, which may be unavailable outside of the hospital (ambulances/aircraft/field); 2) At connection between transducer USB cable to tablet/phone, there may be an unstable...
Cause as FDA categorized it: Device Design
- Philips EPIQ Elite Diagnostic Ultrasound System Model # 795234Open, Classified
Z-3249-2026 · Started Aug 7, 2026 · Posted by FDA Sep 18, 2026
Reason given: Due to software issue with ultrasound system, doppler sample is displayed shallower than its actual measurement location, potentially causing inaccurate blood flow velocity measurements and misdiagnosis.
Cause as FDA categorized it: Software Design Change
Z-0883-2026 · Started Oct 31, 2025 · Posted by FDA Dec 4, 2025
Reason given: Ultrasound system compatibility issues with Apple devices running iOS 18 may cause a failure to perform live imagining.
Cause as FDA categorized it: Software design
Z-0435-2026 · Started Oct 9, 2025 · Posted by FDA Nov 6, 2025
Reason given: GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.
Cause as FDA categorized it: Device Design
- Ultrasound System 1300, bkSpecto w/battery; Model No. 1300-21 or 1300-S1; System, Imaging, Pulsed Doppler, UltrasonicOpen, Classified
Z-0436-2026 · Started Oct 9, 2025 · Posted by FDA Nov 6, 2025
Reason given: GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.
Cause as FDA categorized it: Device Design
- UUltrasound System 2300, bkActiv w/battery; Model No. 2300-11; System, Imaging, Pulsed Doppler, UltrasonicOpen, Classified
Z-0437-2026 · Started Oct 9, 2025 · Posted by FDA Nov 6, 2025
Reason given: GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.
Cause as FDA categorized it: Device Design
- bk3000 Ultrasound System w/battery; Model No. 2300-66; System, Imaging, Pulsed Doppler, UltrasonicOpen, Classified
Z-0438-2026 · Started Oct 9, 2025 · Posted by FDA Nov 6, 2025
Reason given: GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.
Cause as FDA categorized it: Device Design
- bk3500 Ultrasound System w/battery; Model No. 2300-41; System, Imaging, Pulsed Doppler, UltrasonicOpen, Classified
Z-0439-2026 · Started Oct 9, 2025 · Posted by FDA Nov 6, 2025
Reason given: GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.
Cause as FDA categorized it: Device Design
Showing the 8 newest of 300 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.