Device type · HST
Apparatus, Traction, Non-Powered
Last checked: Nov 20, 2026Source: openFDAMethods & limitations
Product code
HST
Neurology
Device class
Class I
FDA decisions since 2015
1
510(k), De Novo and PMA
Adverse event reports
45
Received Nov 1, 2025 to Nov 20, 2026
How FDA classifies it
- Device type
- Apparatus, Traction, Non-Powered
- Product code
- HST
- Device class
- Class I
- Review panel
- Neurology
- Regulation
- 21 CFR 888.5850
- Specialty
- Neurology & Neurosurgery
- FDA record
- Classification database entry for HST
Latest FDA decisions
K162702 · Decided Dec 30, 2016 · 93 days after receipt
Device type: Apparatus, Traction, Non-Powered (HST) · Orthopedic Surgery
Reports received
45
Received Nov 1, 2025 to Nov 20, 2026
Malfunction reports
34
76% of reports
Injury reports
11
24% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 2 | 0 | 0 | 0 |
| Jan 24 | 1 | 0 | 0 | 0 |
| Feb 24 | 4 | 0 | 0 | 0 |
| Mar 24 | 2 | 0 | 0 | 0 |
| Apr 24 | 5 | 0 | 0 | 0 |
| May 24 | 15 | 0 | 0 | 0 |
| Jun 24 | 11 | 0 | 0 | 0 |
| Jul 24 | 3 | 0 | 0 | 0 |
| Aug 24 | 1 | 0 | 0 | 0 |
| Sep 24 | 1 | 0 | 0 | 0 |
| Oct 24 | 3 | 0 | 0 | 0 |
| Nov 24 | 1 | 0 | 0 | 0 |
| Dec 24 | 0 | 0 | 0 | 0 |
| Jan 25 | 3 | 0 | 0 | 0 |
| Feb 25 | 0 | 0 | 0 | 0 |
| Mar 25 | 1 | 0 | 0 | 0 |
| Apr 25 | 2 | 0 | 0 | 0 |
| May 25 | 0 | 0 | 0 | 0 |
| Jun 25 | 0 | 0 | 0 | 0 |
| Jul 25 | 0 | 0 | 0 | 0 |
| Aug 25 | 1 | 0 | 0 | 0 |
| Sep 25 | 0 | 0 | 0 | 0 |
| Oct 25 | 1 | 0 | 0 | 0 |
| Nov 25 | 6 | 2 | 0 | 0 |
| Dec 25 | 2 | 2 | 0 | 0 |
| Jan 26 | 4 | 1 | 0 | 0 |
| Feb 26 | 1 | 4 | 0 | 0 |
| Mar 26 | 3 | 0 | 0 | 0 |
| Apr 26 | 3 | 0 | 0 | 0 |
| May 26 | 2 | 1 | 0 | 0 |
| Jun 26 | 4 | 1 | 0 | 0 |
| Jul 26 | 3 | 0 | 0 | 0 |
| Aug 26 | 1 | 0 | 0 | 0 |
| Sep 26 | 0 | 0 | 0 | 0 |
| Oct 26 | 5 | 0 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Synthes13
5 more in this ranking with a free account
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Material Twisted/Bent8
- Unintended Movement6
- Break5
7 more in this ranking with a free account
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions36
- Bone Fracture(s)4
- Compartment syndrome3
2 more in this ranking with a free account
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
- HDS Traction Boot II, REF AR-6529-23, The Traction Boot is designed to position, support and/or distract the patient's foot for hip surgeryOpen, Classified
Z-1329-2024 · Started Mar 11, 2024 · Posted by FDA Mar 20, 2024
Reason given: Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for the Hip Distraction System, Traction Boot II, due to the boots being assembled incorrectly. The locking ridge on the front face of the boot key socket is facing upwards, which prevents the socket mechanism on the traction boot from attaching...
Cause as FDA categorized it: Under Investigation by firm
Z-1198-2017 · Started Jan 3, 2017 · Posted by FDA Feb 14, 2017
Reason given: During use, there is a potential scenario which could allow the fastening screws attaching covers to come loose and potentially land in the sterile field causing a delay in surgery.
Cause as FDA categorized it: Device Design
Z-0928-2017 · Started Nov 23, 2016
Reason given: During use, the unit could separate un-expectantly. This failure would allow the patient's arm to fall out of suspension during a procedure, possibly causing moderate injury to the patient. This potential defect is not apparent to the user.
Cause as FDA categorized it: Under Investigation by firm
8 more recall records with a free account
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.