Device type · HSN
Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
HSN
Orthopedic
Device class
Class II
FDA decisions since 2015
45
510(k), De Novo and PMA
Adverse event reports
1,712
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer
- Product code
- HSN
- Device class
- Class II
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3110
- Specialty
- Orthopedic Surgery, Joint Reconstruction
- FDA record
- Classification database entry for HSN
Latest FDA decisions
The 8 newest of 45 since 2015.
- APEX 3D S Diamond Tibia510(k)
K261120 · Decided Oct 7, 2026 · 187 days after receipt
Device type: Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer (HSN) · Orthopedic Surgery
Wright Medical Technology, Inc. (Stryker)
K253159 · Decided Jun 16, 2026 · 263 days after receipt
Device type: Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer (HSN) · Orthopedic Surgery
- Kinos Total Ankle System510(k)
K252454 · Decided Sep 3, 2025 · 30 days after receipt
Device type: Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer (HSN) · Orthopedic Surgery
Wright Medical Technology, Inc. (Stryker)
K250037 · Decided Jun 20, 2025 · 163 days after receipt
Device type: Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer (HSN) · Orthopedic Surgery
K250641 · Decided May 1, 2025 · 58 days after receipt
Device type: Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer (HSN) · Orthopedic Surgery
K240475 · Decided Dec 2, 2024 · 286 days after receipt
Device type: Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer (HSN) · Orthopedic Surgery
Wright Medical Technology, Inc. (Stryker)
K241999 · Decided Nov 1, 2024 · 115 days after receipt
Device type: Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer (HSN) · Orthopedic Surgery
- Kinos Total Ankle System510(k)
K242868 · Decided Oct 18, 2024 · 28 days after receipt
Device type: Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer (HSN) · Orthopedic Surgery
Reports received
1,712
Received Nov 1, 2025 to Nov 19, 2026
Malfunction reports
33
1.9% of reports
Injury reports
1,679
98% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 3 | 68 | 0 | 0 |
| Jan 24 | 1 | 40 | 0 | 0 |
| Feb 24 | 1 | 90 | 0 | 0 |
| Mar 24 | 1 | 50 | 0 | 0 |
| Apr 24 | 1 | 56 | 0 | 0 |
| May 24 | 1 | 71 | 0 | 0 |
| Jun 24 | 1 | 83 | 0 | 0 |
| Jul 24 | 3 | 62 | 0 | 0 |
| Aug 24 | 0 | 65 | 0 | 0 |
| Sep 24 | 2 | 101 | 0 | 0 |
| Oct 24 | 3 | 153 | 1 | 0 |
| Nov 24 | 3 | 108 | 0 | 0 |
| Dec 24 | 2 | 196 | 0 | 0 |
| Jan 25 | 1 | 70 | 0 | 0 |
| Feb 25 | 5 | 121 | 0 | 0 |
| Mar 25 | 0 | 108 | 0 | 0 |
| Apr 25 | 1 | 102 | 0 | 0 |
| May 25 | 4 | 161 | 1 | 0 |
| Jun 25 | 0 | 89 | 0 | 0 |
| Jul 25 | 2 | 103 | 0 | 0 |
| Aug 25 | 2 | 86 | 0 | 0 |
| Sep 25 | 2 | 107 | 0 | 0 |
| Oct 25 | 1 | 175 | 0 | 0 |
| Nov 25 | 2 | 135 | 0 | 0 |
| Dec 25 | 5 | 381 | 0 | 0 |
| Jan 26 | 1 | 102 | 0 | 0 |
| Feb 26 | 4 | 87 | 0 | 0 |
| Mar 26 | 9 | 130 | 0 | 0 |
| Apr 26 | 0 | 157 | 0 | 0 |
| May 26 | 1 | 167 | 0 | 0 |
| Jun 26 | 1 | 99 | 0 | 0 |
| Jul 26 | 0 | 94 | 0 | 0 |
| Aug 26 | 6 | 136 | 0 | 0 |
| Sep 26 | 2 | 80 | 0 | 0 |
| Oct 26 | 2 | 111 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Paragon 28122
- Exactech88
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Adverse Event Without Identified Device or Use Problem801
- Insufficient Device Problem Information465
- Osseointegration Problem243
- Loss of Osseointegration145
- Implant subsidence78
- Fracture47
- Migration26
- Loosening of Implant Not Related to Bone-Ingrowth23
- Naturally Worn15
- Malposition of Device14
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Insufficient Information550
- Inadequate Osseointegration358
- Pain358
- Cyst(s)169
- Joint Impingement106
- Unspecified Infection99
- Osteolysis83
- Extraskeletal ossification80
- Post Operative Wound Infection76
- Swelling/ Edema58
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Wright Medical Technology, Inc.
Z-1477-2024 · Started Mar 6, 2024 · Posted by FDA Apr 4, 2024
Reason given: Stryker has identified an issue that impacts specific lots of Infinity" Resection Guide Adjustment Blocks. The parts within these three lots were found to have been missing an internal screw in the finished instrument assembly. The missing screw is used to lockout the Medial/Lateral adjustment on the INFINITY...
Cause as FDA categorized it: Process change control
Z-0587-2024 · Started Nov 21, 2023 · Posted by FDA Jan 18, 2024
Reason given: Exactech has received 3 complaints for 4 devices related to a loss of vacuum in the inner-most vacuum bag.
Cause as FDA categorized it: Packaging change control
- Stryker Infinity Alignment Frame Distal Sub-Assembly, Catalog #33600020, non-sterile.Open, Classified
Wright Medical Technology, Inc.
Z-2105-2023 · Started Jun 13, 2023 · Posted by FDA Jul 7, 2023
Reason given: The devices were manufactured with the pin sleeve holes too narrow, which does not allow the pin sleeves to pass through the pin sleeve holes, thereby preventing pin placement through the pin holes.
Cause as FDA categorized it: Process control
Wright Medical Technology, Inc.
Z-0139-2023 · Started Sep 30, 2022 · Posted by FDA Oct 27, 2022
Reason given: The tibial tray lock detail is oversized (larger than specification).
Cause as FDA categorized it: Device Design
Wright Medical Technology, Inc.
Z-0140-2023 · Started Sep 30, 2022 · Posted by FDA Oct 27, 2022
Reason given: The tibial tray lock detail is oversized (larger than specification).
Cause as FDA categorized it: Device Design
- Stryker INFINITY Resect Guide for INBONE Talus, Size 4, Model #33620254, non-sterile, an instrument used with the INFINITY Total Ankle System.Open, Classified
Wright Medical Technology, Inc.
Z-1291-2022 · Started Jun 2, 2022
Reason given: The instrument is incorrectly color coded.
Cause as FDA categorized it: Nonconforming Material/Component
Z-0315-2022 · Started Nov 4, 2021
Reason given: The package contents and package labeling do not match.
Cause as FDA categorized it: Nonconforming Material/Component
Z-0316-2022 · Started Nov 4, 2021
Reason given: The package contents and package labeling do not match.
Cause as FDA categorized it: Nonconforming Material/Component
Showing the 8 newest of 47 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.