Device type · HJO
Biomicroscope, Slit-Lamp, Ac-Powered
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
HJO
Ophthalmic
Device class
Class II
FDA decisions since 2015
23
510(k), De Novo and PMA
Adverse event reports
16
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Biomicroscope, Slit-Lamp, Ac-Powered
- Product code
- HJO
- Device class
- Class II
- Review panel
- Ophthalmic
- Regulation
- 21 CFR 886.1850
- Specialty
- Ophthalmology
- FDA record
- Classification database entry for HJO
Latest FDA decisions
The 8 newest of 23 since 2015.
K262637 · Decided Sep 18, 2026 · 51 days after receipt
Device type: Biomicroscope, Slit-Lamp, Ac-Powered (HJO) · Ophthalmology
- VISUREF 600510(k)
K253834 · Decided Mar 18, 2026 · 107 days after receipt
Device type: Biomicroscope, Slit-Lamp, Ac-Powered (HJO) · Ophthalmology
- HP-OCT (HP-OCT)510(k)
K231760 · Decided Mar 8, 2024 · 266 days after receipt
Device type: Biomicroscope, Slit-Lamp, Ac-Powered (HJO) · Ophthalmology
- Kowa SL-19510(k)
K222372 · Decided Nov 21, 2022 · 108 days after receipt
Device type: Biomicroscope, Slit-Lamp, Ac-Powered (HJO) · Ophthalmology
Medimaging Integrated Solution, Inc (Miis)
K193188 · Decided Jan 14, 2020 · 57 days after receipt
Device type: Biomicroscope, Slit-Lamp, Ac-Powered (HJO) · Ophthalmology
- Sunkingdom Slit Lamp510(k)
Chongqing Sunkingdom Medical Instrument Co., Ltd.
K182306 · Decided Nov 22, 2018 · 90 days after receipt
Device type: Biomicroscope, Slit-Lamp, Ac-Powered (HJO) · Ophthalmology
- IOLMaster 700510(k)
K173771 · Decided Aug 24, 2018 · 256 days after receipt
Device type: Biomicroscope, Slit-Lamp, Ac-Powered (HJO) · Ophthalmology
Chongqing Kanghua Ruiming S&T Co., Ltd.
K171877 · Decided Dec 21, 2017 · 181 days after receipt
Device type: Biomicroscope, Slit-Lamp, Ac-Powered (HJO) · Ophthalmology
Reports received
16
Received Nov 1, 2025 to Nov 19, 2026
Malfunction reports
6
38% of reports
Injury reports
10
63% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 0 | 13 | 0 | 0 |
| Jan 24 | 1 | 1 | 0 | 0 |
| Feb 24 | 0 | 0 | 0 | 0 |
| Mar 24 | 0 | 5 | 0 | 0 |
| Apr 24 | 0 | 4 | 0 | 0 |
| May 24 | 0 | 0 | 0 | 0 |
| Jun 24 | 0 | 1 | 0 | 0 |
| Jul 24 | 0 | 2 | 0 | 0 |
| Aug 24 | 0 | 0 | 0 | 0 |
| Sep 24 | 0 | 3 | 0 | 0 |
| Oct 24 | 0 | 0 | 0 | 0 |
| Nov 24 | 1 | 0 | 0 | 0 |
| Dec 24 | 0 | 0 | 0 | 0 |
| Jan 25 | 0 | 0 | 0 | 0 |
| Feb 25 | 1 | 0 | 0 | 0 |
| Mar 25 | 0 | 4 | 0 | 0 |
| Apr 25 | 0 | 0 | 0 | 0 |
| May 25 | 0 | 0 | 0 | 0 |
| Jun 25 | 0 | 2 | 0 | 0 |
| Jul 25 | 0 | 0 | 0 | 0 |
| Aug 25 | 0 | 0 | 0 | 0 |
| Sep 25 | 1 | 0 | 0 | 0 |
| Oct 25 | 0 | 2 | 0 | 0 |
| Nov 25 | 2 | 0 | 0 | 0 |
| Dec 25 | 1 | 7 | 0 | 0 |
| Jan 26 | 0 | 1 | 0 | 0 |
| Feb 26 | 0 | 0 | 0 | 0 |
| Mar 26 | 0 | 1 | 0 | 0 |
| Apr 26 | 0 | 0 | 0 | 0 |
| May 26 | 0 | 0 | 0 | 0 |
| Jun 26 | 0 | 0 | 0 | 0 |
| Jul 26 | 2 | 0 | 0 | 0 |
| Aug 26 | 1 | 1 | 0 | 0 |
| Sep 26 | 0 | 0 | 0 | 0 |
| Oct 26 | 0 | 0 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Incorrect, Inadequate or Imprecise Result or Readings11
- Misassembly During Maintenance/Repair10
- Adverse Event Without Identified Device or Use Problem2
- Insufficient Device Problem Information1
- Low Readings1
- Output Problem1
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Visual Impairment10
- No Clinical Signs, Symptoms or Conditions4
- Insufficient Information2
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-1560-2021 · Started Apr 15, 2021
Reason given: Examination data and patient name may be mixed up when printing or generating a pdf with the interface at Pacific Coast and Laser Institute (PCLI).
Cause as FDA categorized it: Software design
- IOLMaster 700Terminated
Z-1133-2021 · Started Dec 7, 2020
Reason given: When using software 1.90.2.09 or 1.90.8.06 and using modality worklist functionality for patient data transfer, the selection of the patient in the patient list may not match the patient information displayed on the right side of the screen
Cause as FDA categorized it: Software design
Z-0108-2019 · Started Jun 25, 2018
Reason given: Calibration setting was incorrectly entered during manufacturing causing the optical coherence tomography (OCT) to obtain the incorrect two-dimensional images of ocular structures of the anterior and posterior segments of the eye.
Cause as FDA categorized it: Employee error
Z-2587-2016 · Started Jul 1, 2016
Reason given: Nidek Inc. received information from our manufacturer NIDEK CO. Japan that the shown axial lengths were much shorter at AL-Scan measurement.
Cause as FDA categorized it: Software in the Use Environment
Z-0357-2016 · Started Oct 27, 2015 · Posted by FDA Dec 1, 2015
Reason given: IOL Master software versions 7.5 and 7.7 calculation printouts and exported reports can contain the wrong IOL power data.
Cause as FDA categorized it: Software design
Z-0358-2016 · Started Oct 27, 2015 · Posted by FDA Dec 1, 2015
Reason given: IOL Master software versions 7.5 and 7.7 calculation printouts and exported reports can contain the wrong IOL power data.
Cause as FDA categorized it: Software design
Z-0333-2014 · Started Oct 23, 2013 · Posted by FDA Nov 15, 2013
Reason given: The user database delivered for download contained an erroneous value. Specifically the "aO" constant of the lens AMO Sensar AR40E is listed as "0.0" in the internal lens database of the IOLMaster. The correct value is -2.420. This could lead to the implantation of lens with a wrong diopter value.
Cause as FDA categorized it: Software in the Use Environment
Z-0334-2014 · Started Oct 23, 2013 · Posted by FDA Nov 15, 2013
Reason given: The user database delivered for download contained an erroneous value. Specifically the "aO" constant of the lens AMO Sensar AR40E is listed as "0.0" in the internal lens database of the IOLMaster. The correct value is -2.420. This could lead to the implantation of lens with a wrong diopter value.
Cause as FDA categorized it: Software in the Use Environment
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.