Device type · GEH
Unit, Cryosurgical, Accessories
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
GEH
General & Plastic Surgery
Device class
Class II
FDA decisions since 2015
106
510(k), De Novo and PMA
Adverse event reports
282
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Unit, Cryosurgical, Accessories
- Product code
- GEH
- Device class
- Class II
- Review panel
- General & Plastic Surgery
- Regulation
- 21 CFR 878.4350
- Specialty
- General & Plastic Surgery
- FDA record
- Classification database entry for GEH
Latest FDA decisions
The 8 newest of 106 since 2015.
K260520 · Decided Oct 27, 2026 · 252 days after receipt
Device type: Unit, Cryosurgical, Accessories (GEH) · General & Plastic Surgery
K262420 · Decided Sep 2, 2026 · 48 days after receipt
Device type: Unit, Cryosurgical, Accessories (GEH) · General & Plastic Surgery
K261068 · Decided Jun 30, 2026 · 90 days after receipt
Device type: Unit, Cryosurgical, Accessories (GEH) · General & Plastic Surgery
K261806 · Decided Jun 9, 2026 · 8 days after receipt
Device type: Unit, Cryosurgical, Accessories (GEH) · General & Plastic Surgery
- ERBECRYO 2 Cryosurgical Unit and Accessories: ERBECRYO 2 cryosurgical unit; Erbe Flexible Cryoprobe510(k)
K253230 · Decided Apr 22, 2026 · 205 days after receipt
Device type: Unit, Cryosurgical, Accessories (GEH) · General & Plastic Surgery
Medical Brands Laboratories B.V.
K252198 · Decided Apr 2, 2026 · 262 days after receipt
Device type: Unit, Cryosurgical, Accessories (GEH) · General & Plastic Surgery
K260377 · Decided Feb 24, 2026 · 19 days after receipt
Device type: Unit, Cryosurgical, Accessories (GEH) · General & Plastic Surgery
K251928 · Decided Feb 13, 2026 · 235 days after receipt
Device type: Unit, Cryosurgical, Accessories (GEH) · General & Plastic Surgery
Reports received
282
Received Nov 1, 2025 to Nov 19, 2026. Includes 2 of another or unstated type.
Malfunction reports
177
63% of reports
Injury reports
94
33% of reports
Death reports
9
3.2% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 6 | 3 | 0 | 0 |
| Jan 24 | 3 | 3 | 0 | 0 |
| Feb 24 | 4 | 8 | 1 | 0 |
| Mar 24 | 8 | 3 | 0 | 0 |
| Apr 24 | 4 | 0 | 0 | 0 |
| May 24 | 3 | 8 | 0 | 0 |
| Jun 24 | 13 | 7 | 0 | 0 |
| Jul 24 | 8 | 4 | 0 | 0 |
| Aug 24 | 7 | 9 | 0 | 0 |
| Sep 24 | 7 | 2 | 0 | 0 |
| Oct 24 | 8 | 2 | 0 | 0 |
| Nov 24 | 10 | 1 | 0 | 0 |
| Dec 24 | 11 | 7 | 0 | 0 |
| Jan 25 | 5 | 2 | 0 | 0 |
| Feb 25 | 8 | 3 | 0 | 0 |
| Mar 25 | 9 | 2 | 0 | 0 |
| Apr 25 | 10 | 1 | 0 | 0 |
| May 25 | 15 | 5 | 0 | 0 |
| Jun 25 | 3 | 1 | 0 | 0 |
| Jul 25 | 17 | 1 | 0 | 0 |
| Aug 25 | 14 | 11 | 0 | 0 |
| Sep 25 | 8 | 5 | 1 | 0 |
| Oct 25 | 6 | 4 | 0 | 0 |
| Nov 25 | 17 | 3 | 0 | 0 |
| Dec 25 | 19 | 4 | 0 | 0 |
| Jan 26 | 22 | 5 | 1 | 0 |
| Feb 26 | 21 | 2 | 0 | 0 |
| Mar 26 | 18 | 2 | 7 | 0 |
| Apr 26 | 10 | 10 | 0 | 0 |
| May 26 | 13 | 4 | 0 | 0 |
| Jun 26 | 9 | 8 | 0 | 0 |
| Jul 26 | 12 | 31 | 0 | 0 |
| Aug 26 | 11 | 11 | 0 | 2 |
| Sep 26 | 16 | 6 | 0 | 0 |
| Oct 26 | 9 | 8 | 1 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Adverse Event Without Identified Device or Use Problem76
- Break70
- Material Rupture66
- Insufficient Cooling20
- Thermal Decomposition of Device20
- Gas/Air Leak12
- Nonstandard Device11
- Insufficient Device Problem Information10
- Noise, Audible6
- Free or Unrestricted Flow5
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions129
- Pneumothorax28
- Pain26
- Hemorrhage/Blood Loss/Bleeding22
- Tinnitus14
- Hearing Impairment13
- Pleural Effusion9
- Burn(s)8
- Pneumonia8
- Superficial (First Degree) Burn8
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
- Glypto Device, Glypto Cryo Machine;Open, Classified
Z-0080-2027 · Started Aug 24, 2026 · Posted by FDA Oct 2, 2026
Reason given: Lack of 510(k) clearance.
Cause as FDA categorized it: Under Investigation by firm
- ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000Open, Classified
Z-2938-2026 · Started Jun 26, 2026 · Posted by FDA Aug 12, 2026
Reason given: A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.
Cause as FDA categorized it: Device Design
- Flexible Cryoprobe (OD 1.1mm, L1.15mm) w/ oversheath (OD2.6mm, L817mm) REF: 20402-401 STERILE EO. For surgical useOpen, Classified
Z-1566-2026 · Started Feb 12, 2026 · Posted by FDA Mar 20, 2026
Reason given: Probes may rupture/burst during activation
Cause as FDA categorized it: Process control
Z-1567-2026 · Started Feb 12, 2026 · Posted by FDA Mar 20, 2026
Reason given: Probes may rupture/burst during activation
Cause as FDA categorized it: Process control
Z-1568-2026 · Started Feb 12, 2026 · Posted by FDA Mar 20, 2026
Reason given: Probes may rupture/burst during activation
Cause as FDA categorized it: Process control
- ICEfx Cryoablation System, portable helium-free desktop cryoablation unit, Material Numbers (UPN) FPRCH8000, FPRCH8000-02, H749396118000Open, Classified
Z-2550-2025 · Started Aug 18, 2025 · Posted by FDA Sep 11, 2025
Reason given: Certain desiccant tube subassemblies were built with end caps that were not correctly tightened.
Cause as FDA categorized it: Process control
Z-2551-2025 · Started Aug 18, 2025 · Posted by FDA Sep 11, 2025
Reason given: Certain desiccant tube subassemblies were built with end caps that were not correctly tightened.
Cause as FDA categorized it: Process control
Z-1686-2025 · Started Mar 24, 2025 · Posted by FDA Apr 29, 2025
Reason given: Boston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to...
Cause as FDA categorized it: Process control
Showing the 8 newest of 107 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.