Device type · FAJ
Cystoscope And Accessories, Flexible/Rigid
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
FAJ
Gastroenterology & Urology
Device class
Class II
FDA decisions since 2015
46
510(k), De Novo and PMA
Adverse event reports
6,303
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Cystoscope And Accessories, Flexible/Rigid
- Product code
- FAJ
- Device class
- Class II
- Review panel
- Gastroenterology & Urology
- Regulation
- 21 CFR 876.1500
- Specialty
- GI, Urology & Nephrology, Urology
- FDA record
- Classification database entry for FAJ
Latest FDA decisions
The 8 newest of 46 since 2015.
K263075 · Decided Oct 30, 2026 · 58 days after receipt
Device type: Cystoscope And Accessories, Flexible/Rigid (FAJ) · GI, Urology & Nephrology
K260649 · Decided May 21, 2026 · 83 days after receipt
Device type: Cystoscope And Accessories, Flexible/Rigid (FAJ) · GI, Urology & Nephrology
K253696 · Decided May 11, 2026 · 171 days after receipt
Device type: Cystoscope And Accessories, Flexible/Rigid (FAJ) · GI, Urology & Nephrology
K253892 · Decided Apr 16, 2026 · 133 days after receipt
Device type: Cystoscope And Accessories, Flexible/Rigid (FAJ) · GI, Urology & Nephrology
K253905 · Decided Apr 7, 2026 · 123 days after receipt
Device type: Cystoscope And Accessories, Flexible/Rigid (FAJ) · GI, Urology & Nephrology
American Medical Endoscopy, Inc.
K260271 · Decided Mar 27, 2026 · 58 days after receipt
Device type: Cystoscope And Accessories, Flexible/Rigid (FAJ) · GI, Urology & Nephrology
- UV5000W Handle510(k)
K253226 · Decided Oct 23, 2025 · 24 days after receipt
Device type: Cystoscope And Accessories, Flexible/Rigid (FAJ) · GI, Urology & Nephrology
- Zenflow Spring Scope510(k)
K251140 · Decided Sep 23, 2025 · 162 days after receipt
Device type: Cystoscope And Accessories, Flexible/Rigid (FAJ) · GI, Urology & Nephrology
Reports received
6,303
Received Nov 1, 2025 to Nov 19, 2026. Includes 4 of another or unstated type.
Malfunction reports
6,222
99% of reports
Injury reports
76
1.2% of reports
Death reports
1
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 115 | 53 | 1 | 0 |
| Jan 24 | 81 | 13 | 0 | 0 |
| Feb 24 | 70 | 37 | 0 | 0 |
| Mar 24 | 68 | 64 | 0 | 0 |
| Apr 24 | 51 | 2 | 0 | 0 |
| May 24 | 59 | 4 | 0 | 0 |
| Jun 24 | 67 | 6 | 0 | 0 |
| Jul 24 | 60 | 5 | 0 | 0 |
| Aug 24 | 68 | 24 | 0 | 0 |
| Sep 24 | 43 | 1 | 0 | 0 |
| Oct 24 | 75 | 0 | 0 | 0 |
| Nov 24 | 51 | 3 | 0 | 0 |
| Dec 24 | 33 | 8 | 0 | 0 |
| Jan 25 | 55 | 5 | 0 | 0 |
| Feb 25 | 96 | 3 | 0 | 0 |
| Mar 25 | 87 | 0 | 0 | 0 |
| Apr 25 | 79 | 1 | 0 | 0 |
| May 25 | 89 | 1 | 0 | 0 |
| Jun 25 | 102 | 5 | 0 | 0 |
| Jul 25 | 182 | 20 | 0 | 0 |
| Aug 25 | 264 | 0 | 0 | 0 |
| Sep 25 | 243 | 0 | 0 | 0 |
| Oct 25 | 789 | 15 | 0 | 0 |
| Nov 25 | 376 | 12 | 0 | 0 |
| Dec 25 | 275 | 7 | 0 | 0 |
| Jan 26 | 384 | 0 | 0 | 0 |
| Feb 26 | 356 | 16 | 0 | 0 |
| Mar 26 | 393 | 2 | 0 | 0 |
| Apr 26 | 924 | 4 | 0 | 0 |
| May 26 | 719 | 7 | 0 | 0 |
| Jun 26 | 524 | 0 | 0 | 0 |
| Jul 26 | 715 | 3 | 0 | 0 |
| Aug 26 | 556 | 0 | 0 | 0 |
| Sep 26 | 582 | 7 | 1 | 0 |
| Oct 26 | 418 | 18 | 0 | 4 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Shirakawa Olympus5,876
- Aizu Olympus117
- Olympus99
- Ambu7
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Detachment of Device or Device Component2,551
- Corroded841
- Loss of or Failure to Bond794
- Failure to Clean Adequately474
- Device Reprocessing Problem324
- Material Separation250
- Break238
- Crack224
- Degraded215
- Material Fragmentation195
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions6,186
- Insufficient Information47
- Urinary Tract Infection27
- Unspecified Infection19
- Dysuria18
- Bacterial Infection16
- Hematuria8
- Sepsis8
- Foreign Body In Patient7
- Urinary Frequency / Polyuria6
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
- Model No. CYF-5, CystonephrofiberscopeOpen, Classified
Z-0684-2024 · Started Dec 4, 2023 · Posted by FDA Jan 9, 2024
Reason given: Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
Cause as FDA categorized it: Under Investigation by firm
- Model No. CYF-5R, CystonephrofiberscopeOpen, Classified
Z-0685-2024 · Started Dec 4, 2023 · Posted by FDA Jan 9, 2024
Reason given: Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
Cause as FDA categorized it: Under Investigation by firm
- Model No. CYF-VH, CYSTO-NEPHRO VIDEOSCOPEOpen, Classified
Z-0689-2024 · Started Dec 4, 2023 · Posted by FDA Jan 9, 2024
Reason given: Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
Cause as FDA categorized it: Under Investigation by firm
- Model No. CYF-VHR, CYSTO-NEPHRO VIDEOSCOPEOpen, Classified
Z-0690-2024 · Started Dec 4, 2023 · Posted by FDA Jan 9, 2024
Reason given: Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
Cause as FDA categorized it: Under Investigation by firm
- Model No. CYF-V2, VISERA Cysto-Nephro VideoscopeOpen, Classified
Z-0691-2024 · Started Dec 4, 2023 · Posted by FDA Jan 9, 2024
Reason given: Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
Cause as FDA categorized it: Under Investigation by firm
- Model No. CYF-V2R, VISERA Cysto-Nephro VideoscopeOpen, Classified
Z-0692-2024 · Started Dec 4, 2023 · Posted by FDA Jan 9, 2024
Reason given: Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
Cause as FDA categorized it: Under Investigation by firm
- Cystonephroscope-endoscopic diagnosis and treatment within the bladder, urethra, and kidney Model Numbers: CYF-5, CYF-VH, CYF-VHR.Open, Classified
Z-0331-2024 · Started Sep 25, 2023 · Posted by FDA Nov 17, 2023
Reason given: Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Cause as FDA categorized it: Under Investigation by firm
Z-1234-2022 · Started May 3, 2022
Reason given: The endoscope may experience a loss of image when activating monopolar electrocautery, due a manufacturing assembly process issue.
Cause as FDA categorized it: Process control
Showing the 8 newest of 22 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.