Device type · DWA
Control, Pump Speed, Cardiopulmonary Bypass
Last checked: Nov 20, 2026Source: openFDAMethods & limitations
Product code
DWA
Cardiovascular
Device class
Class II
FDA decisions since 2015
6
510(k), De Novo and PMA
Adverse event reports
317
Received Nov 1, 2025 to Nov 20, 2026
How FDA classifies it
- Device type
- Control, Pump Speed, Cardiopulmonary Bypass
- Product code
- DWA
- Device class
- Class II
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.4380
- Specialty
- Cardiology
- FDA record
- Classification database entry for DWA
Latest FDA decisions
- INSPIRA ART100510(k)
Inspira Technologies Oxy B.H.N. , Ltd.
K232788 · Decided May 24, 2024 · 256 days after receipt
Device type: Control, Pump Speed, Cardiopulmonary Bypass (DWA) · Cardiology
K230698 · Decided Apr 18, 2023 · 36 days after receipt
Device type: Control, Pump Speed, Cardiopulmonary Bypass (DWA) · Cardiology
K221491 · Decided Feb 3, 2023 · 256 days after receipt
Device type: Control, Pump Speed, Cardiopulmonary Bypass (DWA) · Cardiology
3 more FDA decisions with a free account
Reports received
317
Received Nov 1, 2025 to Nov 20, 2026
Malfunction reports
308
97% of reports
Injury reports
8
2.5% of reports
Death reports
1
0.3% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 24 | 0 | 0 | 0 |
| Jan 24 | 40 | 1 | 1 | 0 |
| Feb 24 | 35 | 3 | 0 | 0 |
| Mar 24 | 18 | 2 | 0 | 0 |
| Apr 24 | 37 | 2 | 0 | 0 |
| May 24 | 27 | 2 | 0 | 0 |
| Jun 24 | 36 | 3 | 1 | 0 |
| Jul 24 | 38 | 0 | 0 | 0 |
| Aug 24 | 19 | 0 | 0 | 0 |
| Sep 24 | 30 | 0 | 0 | 0 |
| Oct 24 | 40 | 0 | 0 | 0 |
| Nov 24 | 30 | 0 | 0 | 0 |
| Dec 24 | 25 | 0 | 0 | 0 |
| Jan 25 | 35 | 0 | 0 | 0 |
| Feb 25 | 31 | 2 | 3 | 0 |
| Mar 25 | 28 | 0 | 0 | 0 |
| Apr 25 | 20 | 0 | 0 | 0 |
| May 25 | 32 | 1 | 0 | 0 |
| Jun 25 | 29 | 0 | 0 | 0 |
| Jul 25 | 21 | 0 | 0 | 0 |
| Aug 25 | 16 | 0 | 0 | 0 |
| Sep 25 | 44 | 0 | 0 | 0 |
| Oct 25 | 32 | 3 | 0 | 0 |
| Nov 25 | 29 | 1 | 0 | 0 |
| Dec 25 | 39 | 0 | 1 | 0 |
| Jan 26 | 28 | 0 | 0 | 0 |
| Feb 26 | 24 | 0 | 0 | 0 |
| Mar 26 | 24 | 1 | 0 | 0 |
| Apr 26 | 22 | 1 | 0 | 0 |
| May 26 | 21 | 0 | 0 | 0 |
| Jun 26 | 28 | 2 | 0 | 0 |
| Jul 26 | 16 | 1 | 0 | 0 |
| Aug 26 | 26 | 0 | 0 | 0 |
| Sep 26 | 24 | 1 | 0 | 0 |
| Oct 26 | 27 | 1 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Thoratec18
3 more in this ranking with a free account
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Insufficient Device Problem Information82
- Battery Problem49
- Device Alarm System37
7 more in this ranking with a free account
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions311
- Insufficient Information5
- Convulsion/Seizure2
2 more in this ranking with a free account
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-0926-2023 · Started Dec 5, 2022 · Posted by FDA Jan 19, 2023
Reason given: A software update (v1.1.5) has been developed to address the issue of Critical Failure which can occur on the LifeSPARC Controller. The Critical Failure was first addressed in the firm's recall initiated July 21, 2022. During the Critical Failure, the software freezes or crashes and the screen does not display data.
Cause as FDA categorized it: Software design
Z-1763-2022 · Started Jul 21, 2022 · Posted by FDA Sep 23, 2022
Reason given: Critical Failure of the LifeSPARC controller when the software freezes or crashes and the screen does not display data.
Cause as FDA categorized it: Under Investigation by firm
Z-0787-2022 · Started Dec 17, 2021
Reason given: One ferrite ring expected to be installed on centrifugal pump system power cable may not be present.
Cause as FDA categorized it: Process control
9 more recall records with a free account
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.