All In MedTech

Device type · DPT

Probe, Blood-Flow, Extravascular

FDA groups similar devices under a three-letter product code. This page gathers what FDA's public data says about DPT: how it is classified, what has been cleared or approved, and what has been reported and recalled.

Last checked: Nov 19, 2026Source: openFDAMethods & limitations

Product code

DPT

Cardiovascular

Device class

Class II

FDA decisions since 2015

4

510(k), De Novo and PMA

Adverse event reports

76

Received Nov 1, 2025 to Nov 20, 2026

How FDA classifies it

Device type
Probe, Blood-Flow, Extravascular
Product code
DPT
Device class
Class II
Review panel
Cardiovascular
Regulation
21 CFR 870.2120
Specialty
Cardiology

Latest FDA decisions

All decisions for DPT

1 more FDA decisions with a free account

Reports received

76

Received Nov 1, 2025 to Nov 20, 2026

Malfunction reports

75

99% of reports

Injury reports

1

1.3% of reports

Death reports

0

0.0% of reports

Adverse event reports received per month

By the date FDA received each report, stacked by event type. The newest month is incomplete.

Open in the MAUDE Monitor
  • Malfunction
  • Injury
  • Death
  • Other
View as table
MonthMalfunctionInjuryDeathOther
Dec 232000
Jan 2418000
Feb 246000
Mar 245000
Apr 2428000
May 2414000
Jun 2411000
Jul 245000
Aug 247000
Sep 2410000
Oct 248000
Nov 2420000
Dec 2437000
Jan 258000
Feb 253000
Mar 251000
Apr 252000
May 253000
Jun 251000
Jul 252000
Aug 258000
Sep 252000
Oct 254000
Nov 2514000
Dec 255000
Jan 263000
Feb 266100
Mar 2616000
Apr 266000
May 260000
Jun 260000
Jul 2611000
Aug 266000
Sep 264000
Oct 264000
Nov 260000

Manufacturers named most often

Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.

Device problems as coded

Last 12 months. FDA's terms. A report can carry several.

  • Fluid/Blood Leak50
  • Incorrect, Inadequate or Imprecise Result or Readings6
  • Leak/Splash6

7 more in this ranking with a free account

Patient problems as coded

Last 12 months. FDA's terms. A report can carry several.

  • No Clinical Signs, Symptoms or Conditions75
  • Dyspnea1
  • Perforation of Vessels1

Recent recalls

From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.

  • Smiths Medical ASD, Inc.

    Z-0184-2025 · Started Sep 24, 2024 · Posted by FDA Oct 23, 2024

    Reason given: Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    Cause as FDA categorized it: Process control

  • Smiths Medical ASD, Inc.

    Z-0185-2025 · Started Sep 24, 2024 · Posted by FDA Oct 23, 2024

    Reason given: Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    Cause as FDA categorized it: Process control

  • Smiths Medical ASD, Inc.

    Z-0186-2025 · Started Sep 24, 2024 · Posted by FDA Oct 23, 2024

    Reason given: Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

    Cause as FDA categorized it: Process control

21 more recall records with a free account

Back to all tools

For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.