Device type · DPS
Electrocardiograph
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
DPS
Cardiovascular
Device class
Class II
FDA decisions since 2015
91
510(k), De Novo and PMA
Adverse event reports
127
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Electrocardiograph
- Product code
- DPS
- Device class
- Class II
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.2340
- Specialty
- Cardiology, Electrophysiology
- FDA record
- Classification database entry for DPS
Latest FDA decisions
The 8 newest of 91 since 2015.
- Kinetic Software Library510(k)
K260640 · Decided Oct 23, 2026 · 238 days after receipt
Device type: Electrocardiograph (DPS) · Cardiology
K260005 · Decided Oct 5, 2026 · 276 days after receipt
Device type: Electrocardiograph (DPS) · Cardiology
K260250 · Decided Sep 29, 2026 · 245 days after receipt
Device type: Electrocardiograph (DPS) · Cardiology
K252463 · Decided Jul 2, 2026 · 330 days after receipt
Device type: Electrocardiograph (DPS) · Cardiology
Ge Medical Systems Information Technologies, Inc.
K251670 · Decided Jan 26, 2026 · 241 days after receipt
Device type: Electrocardiograph (DPS) · Cardiology
K252361 · Decided Dec 22, 2025 · 146 days after receipt
Device type: Electrocardiograph (DPS) · Cardiology
- Withings BeamO (SCT02)510(k)
K252474 · Decided Nov 10, 2025 · 96 days after receipt
Device type: Electrocardiograph (DPS) · Cardiology
K250569 · Decided Aug 6, 2025 · 161 days after receipt
Device type: Electrocardiograph (DPS) · Cardiology
Reports received
127
Received Nov 1, 2025 to Nov 19, 2026. Includes 1 of another or unstated type.
Malfunction reports
115
91% of reports
Injury reports
9
7.1% of reports
Death reports
2
1.6% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 2 | 1 | 0 | 0 |
| Jan 24 | 4 | 0 | 0 | 0 |
| Feb 24 | 2 | 0 | 0 | 0 |
| Mar 24 | 3 | 0 | 0 | 0 |
| Apr 24 | 6 | 1 | 0 | 0 |
| May 24 | 5 | 0 | 0 | 0 |
| Jun 24 | 3 | 0 | 0 | 0 |
| Jul 24 | 14 | 0 | 0 | 0 |
| Aug 24 | 7 | 0 | 0 | 0 |
| Sep 24 | 6 | 0 | 0 | 0 |
| Oct 24 | 4 | 0 | 0 | 0 |
| Nov 24 | 5 | 1 | 0 | 0 |
| Dec 24 | 3 | 0 | 0 | 0 |
| Jan 25 | 2 | 0 | 0 | 0 |
| Feb 25 | 2 | 0 | 0 | 0 |
| Mar 25 | 3 | 0 | 0 | 0 |
| Apr 25 | 3 | 0 | 0 | 0 |
| May 25 | 19 | 1 | 0 | 0 |
| Jun 25 | 0 | 0 | 1 | 0 |
| Jul 25 | 2 | 1 | 0 | 0 |
| Aug 25 | 4 | 1 | 0 | 0 |
| Sep 25 | 3 | 2 | 0 | 0 |
| Oct 25 | 3 | 1 | 3 | 0 |
| Nov 25 | 3 | 2 | 0 | 0 |
| Dec 25 | 1 | 0 | 0 | 0 |
| Jan 26 | 1 | 0 | 0 | 0 |
| Feb 26 | 0 | 0 | 0 | 0 |
| Mar 26 | 3 | 0 | 0 | 0 |
| Apr 26 | 14 | 0 | 0 | 0 |
| May 26 | 25 | 2 | 1 | 0 |
| Jun 26 | 13 | 3 | 0 | 0 |
| Jul 26 | 17 | 1 | 1 | 0 |
| Aug 26 | 10 | 1 | 0 | 0 |
| Sep 26 | 19 | 0 | 0 | 0 |
| Oct 26 | 9 | 0 | 0 | 1 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Incorrect Measurement43
- High Readings35
- Incorrect, Inadequate or Imprecise Result or Readings18
- Patient Data Problem17
- Imprecise or Erratic Results7
- Wireless Communication Problem4
- Overheating of Device3
- Human-Device Interface Problem2
- Poor Quality Image2
- Sparking2
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions117
- Asystole2
- Insufficient Information2
- Tachycardia2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Arrhythmia1
- Bacterial Infection1
- Bradycardia1
- Cardiac Arrest1
- Dyspnea1
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-3012-2026 · Started Jul 20, 2026 · Posted by FDA Aug 25, 2026
Reason given: A communication anomoly may result in incorrect electrocardiogram waveforms of a patient and an inaccurate beats-per-minute calculation.
Cause as FDA categorized it: Software design
- Cardiac Workstation 5000. Product Number: 860439.Open, Classified
Z-2955-2026 · Started Jul 1, 2026 · Posted by FDA Aug 14, 2026
Reason given: Device will intermittently provide inaccurate data related to native heart rate.
Cause as FDA categorized it: Software design
- Cardiac Workstation 7000. Product Number: 860441.Open, Classified
Z-2956-2026 · Started Jul 1, 2026 · Posted by FDA Aug 14, 2026
Reason given: Device will intermittently provide inaccurate data related to native heart rate.
Cause as FDA categorized it: Software design
GE Medical Systems Information Technologies Inc
Z-2876-2026 · Started Jun 29, 2026 · Posted by FDA Aug 4, 2026
Reason given: Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity. As a result, data from a previous patient may remain on the screen and could be incorrectly associated with a subsequent patient.
Cause as FDA categorized it: Software design
- Cardiac Workstation 5000; Model Number: 860439;Open, Classified
Z-0165-2026 · Started Sep 12, 2025 · Posted by FDA Oct 16, 2025
Reason given: Failure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short.
Cause as FDA categorized it: Process control
- Cardiac Workstation 7000; Model Number: 860441;Open, Classified
Z-0166-2026 · Started Sep 12, 2025 · Posted by FDA Oct 16, 2025
Reason given: Failure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short.
Cause as FDA categorized it: Process control
Z-0231-2026 · Started Aug 22, 2025 · Posted by FDA Oct 22, 2025
Reason given: If a user incorrectly places the Acquisition Module into the Acquisition Module Holder, it can result in excessive bending and wear of the Acquisition Trunk Cable. This is visible as a damaged sheath or exposed wires. If a damaged Acquisition Trunk Cable is not recognized and removed from service, the cable can...
Cause as FDA categorized it: Device Design
- Welch Allyn CP150 Electrocardiograph:Open, Classified
Z-1560-2025 · Started Mar 24, 2025 · Posted by FDA Apr 14, 2025
Reason given: There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.
Cause as FDA categorized it: Process design
Showing the 8 newest of 98 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.