Device type · CJE
Nitrophenylphosphate, Alkaline Phosphatase Or Isoenzymes
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
CJE
Clinical Chemistry
Device class
Class II
FDA decisions since 2015
7
510(k), De Novo and PMA
Adverse event reports
6
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Nitrophenylphosphate, Alkaline Phosphatase Or Isoenzymes
- Product code
- CJE
- Device class
- Class II
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.1050
- Specialty
- Laboratory Medicine & Pathology
- FDA record
- Classification database entry for CJE
Latest FDA decisions
K251074 · Decided Dec 19, 2025 · 255 days after receipt
Device type: Nitrophenylphosphate, Alkaline Phosphatase Or Isoenzymes (CJE) · Laboratory Medicine & Pathology
- Alkaline Phosphatase510(k)
Abbott Laboratories Diagnostics Division
K240468 · Decided Oct 16, 2024 · 239 days after receipt
Device type: Nitrophenylphosphate, Alkaline Phosphatase Or Isoenzymes (CJE) · Laboratory Medicine & Pathology
- Alkaline Phosphatase2510(k)
Abbott Ireland Diagnostics Division
K223317 · Decided Jul 21, 2023 · 266 days after receipt
Device type: Nitrophenylphosphate, Alkaline Phosphatase Or Isoenzymes (CJE) · Laboratory Medicine & Pathology
4 more FDA decisions with a free account
Reports received
6
Received Nov 1, 2025 to Nov 19, 2026
Malfunction reports
6
100% of reports
Injury reports
0
0.0% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 2 | 0 | 0 | 0 |
| Jan 24 | 3 | 0 | 0 | 0 |
| Feb 24 | 0 | 0 | 0 | 0 |
| Mar 24 | 0 | 0 | 0 | 0 |
| Apr 24 | 4 | 0 | 0 | 0 |
| May 24 | 0 | 0 | 0 | 0 |
| Jun 24 | 0 | 0 | 0 | 0 |
| Jul 24 | 1 | 0 | 0 | 0 |
| Aug 24 | 0 | 0 | 0 | 0 |
| Sep 24 | 1 | 0 | 0 | 0 |
| Oct 24 | 0 | 0 | 0 | 0 |
| Nov 24 | 2 | 0 | 0 | 0 |
| Dec 24 | 0 | 0 | 0 | 0 |
| Jan 25 | 1 | 0 | 0 | 0 |
| Feb 25 | 1 | 0 | 0 | 0 |
| Mar 25 | 0 | 0 | 0 | 0 |
| Apr 25 | 0 | 0 | 0 | 0 |
| May 25 | 1 | 0 | 0 | 0 |
| Jun 25 | 0 | 0 | 0 | 0 |
| Jul 25 | 1 | 0 | 0 | 0 |
| Aug 25 | 1 | 0 | 0 | 0 |
| Sep 25 | 2 | 0 | 0 | 0 |
| Oct 25 | 1 | 0 | 0 | 0 |
| Nov 25 | 2 | 0 | 0 | 0 |
| Dec 25 | 0 | 0 | 0 | 0 |
| Jan 26 | 1 | 0 | 0 | 0 |
| Feb 26 | 1 | 0 | 0 | 0 |
| Mar 26 | 0 | 0 | 0 | 0 |
| Apr 26 | 0 | 0 | 0 | 0 |
| May 26 | 1 | 0 | 0 | 0 |
| Jun 26 | 1 | 0 | 0 | 0 |
| Jul 26 | 0 | 0 | 0 | 0 |
| Aug 26 | 0 | 0 | 0 | 0 |
| Sep 26 | 0 | 0 | 0 | 0 |
| Oct 26 | 0 | 0 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- High Test Results4
- Non Reproducible Results4
- Incorrect, Inadequate or Imprecise Result or Readings2
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions6
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Mckesson Medical-Surgical Inc. Corporate Office
Z-1835-2025 · Started Apr 24, 2025 · Posted by FDA May 27, 2025
Reason given: transit delays of certain cold chain products due to severe thunderstorms which may caused the products to be stored outside the recommended labeled storage conditions. This failure could increase the probability of an incorrect test result or a delayed test result.
Cause as FDA categorized it: Storage
- VITROS XT Chemistry Products ALKP SlidesOpen, Classified
Ortho-Clinical Diagnostics, INc.
Z-2624-2023 · Started Aug 5, 2023 · Posted by FDA Sep 21, 2023
Reason given: Samples containing high levels of ALKP (or an interfering substance) could deplete the available substrate in the reagent layer of a VITROS ALKP Slide and not post the substrate deleption (DP) code, which may result in the System reporting an incorrect value (lower than the actual value), which may result in missed or...
Cause as FDA categorized it: Under Investigation by firm
Z-0296-2014 · Started Jun 12, 2013 · Posted by FDA Nov 15, 2013
Reason given: Ortho Clinical Diagnostics is recalling Slide Dispense cartridges used with some VITROS Chemistry Product Slides due to slide dispense issues.
Cause as FDA categorized it: Equipment maintenance
19 more recall records with a free account
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.