Device type · BTT
Humidifier, Respiratory Gas, (Direct Patient Interface)
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
BTT
Anesthesiology
Device class
Class II
FDA decisions since 2015
39
510(k), De Novo and PMA
Adverse event reports
4,478
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Humidifier, Respiratory Gas, (Direct Patient Interface)
- Product code
- BTT
- Device class
- Class II
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.5450
- Specialty
- Anesthesiology & Respiratory, Pulmonology
- FDA record
- Classification database entry for BTT
Latest FDA decisions
The 8 newest of 39 since 2015.
Fisher and Paykel Healthcare Limited
K262268 · Decided Aug 27, 2026 · 52 days after receipt
Device type: Humidifier, Respiratory Gas, (Direct Patient Interface) (BTT) · Anesthesiology & Respiratory
- HVT 2.0510(k)
K254078 · Decided Apr 13, 2026 · 116 days after receipt
Device type: Humidifier, Respiratory Gas, (Direct Patient Interface) (BTT) · Anesthesiology & Respiratory
K251448 · Decided Jan 16, 2026 · 252 days after receipt
Device type: Humidifier, Respiratory Gas, (Direct Patient Interface) (BTT) · Anesthesiology & Respiratory
Laboratorios Biogalenic S.A. DE C.V.
K251419 · Decided Jan 14, 2026 · 252 days after receipt
Device type: Humidifier, Respiratory Gas, (Direct Patient Interface) (BTT) · Anesthesiology & Respiratory
Fisher & Paykel Healthcare, Ltd.
K253479 · Decided Dec 19, 2025 · 66 days after receipt
Device type: Humidifier, Respiratory Gas, (Direct Patient Interface) (BTT) · Anesthesiology & Respiratory
K250312 · Decided Jul 2, 2025 · 149 days after receipt
Device type: Humidifier, Respiratory Gas, (Direct Patient Interface) (BTT) · Anesthesiology & Respiratory
Fisher & Paykel Healthcare, Ltd.
K251611 · Decided Jun 25, 2025 · 29 days after receipt
Device type: Humidifier, Respiratory Gas, (Direct Patient Interface) (BTT) · Anesthesiology & Respiratory
Laboratorios Biogalenic S.A. DE C.V.
K242717 · Decided Jan 30, 2025 · 142 days after receipt
Device type: Humidifier, Respiratory Gas, (Direct Patient Interface) (BTT) · Anesthesiology & Respiratory
Reports received
4,478
Received Nov 1, 2025 to Nov 19, 2026. Includes 6 of another or unstated type.
Malfunction reports
4,445
99% of reports
Injury reports
23
0.5% of reports
Death reports
4
0.1% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 65 | 5 | 2 | 0 |
| Jan 24 | 28 | 1 | 3 | 0 |
| Feb 24 | 45 | 2 | 1 | 0 |
| Mar 24 | 41 | 0 | 0 | 0 |
| Apr 24 | 33 | 1 | 1 | 0 |
| May 24 | 36 | 4 | 2 | 0 |
| Jun 24 | 40 | 1 | 0 | 0 |
| Jul 24 | 47 | 2 | 0 | 0 |
| Aug 24 | 37 | 0 | 0 | 0 |
| Sep 24 | 28 | 4 | 3 | 0 |
| Oct 24 | 49 | 1 | 0 | 0 |
| Nov 24 | 36 | 4 | 0 | 0 |
| Dec 24 | 49 | 2 | 0 | 0 |
| Jan 25 | 48 | 0 | 0 | 0 |
| Feb 25 | 52 | 3 | 1 | 0 |
| Mar 25 | 51 | 2 | 0 | 0 |
| Apr 25 | 69 | 0 | 0 | 0 |
| May 25 | 63 | 2 | 0 | 0 |
| Jun 25 | 63 | 0 | 0 | 0 |
| Jul 25 | 64 | 0 | 1 | 0 |
| Aug 25 | 67 | 0 | 1 | 0 |
| Sep 25 | 78 | 4 | 3 | 0 |
| Oct 25 | 218 | 5 | 0 | 0 |
| Nov 25 | 477 | 4 | 0 | 0 |
| Dec 25 | 481 | 2 | 0 | 0 |
| Jan 26 | 389 | 0 | 0 | 0 |
| Feb 26 | 286 | 4 | 0 | 0 |
| Mar 26 | 349 | 1 | 1 | 0 |
| Apr 26 | 347 | 1 | 0 | 0 |
| May 26 | 227 | 0 | 0 | 0 |
| Jun 26 | 205 | 4 | 2 | 0 |
| Jul 26 | 249 | 3 | 0 | 1 |
| Aug 26 | 455 | 2 | 0 | 0 |
| Sep 26 | 648 | 1 | 0 | 4 |
| Oct 26 | 332 | 1 | 1 | 1 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Defective Alarm3,000
- Unexpected Shutdown214
- Detachment of Device or Device Component197
- No Audible Alarm134
- Insufficient Heating124
- Material Split, Cut or Torn114
- Fluid/Blood Leak87
- Failure to Power Up73
- Gas/Air Leak71
- Crack69
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions4,401
- Low Oxygen Saturation58
- Insufficient Information20
- Airway Obstruction5
- Bradycardia3
- Burn(s)3
- Superficial (First Degree) Burn3
- Blister2
- Partial thickness (Second Degree) Burn2
- Aspiration/Inhalation1
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-2991-2026 · Started Jul 13, 2026 · Posted by FDA Aug 21, 2026
Reason given: Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions,...
Cause as FDA categorized it: Under Investigation by firm
- Airvo 2, REF: PT101US, and myAirvo 2, REF: PT100US, used with Disinfection Kit User Manual, REF: UI-185043723Open, Classified
Fisher & Paykel Healthcare, Ltd.
Z-0158-2026 · Started Sep 18, 2025 · Posted by FDA Oct 15, 2025
Reason given: When unintentionally disconnected from power source, humidifier device (used to deliver high flow warmed and humidified respiratory gases) may sound audible alert for less than 120 seconds, which may delay user awareness to reconnect power, which could lead to oxygen desaturation, hypoxia. Disinfection kit user manual...
Cause as FDA categorized it: Nonconforming Material/Component
Fisher & Paykel Healthcare, Ltd.
Z-1823-2024 · Started Apr 2, 2024 · Posted by FDA May 15, 2024
Reason given: Humidifier devices, used to deliver high flow respiratory therapy to patients, manufactured before 14 August 2017, have a speaker configuration that may result in distorted, intermittent or inaudible alarm sound levels. In the absence of an audible alarm, if there is an interruption to therapy, a patient may...
Cause as FDA categorized it: Nonconforming Material/Component
- Low Flow Disposable Patient Circuit for the Precision Flow System, Product Code PF-DPC-LOWOpen, Classified
Z-1272-2023 · Started Dec 16, 2022 · Posted by FDA Mar 17, 2023
Reason given: There have been reports of excessive condensation associated with certain lots of the Low Flow Disposable Patient Circuits (DPCs) used with Precision Flow System which may lead to excessive rainout.
Cause as FDA categorized it: Nonconforming Material/Component
Z-1385-2022 · Started Jan 5, 2021
Reason given: When starting a humidifier, intended for respiratory gas conditioning during invasive and noninvasive mechanical ventilation, with an empty chamber, and water is not filled into the chamber, the low water level alarm will be delayed and medical staff are unaware that there is no water in the chamber.
Cause as FDA categorized it: Employee error
- HUDSON RCI COMFORT FLO Corrugated Humidification System, REF 2415, humidifier nebulizer kitTerminated
Z-1112-2020 · Started Dec 13, 2019 · Posted by FDA Feb 7, 2020
Reason given: There is the potential for water to flood the column and enter the circuit under circumstances where an abnormal pressure differential is created between the water bottle and the column during high flow oxygen therapy.
Cause as FDA categorized it: Device Design
Z-1113-2020 · Started Dec 13, 2019 · Posted by FDA Feb 7, 2020
Reason given: There is the potential for water to flood the column and enter the circuit under circumstances where an abnormal pressure differential is created between the water bottle and the column during high flow oxygen therapy.
Cause as FDA categorized it: Device Design
Z-1114-2020 · Started Dec 13, 2019 · Posted by FDA Feb 7, 2020
Reason given: There is the potential for water to flood the column and enter the circuit under circumstances where an abnormal pressure differential is created between the water bottle and the column during high flow oxygen therapy.
Cause as FDA categorized it: Device Design
Showing the 8 newest of 70 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.