Device type · BTR
Tube, Tracheal (W/Wo Connector)
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
BTR
Anesthesiology
Device class
Class II
FDA decisions since 2015
32
510(k), De Novo and PMA
Adverse event reports
1,623
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Tube, Tracheal (W/Wo Connector)
- Product code
- BTR
- Device class
- Class II
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.5730
- Specialty
- Anesthesiology & Respiratory
- FDA record
- Classification database entry for BTR
Latest FDA decisions
The 8 newest of 32 since 2015.
K251794 · Decided Sep 3, 2026 · 448 days after receipt
Device type: Tube, Tracheal (W/Wo Connector) (BTR) · Anesthesiology & Respiratory
- Spiro-VISTA510(k)
K253437 · Decided Jun 12, 2026 · 254 days after receipt
Device type: Tube, Tracheal (W/Wo Connector) (BTR) · Anesthesiology & Respiratory
K250173 · Decided Oct 6, 2025 · 258 days after receipt
Device type: Tube, Tracheal (W/Wo Connector) (BTR) · Anesthesiology & Respiratory
K243785 · Decided Sep 25, 2025 · 290 days after receipt
Device type: Tube, Tracheal (W/Wo Connector) (BTR) · Anesthesiology & Respiratory
Pipeline Medical Products, LLC
K243579 · Decided Aug 13, 2025 · 267 days after receipt
Device type: Tube, Tracheal (W/Wo Connector) (BTR) · Anesthesiology & Respiratory
K251499 · Decided Jun 11, 2025 · 27 days after receipt
Device type: Tube, Tracheal (W/Wo Connector) (BTR) · Anesthesiology & Respiratory
K250243 · Decided May 20, 2025 · 113 days after receipt
Device type: Tube, Tracheal (W/Wo Connector) (BTR) · Anesthesiology & Respiratory
- LMA Fastrach ETT SU510(k)
K242495 · Decided Apr 29, 2025 · 250 days after receipt
Device type: Tube, Tracheal (W/Wo Connector) (BTR) · Anesthesiology & Respiratory
Reports received
1,623
Received Nov 1, 2025 to Nov 19, 2026. Includes 5 of another or unstated type.
Malfunction reports
1,396
86% of reports
Injury reports
212
13% of reports
Death reports
10
0.6% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 82 | 9 | 4 | 0 |
| Jan 24 | 123 | 18 | 0 | 0 |
| Feb 24 | 85 | 20 | 0 | 0 |
| Mar 24 | 111 | 14 | 0 | 0 |
| Apr 24 | 133 | 18 | 2 | 0 |
| May 24 | 146 | 26 | 0 | 0 |
| Jun 24 | 113 | 17 | 0 | 1 |
| Jul 24 | 130 | 13 | 0 | 0 |
| Aug 24 | 149 | 17 | 3 | 0 |
| Sep 24 | 161 | 20 | 1 | 0 |
| Oct 24 | 101 | 11 | 1 | 0 |
| Nov 24 | 103 | 19 | 0 | 0 |
| Dec 24 | 103 | 18 | 1 | 0 |
| Jan 25 | 128 | 13 | 0 | 0 |
| Feb 25 | 111 | 8 | 3 | 0 |
| Mar 25 | 139 | 12 | 2 | 0 |
| Apr 25 | 155 | 12 | 0 | 0 |
| May 25 | 167 | 13 | 1 | 0 |
| Jun 25 | 132 | 13 | 1 | 1 |
| Jul 25 | 125 | 13 | 0 | 0 |
| Aug 25 | 133 | 15 | 1 | 0 |
| Sep 25 | 145 | 3 | 0 | 0 |
| Oct 25 | 124 | 13 | 0 | 0 |
| Nov 25 | 131 | 16 | 0 | 0 |
| Dec 25 | 159 | 26 | 4 | 0 |
| Jan 26 | 160 | 49 | 0 | 0 |
| Feb 26 | 107 | 21 | 2 | 0 |
| Mar 26 | 131 | 6 | 0 | 0 |
| Apr 26 | 116 | 16 | 1 | 0 |
| May 26 | 100 | 13 | 0 | 0 |
| Jun 26 | 110 | 13 | 1 | 0 |
| Jul 26 | 95 | 13 | 0 | 1 |
| Aug 26 | 62 | 13 | 0 | 0 |
| Sep 26 | 107 | 17 | 2 | 4 |
| Oct 26 | 118 | 9 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Medtronic995
- ICU Medical218
- Teleflex161
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Gas/Air Leak500
- Use of Device Problem448
- Disconnection125
- Material Split, Cut or Torn98
- Leak/Splash55
- Inflation Problem40
- Material Deformation40
- Defective Component32
- Deflation Problem26
- Material Integrity Problem25
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions1,011
- Unintended Extubation365
- Insufficient Information135
- Low Oxygen Saturation57
- Hypoventilation52
- Cardiac Arrest48
- Unspecified Tissue Injury24
- Aspiration/Inhalation17
- Hemorrhage/Blood Loss/Bleeding15
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available8
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
- SunMed Tracheal Tube, Size 2.5; Model Number: 1-7333-25Open, Classified
Z-3172-2026 · Started Jul 31, 2026 · Posted by FDA Sep 15, 2026
Reason given: The 6Fr suction catheter could not pass through the samples smoothly.
Cause as FDA categorized it: Process control
Z-1289-2025 · Started Feb 13, 2025 · Posted by FDA Mar 12, 2025
Reason given: Affected devices have a smaller diameter than expected and may potentially result in inadequate ventilation to the patient. Patient may experience hypoxia, underdose, or cardiopulmonary collapse.
Cause as FDA categorized it: Process change control
MEDLINE INDUSTRIES, LP - Northfield
Z-1876-2024 · Started Feb 26, 2024 · Posted by FDA May 28, 2024
Reason given: Complaints have been received that the device inflation tube detached and/or tore from the main tube, resulting in potential moisture buildup, loss of pressure, or inability to inflate. There were also reports the suction pump is difficult to connect or detaches during use.
Cause as FDA categorized it: Under Investigation by firm
Z-1827-2023 · Started May 25, 2023 · Posted by FDA Jun 29, 2023
Reason given: Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
Cause as FDA categorized it: Under Investigation by firm
Z-1828-2023 · Started May 25, 2023 · Posted by FDA Jun 29, 2023
Reason given: Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
Cause as FDA categorized it: Under Investigation by firm
Z-1829-2023 · Started May 25, 2023 · Posted by FDA Jun 29, 2023
Reason given: Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
Cause as FDA categorized it: Under Investigation by firm
Z-1830-2023 · Started May 25, 2023 · Posted by FDA Jun 29, 2023
Reason given: Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
Cause as FDA categorized it: Under Investigation by firm
Z-1831-2023 · Started May 25, 2023 · Posted by FDA Jun 29, 2023
Reason given: Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
Cause as FDA categorized it: Under Investigation by firm
Showing the 8 newest of 227 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.