Proclaim SCS Neurostimulation System

A PMA supplement is a change to a device that already has premarket approval, such as a new indication, a design change or a manufacturing change.

Last checked: Nov 20, 2026Latest source record: Nov 17, 2021 (this decision)Source: openFDAMethods & limitations

PMA supplement

Decided

Nov 17, 2021

Approved

Time to decision

169 days

Received Jun 1, 2021. Calendar days, including time FDA spent waiting on the applicant.

Pathway

PMA supplement

Normal 180 Day Track

Device class

Class III

The decision

FDA number
P010032-S176
Device name
Proclaim SCS Neurostimulation System
Applicant location
Plano, TX
Received
Jun 1, 2021
Decided
Nov 17, 2021
Decision
Approved
Supplement type
Normal 180 Day Track
Reason for supplement
Labeling Change - Indications/instructions/shelf life/tradename
Original approval
P010032

Device type

Device class
Class III

Every decision under product code LGW

FDA's approval statement

As written in FDA's record.

Approval for expanded magnetic resonance (MR) Conditional labeling for the Octrode 60cm Lead when implanted with the Proclaim XR IPGs (Models 3660 and 3662) and their associated accessories, including the port plug and lead anchors. This supplement requested expansion of the conditions of use within the MRI labeling to increase radiofrequency (RF) power limit from whole body SAR <=0.8 W/kg to normal operating mode (NOM) for scans using integrated body RF transmit coil with any receive coil, and also to expand implant locations eligible for MRI scans for lead tips in epidural space from T7 to T12 to C1 to S2.

Source documents

Other decisions for ABBOTT MEDICAL

Matched on the applicant name as FDA recorded it. A company that files under several names will have more.

Company page

5 more FDA decisions with a free account

For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.