PMA device
MiniMed 630G Insulin Pump, Guardian Link (3) Transmitter
Medtronic Minimed · Automated Insulin Dosing , Threshold Suspend (OZO, Class III) · First FDA decision Aug 10, 2016, newest Nov 20, 2018Diabetes
Last checked: Nov 20, 2026Source: FDA, CMS, ClinicalTrials.gov, PubMed
FDA decisions
5
PMA pathway
Device trials
0
Naming "MiniMed 630G"
MAUDE reports
30
Brand "MiniMed 630G", since 2015
PubMed studies
3
Mentioning "MiniMed 630G"
AI summary across all sources
Written by AI from the FDA, CMS, ClinicalTrials.gov, MAUDE and PubMed data on this page, plus web searches for payment facts.
Product review deck
A slide deck reviewing MiniMed 630G Insulin Pump, Guardian Link (3) Transmitter
About ten slides on the technology, clinical evidence, regulatory history and commercialization, built with AI from the data on this page and web research, with sources. Download it as PowerPoint to edit or as PDF to share.
- Executive summary and key facts
- Technology
- Clinical evidence and trials
- Regulatory history and safety
- Reimbursement, market and competition
- Risks, outlook and sources
AI-written from public data and may contain mistakes. Check each fact against its source before relying on it. Not investment, legal, reimbursement or medical advice.
FDA decisions
The newest 5 of 5 decisions filed under this name.
- MiniMed 630G Insulin Pump, Guardian Link (3) TransmitterPMA supplement
P150001-S050 · Decided Nov 20, 2018 · 120 days after receipt · Normal 180 Day Track
Device type: Automated insulin dosing , threshold suspend (OZO) · Laboratory Medicine & Pathology
- MiniMed 630G Insulin PumpPMA supplement
P150001-S028 · Decided Dec 20, 2017 · 51 days after receipt · Normal 180 Day Track
Device type: Automated insulin dosing , threshold suspend (OZO) · Laboratory Medicine & Pathology
- MiniMed 630G SystemPMA supplement
P150001-S021 · Decided Feb 13, 2018 · 195 days after receipt · Panel Track
Device type: Automated insulin dosing , threshold suspend (OZO) · Laboratory Medicine & Pathology
2 more FDA decisions with a free account
Medicare coverage
National and local coverage documents whose titles match this device or its device type. Payment add-ons (NTAP, pass-through, TCET) and likely codes are covered in the summary above.
No Medicare coverage document has a title matching this device or device type. Many devices are paid under existing codes without one.
Clinical trials
Device studies on ClinicalTrials.gov that name "MiniMed 630G", any status and sponsor, newest update first.
No registered device study names "MiniMed 630G".
Adverse event reports (MAUDE)
Reports received since 2015 naming the brand "MiniMed 630G", and for the whole product code for context.
This brand
30 reports: 11 malfunction, 17 injury, 2 death
Whole product code OZO
424,597 reports: 337,031 malfunction, 86,136 injury, 1,411 death
Newest reports for this brand
- MINIMED 630GInjury
MW5158339 · Received Aug 12, 2024 · Event dated Aug 7, 2024
Device type: Automated Insulin Dosing , Threshold Suspend (OZO)
Patient problem as coded: Hyperglycemia
MW5118970 · Received Jun 28, 2023 · Event dated Jun 26, 2023
Device type: Automated Insulin Dosing Device System, Single Hormonal Control (OZP)
Device problem as coded: Crack; Positioning Failure
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- MINIMED 630GMalfunction
MW5116449 · Received Apr 6, 2023 · Event dated Apr 3, 2023
Device type: Automated Insulin Dosing , Threshold Suspend (OZO)
Device problem as coded: Fitting Problem
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
2 more reports with a free account
Recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-1740-2026 · Started Feb 13, 2026 · Posted by FDA Apr 2, 2026
Reason given: All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and...
Cause as FDA categorized it: Device Design · Device type: Automated Insulin Dosing , Threshold Suspend (OZO)
Z-1259-2025 · Started Jan 31, 2025 · Posted by FDA Mar 3, 2025
Reason given: All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can...
Cause as FDA categorized it: Device Design · Device type: Automated Insulin Dosing , Threshold Suspend (OZO)
Z-0001-2025 · Started Jul 31, 2024 · Posted by FDA Oct 3, 2024
Reason given: Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user...
Cause as FDA categorized it: Device Design · Device type: Automated Insulin Dosing , Threshold Suspend (OZO)
7 more recall records with a free account
Published research
The most relevant PubMed records for "MiniMed 630G"[tiab]. The summary above synthesizes these.
- Insulin Pump Alarms During Adverse Events: A Qualitative Descriptive Study.
Estock JL, Codario RA, Zupa MF, et al. · J Diabetes Sci Technol · 2025Research Support, U.S. Gov't, Non-P.H.S.
- Insulin Pump-Associated Adverse Events: A Qualitative Descriptive Study of Clinical Consequences and Potential Root Causes.
Estock JL, Codario RA, Keddem S, et al. · Diabetes Technol Ther · 2023Research Support, Non-U.S. Gov't
- Severe Insulin Pump-Related Adverse Events: Potential Root Causes and Impact of the COVID-19 Pandemic.
Orbell SL, Codario RA, Zupa MF, et al. · J Diabetes Sci Technol · 2025Research Support, N.I.H., Extramural, Research Support, U.S. Gov't, P.H.S.
Other devices of this type
Also filed under product code OZO.
- Minimed 530G SystemMedtronic, Inc. · PMA 2017
- t:slim X2 Insulin Pump with Basal-IQ TechnologyTandem Diabetes Care, Inc. · PMA 2018
- Enlite SensorMedtronic, Inc. · PMA 2015
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.