PMA device
Minimed 530G System
Medtronic, Inc. · Automated Insulin Dosing , Threshold Suspend (OZO, Class III) · First FDA decision Sep 26, 2013, newest Oct 11, 2017Diabetes
Last checked: Nov 20, 2026Source: FDA, CMS, ClinicalTrials.gov, PubMed
FDA decisions
7
PMA pathway
Device trials
1
Naming "Minimed 530G"
MAUDE reports
11
Brand "Minimed 530G", since 2015
PubMed studies
6
Mentioning "Minimed 530G"
AI summary across all sources
Written by AI from the FDA, CMS, ClinicalTrials.gov, MAUDE and PubMed data on this page, plus web searches for payment facts.
Product review deck
A slide deck reviewing Minimed 530G System
About ten slides on the technology, clinical evidence, regulatory history and commercialization, built with AI from the data on this page and web research, with sources. Download it as PowerPoint to edit or as PDF to share.
- Executive summary and key facts
- Technology
- Clinical evidence and trials
- Regulatory history and safety
- Reimbursement, market and competition
- Risks, outlook and sources
AI-written from public data and may contain mistakes. Check each fact against its source before relying on it. Not investment, legal, reimbursement or medical advice.
FDA decisions
The newest 6 of 7 decisions filed under this name.
- MINIMED 530G SYSTEMPMA supplement
P120010-S101 · Decided Oct 11, 2017 · 250 days after receipt · Normal 180 Day Track No User Fee
Device type: Automated insulin dosing , threshold suspend (OZO) · Laboratory Medicine & Pathology
- MINIMED 530G SYSTEMPMA supplement
P120010-S071 · Decided Sep 18, 2015 · 60 days after receipt · Normal 180 Day Track No User Fee
Device type: Automated insulin dosing , threshold suspend (OZO) · Laboratory Medicine & Pathology
- MINIMED 530G SYSTEMPMA supplement
P120010-S070 · Decided Feb 18, 2016 · 254 days after receipt · Normal 180 Day Track
Device type: Automated insulin dosing , threshold suspend (OZO) · Laboratory Medicine & Pathology
3 more FDA decisions with a free account
Older decisions: P120010
Medicare coverage
National and local coverage documents whose titles match this device or its device type. Payment add-ons (NTAP, pass-through, TCET) and likely codes are covered in the summary above.
No Medicare coverage document has a title matching this device or device type. Many devices are paid under existing codes without one.
Clinical trials
Device studies on ClinicalTrials.gov that name "Minimed 530G", any status and sponsor, newest update first.
Medtronic MiniMed, Inc.Public (MDT)
NCT02003898 · 372 participants · Start Nov 26, 2013 · Primary completion Feb 20, 2018
Device: Medtronic MiniMed 530G Insulin Pump · For: Diabetes
Sites: Little Rock, Arkansas; Escondido, California; Fresno, California · 43 sites in all
Adverse event reports (MAUDE)
Reports received since 2015 naming the brand "Minimed 530G", and for the whole product code for context.
This brand
11 reports: 4 malfunction, 7 injury, 0 death
Whole product code OZO
424,597 reports: 337,031 malfunction, 86,136 injury, 1,411 death
Newest reports for this brand
- MINIMED 530GInjury
MW5094419 · Received May 7, 2020 · Event dated May 4, 2020 · 1 other device named on this report
Device type: Automated Insulin Dosing , Threshold Suspend (OZO)
Device problem as coded: Fluid/Blood Leak
Patient problem as coded: Hyperglycemia
- MINIMED 530G W/ CGMInjury
MW5072545 · Received Oct 2, 2017
Device type: Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor (OYC)
Device problem as coded: Nonstandard Device; Difficult to Remove
Patient problem as coded: Test Result
- MINIMED 530G WITH ENLITEInjury
MW5072532 · Received Sep 29, 2017 · Event dated Apr 22, 2017
Device type: Automated Insulin Dosing , Threshold Suspend (OZO)
Device problem as coded: Excess Flow or Over-Infusion; Nonstandard Device
Patient problem as coded: Fall; Hypoglycemia; Injury
2 more reports with a free account
Recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
- MiniMed 530G Insulin Pump (MMT-551, MMT-751)Open, Classified
Z-1751-2026 · Started Feb 13, 2026 · Posted by FDA Apr 2, 2026
Reason given: All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and...
Cause as FDA categorized it: Device Design · Device type: Automated Insulin Dosing , Threshold Suspend (OZO)
Z-2544-2014 · Started Mar 13, 2014 · Posted by FDA Aug 22, 2014
Reason given: Medtronic is recalling the Medtronic MiniMed Paradigm Insulin Infusion Pumps because they have received reports that users have accidentally programmed the pump to deliver the maximum bolus amount.
Cause as FDA categorized it: Device Design · Device type: Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor (OYC)
Published research
The most relevant PubMed records for "Minimed 530G"[tiab]. The summary above synthesizes these.
- MiniMed 530G: an insulin pump with low-glucose suspend automation.
JAMA · 2015
- Design and analysis of crossover trials for investigating high-risk medical devices: A review.
Zhang B, Guo J, Zhang H · Contemp Clin Trials Commun · 2022Review
- MiniMed 530G: an insulin pump with low-glucose suspend automation.
Med Lett Drugs Ther · 2013
- Day and Night Closed-Loop Control Using the Integrated Medtronic Hybrid Closed-Loop System in Type 1 Diabetes at Diabetes Camp.
Ly TT, Roy A, Grosman B, et al. · Diabetes Care · 2015Controlled Clinical Trial, Randomized Controlled Trial
- Retrospective analysis of the real-world use of the threshold suspend feature of sensor-augmented insulin pumps.
Agrawal P, Zhong A, Welsh JB, et al. · Diabetes Technol Ther · 2015Evaluation Study
- Effectiveness of Automated Insulin Management Features of the MiniMed® 640G Sensor-Augmented Insulin Pump.
Zhong A, Choudhary P, McMahon C, et al. · Diabetes Technol Ther · 2016Research Support, Non-U.S. Gov't
Other devices of this type
Also filed under product code OZO.
- MiniMed 630G Insulin Pump, Guardian Link (3) TransmitterMedtronic Minimed · PMA 2018
- t:slim X2 Insulin Pump with Basal-IQ TechnologyTandem Diabetes Care, Inc. · PMA 2018
- Enlite SensorMedtronic, Inc. · PMA 2015
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.