PMA device
CardioMEMS HF System
Abbott Medical · System, Hemodynamic, Implantable (MOM, Class III) · First FDA decision Aug 29, 2014, newest Jun 1, 2026Heart failure
Last checked: Nov 20, 2026Source: FDA, CMS, ClinicalTrials.gov, PubMed
FDA decisions
13
PMA pathway
Device trials
20
Naming "CardioMEMS HF"
MAUDE reports
18
Brand "CardioMEMS HF", since 2015
PubMed studies
57
Mentioning "CardioMEMS HF"
AI summary across all sources
Written by AI from the FDA, CMS, ClinicalTrials.gov, MAUDE and PubMed data on this page, plus web searches for payment facts.
Product review deck
A slide deck reviewing CardioMEMS HF System
About ten slides on the technology, clinical evidence, regulatory history and commercialization, built with AI from the data on this page and web research, with sources. Download it as PowerPoint to edit or as PDF to share.
- Executive summary and key facts
- Technology
- Clinical evidence and trials
- Regulatory history and safety
- Reimbursement, market and competition
- Risks, outlook and sources
AI-written from public data and may contain mistakes. Check each fact against its source before relying on it. Not investment, legal, reimbursement or medical advice.
FDA decisions
The newest 6 of 13 decisions filed under this name.
- CardioMEMS HF SystemPMA supplement
P100045-S090 · Decided Jun 1, 2026 · 42 days after receipt · Normal 180 Day Track No User Fee
Device type: System, hemodynamic, implantable (MOM) · Cardiology
- CardioMEMS HF SystemPMA supplement
P100045-S077 · Decided Feb 6, 2025 · 216 days after receipt · Normal 180 Day Track
Device type: System, hemodynamic, implantable (MOM) · Cardiology
- CardioMEMS HF SystemPMA supplement
P100045-S076 · Decided Aug 29, 2024 · 136 days after receipt · Normal 180 Day Track No User Fee
Device type: System, hemodynamic, implantable (MOM) · Cardiology
3 more FDA decisions with a free account
Older decisions: P100045-S056, P100045-S053, P100045-S052, P100045-S050, P100045-S049, P100045-S018, P100045-S001
Medicare coverage
National and local coverage documents whose titles match this device or its device type. Payment add-ons (NTAP, pass-through, TCET) and likely codes are covered in the summary above.
No Medicare coverage document has a title matching this device or device type. Many devices are paid under existing codes without one.
Clinical trials
Device studies on ClinicalTrials.gov that name "CardioMEMS HF", any status and sponsor, newest update first.
- Trial to Evaluate Safety And Effectiveness of Mechanical Circulatory Support in Patients With Advancing Heart FailurePivotal-lookingRecruiting
Abbott Medical DevicesPublic (ABT)
NCT06526195 · 850 participants (planned) · Start Dec 13, 2024 · Primary completion Sep 2029
Device: CardioMEMS HF System; HeartMate 3 Left Ventricular Assist System · For: Heart Failure; Heart Diseases; Cardiovascular Diseases and 1 more
Sites: Phoenix, Arizona; Little Rock, Arkansas; Sacramento, California · 70 sites in all
Centre Hospitalier Universitaire de NīmesAcademic, hospital or other
NCT05428631 · 4 participants · Start Aug 2, 2022 · Primary completion Oct 25, 2023
Device: Implantation of the CARDIOMEMS™ HF device · For: Heart Failure; Renal Insufficiency
Sites: Nîmes, France; Montpellier, France
Sheffield Teaching Hospitals NHS Foundation TrustAcademic, hospital or other
NCT04078243 · 120 participants (planned) · Start Jan 21, 2020 · Primary completion Jan 6, 2028
Device: CardioMEMS HF System · For: Pulmonary Hypertension
Sites: Sheffield, United Kingdom
17 more trials with a free account
Adverse event reports (MAUDE)
Reports received since 2015 naming the brand "CardioMEMS HF", and for the whole product code for context.
This brand
18 reports: 1 malfunction, 9 injury, 7 death
Whole product code MOM
12,295 reports: 7,687 malfunction, 4,377 injury, 226 death
Newest reports for this brand
3004936110-2025-01729 · Received Sep 23, 2025 · Event dated Sep 11, 2025
Device type: System, Hemodynamic, Implantable (MOM)
Device problem as coded: Difficult to Advance
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
- CARDIOMEMS HF SYSTEMMalfunction
MW5162562 · Received Nov 14, 2024
Device type: System, Hemodynamic, Implantable (MOM)
Patient problem as coded: Insufficient Information
- CARDIOMEMS HF SYSTEMInjury
5488421 · Received Mar 9, 2016 · Event dated Jan 19, 2016
Device type: System, Hemodynamic, Implantable (MOM)
Device problem as coded: Device Operates Differently Than Expected
Patient problem as coded: No Known Impact Or Consequence To Patient
2 more reports with a free account
Recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-1427-2025 · Started Feb 3, 2025 · Posted by FDA Mar 21, 2025
Reason given: Due to issues associated with data migration of patient information and results to a Cloud service, the firm has received complaint associated with some patient profile information being duplicated while other patient information is missing.
Cause as FDA categorized it: Software design · Device type: System, Hemodynamic, Implantable (MOM)
Z-0987-2025 · Started Nov 6, 2024 · Posted by FDA Jan 17, 2025
Reason given: Due to a data migration from on-premises server to the cloud environment, some customers reported that they were not able to transmit readings.
Cause as FDA categorized it: Under Investigation by firm · Device type: System, Hemodynamic, Implantable (MOM)
Z-0988-2025 · Started Oct 5, 2024 · Posted by FDA Jan 17, 2025
Reason given: As a result of a cloud data migration of Patient and Hospital Electronic Systems (PES and HES) reverted to an outdated configuration causing incorrect/inaccurate pulmonary artery (PA) pressure readings.
Cause as FDA categorized it: Under Investigation by firm · Device type: System, Hemodynamic, Implantable (MOM)
10 more recall records with a free account
Published research
The most relevant PubMed records for "CardioMEMS HF"[tiab]. The summary above synthesizes these.
- CardioMEMS: where we are and where can we go?
Pour-Ghaz I, Hana D, Raja J, et al. · Ann Transl Med · 2019Review
- CardioMEMS HF for the Management of Heart Failure-Effectiveness and Value.
Ollendorf DA, Sandhu AT, Pearson SD · JAMA Intern Med · 2016Research Support, Non-U.S. Gov't, Review
- CardioMEMSࡊ: a tool for remote hemodynamic monitoring of chronic heart failure patients.
Radhoe SP, Brugts JJ · Future Cardiol · 2022Review
- A profile on the CardioMEMS HF system in the management of patients with early stages of heart failure: an update.
Vyas R, Patel M, Khouri SJ, et al. · Expert Rev Med Devices · 2023Review
- Concurrent Assessment of the CardioMEMS HF System and HeartLogic HF Diagnostic: A Retrospective Case Series.
Chilcote JL, Summers RP, Vaz DG, et al. · J Card Fail · 2022
- An update on the CardioMEMS pulmonary artery pressure sensor.
Ayyadurai P, Alkhawam H, Saad M, et al. · Ther Adv Cardiovasc Dis · 2019Review
- Understanding adherence to the CardioMEMS pulmonary artery pressure monitoring system for heart failure: A qualitative study.
Haynes SC, Pallin R, Tong K, et al. · Heart Lung · 2020
- Use of the CardioMEMS Device in Children and Patients with Congenital Heart Disease: A Literature Review.
Piccinelli E, Grutter G, Pilati M, et al. · J Clin Med · 2024Review
- Hemodynamic monitoring with the CardioMEMS HF System.
Falls C · Nurse Pract · 2021
- Noninvasive assessment of pulmonary congestion in heart failure: Need of the hour.
Deora S, Choudhary R, Kaushik A, et al. · Indian Heart J · 2023Review
- Postmarketing Adverse Events Related to the CardioMEMS HF System.
Vaduganathan M, DeFilippis EM, Fonarow GC, et al. · JAMA Cardiol · 2017Research Support, N.I.H., Extramural, Research Support, Non-U.S. Gov't
- Large pulmonary artery pseudoaneurysm after CardioMEMS implantation: a case report.
Garg T, Raikhelkar J, Gilkeson R, et al. · Eur Heart J Case Rep · 2022Case Reports
Other devices of this type
Also filed under product code MOM.
- Cordella Pulmonary Artery Sensor SystemEndotronix, Inc. · PMA 2026
- Cardiomems HF Pressure Measurement SystemAbbott Medical · PMA 2016
- Next Gen Patient System CM1200 CardioMEMS HERO DeviceAbbott Medical · PMA 2026
- CorPASS SystemEndotronix, Inc. · PMA 2025
- CardioMEMS patient and hospital electronicsAbbott Medical · PMA 2018
- CardiomemsAbbott Medical · PMA 2016
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.