PMA device
SynCardia temporary Total Artificial Heart (TAH-t) System
Syncardia Systems, LLC · Artificial Heart (LOZ, Class III) · First FDA decision Apr 28, 2010, newest Jul 6, 2023Heart failure
Last checked: Nov 20, 2026Source: FDA, CMS, ClinicalTrials.gov, PubMed
FDA decisions
11
PMA pathway
Device trials
4
Naming "SynCardia temporary"
MAUDE reports
11
Brand "SynCardia temporary", since 2015
PubMed studies
11
Mentioning "SynCardia temporary"
AI summary across all sources
Written by AI from the FDA, CMS, ClinicalTrials.gov, MAUDE and PubMed data on this page, plus web searches for payment facts.
Product review deck
A slide deck reviewing SynCardia temporary Total Artificial Heart (TAH-t) System
About ten slides on the technology, clinical evidence, regulatory history and commercialization, built with AI from the data on this page and web research, with sources. Download it as PowerPoint to edit or as PDF to share.
- Executive summary and key facts
- Technology
- Clinical evidence and trials
- Regulatory history and safety
- Reimbursement, market and competition
- Risks, outlook and sources
AI-written from public data and may contain mistakes. Check each fact against its source before relying on it. Not investment, legal, reimbursement or medical advice.
FDA decisions
The newest 6 of 11 decisions filed under this name.
P030011-S064 · Decided Jul 6, 2023 · 1,756 days after receipt · Normal 180 Day Track No User Fee
Device type: Artificial heart (LOZ) · Cardiology
P030011-S080 · Decided Nov 22, 2022 · 154 days after receipt · Normal 180 Day Track
Device type: Artificial heart (LOZ) · Cardiology
P030011-S070 · Decided Mar 5, 2020 · 276 days after receipt · Normal 180 Day Track
Device type: Artificial heart (LOZ) · Cardiology
3 more FDA decisions with a free account
Older decisions: P030011-S021, P030011-S020, P030011-S011, P030011-S008, P030011-S007
Medicare coverage
National and local coverage documents whose titles match this device or its device type. Payment add-ons (NTAP, pass-through, TCET) and likely codes are covered in the summary above.
- Artificial Hearts and related devices, including Ventricular Assist Devices for Bridge-to-Transplant and Destination TherapyNational analysisReview closedCAG-00453NUpdated Nov 19, 2026
- Artificial HeartsNational analysisReview closedCAG-00322NUpdated Sep 16, 2026
- Artificial Hearts and Related Devices - RETIREDNational determinationOn the books20.9Updated Jun 5, 2024
Clinical trials
Device studies on ClinicalTrials.gov that name "SynCardia temporary", any status and sponsor, newest update first.
- SynCardia 70cc TAH-t for Destination Therapy (DT)Status unknown
SynCardia Systems. LLCNot on our public-company list
NCT02232659 · 38 participants (planned) · Start Feb 2016 · Primary completion Dec 2021
Device: 70cc TAH-t for Destination Therapy · For: Life-threatening; Biventricular Heart Failure; Ineligible for Cardiac Transplant
Sites: Phoenix, Arizona; Tucson, Arizona; Los Angeles, California · 7 sites in all
SynCardia Systems. LLCNot on our public-company list
NCT02459054 · 16 participants · Start Oct 2015 · Primary completion Jul 29, 2020
Device: SynCardia 50cc temporary Total Artificial Heart (TAH-t) · For: Heart Failure, Right-Sided; Cardiac Failure; Heart Failure, Left-Sided
Sites: Tucson, Arizona; Los Angeles, California; Gainesville, Florida · 26 sites in all
- SynCardia Freedom Driver System StudyApproved for marketing
SynCardia Systems. LLCNot on our public-company list
NCT00733447 · Start Mar 2010 · Primary completion Aug 2013
Device: CardioWest temporary Total Artificial Heart (TAH-t) System · For: Biventricular Failure
Sites: Phoenix, Arizona; Philadelphia, Pennsylvania; Richmond, Virginia
1 more trials with a free account
Adverse event reports (MAUDE)
Reports received since 2015 naming the brand "SynCardia temporary", and for the whole product code for context.
This brand
11 reports: 2 malfunction, 5 injury, 4 death
Whole product code LOZ
3,209 reports: 2,545 malfunction, 279 injury, 384 death
Newest reports for this brand
3003761017-2022-00103 · Received Oct 5, 2022 · Event dated Sep 8, 2022
Device type: Artificial Heart (LOZ)
Device problem as coded: Adverse Event Without Identified Device or Use Problem
Patient problem as coded: Suicidal Ideation
SYNCARDIA TEMPORARY TOTAL ARTIFICIAL HEART (TAH-T)
3003761017-2022-00104 · Received Oct 5, 2022 · Event dated Sep 9, 2022
Device type: Artificial Heart (LOZ)
Device problem as coded: Adverse Event Without Identified Device or Use Problem
Patient problem as coded: Multiple Organ Dysfunction Syndrome
3003761017-2022-00102 · Received Sep 29, 2022 · Event dated Sep 3, 2022
Device type: Artificial Heart (LOZ)
Device problem as coded: Adverse Event Without Identified Device or Use Problem
Patient problem as coded: Multiple Organ Dysfunction Syndrome
2 more reports with a free account
Recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-0126-2024 · Started Aug 1, 2023 · Posted by FDA Oct 20, 2023
Reason given: Artificial hearts contain a epoxy resin that has not been reviewed by the FDA for safety and efficacy.
Cause as FDA categorized it: Other · Device type: Artificial Heart (LOZ)
- SynCardia Total Artificial Heart System (70CC): 500101-001 SynCardia Total Artificial Heart System (50CC): 570500-000Open, Classified
Z-1022-2023 · Started Nov 2, 2022 · Posted by FDA Jan 27, 2023
Reason given: Due to tears reported tears in the cannula due to wear and tear stress.
Cause as FDA categorized it: Device Design · Device type: Artificial Heart (LOZ)
Z-0659-2017 · Started Oct 21, 2016
Reason given: The Main Printed Circuit Board Assembly (PCBA) of the affected Freedom Drivers may fail and cause the Freedom Driver to stop functioning without visual or audible alarms, resulting in the loss of life-sustaining function.
Cause as FDA categorized it: Process control · Device type: Artificial Heart (LOZ)
2 more recall records with a free account
Published research
The most relevant PubMed records for "SynCardia temporary"[tiab]. The summary above synthesizes these.
- SynCardia Temporary Total Artificial Heart: Single-Center Experience at a Children's Hospital.
Perry T, Morales DLS, Villa CR, et al. · ASAIO J · 2022
- SynCardia temporary total artificial heart as bridge to transplantation: current results at la pitié hospital.
Kirsch ME, Nguyen A, Mastroianni C, et al. · Ann Thorac Surg · 2013
- Experience With SynCardia Total Artificial Heart as a Bridge to Transplantation in 100 Patients.
Malas J, Chen Q, Akhmerov A, et al. · Ann Thorac Surg · 2023
- The Total Artificial Heart: Where Are We?
Arabía FA · Cardiol Rev · 2020
- Troubleshooting Total Artificial Heart: Novel Use of Implantable Hemodynamic Monitor.
Chau VQ, Oliveros E, Mahmood K, et al. · JACC Case Rep · 2021Case Reports
- Outcomes after SynCardia® temporary total artificial heart implantation: A 20-year single-center experience in 196 patients.
Razumov A, Burri M, Zittermann A, et al. · Artif Organs · 2025
- The total artificial heart in pediatrics: outcomes in an evolving field.
Villa CR, Moore RA, Morales DL, et al. · Ann Cardiothorac Surg · 2020
- A heart transplant after total artificial heart support: initial and long-term results.
David CH, Lacoste P, Nanjaiah P, et al. · Eur J Cardiothorac Surg · 2020
- Status, Indications, and Use of Cardiac Replacement Therapy in the Era of Multimodal Mechanical Approaches to Circulatory Support: A Scoping Review.
Noly PE, Ben Ali W, Lamarche Y, et al. · Can J Cardiol · 2020Scoping Review
- Bridge to transplantation using paracorporeal biventricular assist devices or the syncardia temporary total artificial heart: is there a difference?
Nguyen A, Pozzi M, Mastroianni C, et al. · J Cardiovasc Surg (Torino) · 2015Comparative Study, Multicenter Study
- A device profile of the SynCardia temporary total artificial heart as a bridge-to-transplantation in patients with biventricular failure.
Rubin G, Colón MJ, Mahar J, et al. · Expert Rev Med Devices · 2026Review
Other devices of this type
Also filed under product code LOZ.
- Total Artificial HeartSyncardia Systems, LLC · PMA 2025
- Syncardia Companion 2 Driver SystemSyncardia Systems, LLC · PMA 2012
- Cardiowest Temporary Total Artifical Heart (Tah-T) SystemSyncardia Systems, LLC · PMA 2008
- Syncardia Temporary Cardiowest Total Artificial Heart (Tah-T)Syncardia Systems, LLC · PMA 2005
- Syncardia Temporary Cardio West Total Artificial Heart (Tah-T)Syncardia Systems, LLC · PMA 2004
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.