Device type · LOZ
Artificial Heart
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
LOZ
Cardiovascular
Device class
Class III
FDA decisions since 2015
0
510(k), De Novo and PMA
Adverse event reports
184
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Artificial Heart
- Product code
- LOZ
- Device class
- Class III
- Review panel
- Cardiovascular
- Specialty
- Cardiology
- FDA record
- Classification database entry for LOZ
Latest FDA decisions
No 510(k), De Novo or PMA decisions for this product code since 2015.
Reports received
184
Received Nov 1, 2025 to Nov 19, 2026
Malfunction reports
154
84% of reports
Injury reports
5
2.7% of reports
Death reports
25
14% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 8 | 0 | 3 | 0 |
| Jan 24 | 3 | 0 | 1 | 0 |
| Feb 24 | 8 | 0 | 1 | 0 |
| Mar 24 | 5 | 0 | 3 | 0 |
| Apr 24 | 4 | 0 | 1 | 0 |
| May 24 | 6 | 0 | 1 | 0 |
| Jun 24 | 9 | 0 | 1 | 0 |
| Jul 24 | 3 | 0 | 1 | 0 |
| Aug 24 | 10 | 0 | 0 | 0 |
| Sep 24 | 6 | 0 | 1 | 0 |
| Oct 24 | 3 | 0 | 3 | 0 |
| Nov 24 | 3 | 0 | 1 | 0 |
| Dec 24 | 4 | 0 | 2 | 0 |
| Jan 25 | 9 | 0 | 1 | 0 |
| Feb 25 | 4 | 0 | 1 | 0 |
| Mar 25 | 5 | 0 | 0 | 0 |
| Apr 25 | 0 | 0 | 0 | 0 |
| May 25 | 0 | 0 | 0 | 0 |
| Jun 25 | 6 | 0 | 0 | 0 |
| Jul 25 | 1 | 0 | 0 | 0 |
| Aug 25 | 5 | 0 | 0 | 0 |
| Sep 25 | 3 | 0 | 0 | 0 |
| Oct 25 | 3 | 0 | 0 | 0 |
| Nov 25 | 0 | 0 | 0 | 0 |
| Dec 25 | 8 | 1 | 0 | 0 |
| Jan 26 | 0 | 0 | 0 | 0 |
| Feb 26 | 9 | 0 | 0 | 0 |
| Mar 26 | 3 | 0 | 0 | 0 |
| Apr 26 | 0 | 0 | 0 | 0 |
| May 26 | 0 | 0 | 1 | 0 |
| Jun 26 | 0 | 0 | 0 | 0 |
| Jul 26 | 21 | 1 | 3 | 0 |
| Aug 26 | 12 | 2 | 21 | 0 |
| Sep 26 | 59 | 1 | 0 | 0 |
| Oct 26 | 42 | 0 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Audible Prompt/Feedback Problem93
- Adverse Event Without Identified Device or Use Problem27
- Charging Problem14
- Output Problem11
- Material Integrity Problem9
- Noise, Audible8
- Battery Problem5
- Break5
- Display or Visual Feedback Problem5
- Mechanical Problem4
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions150
- Multiple Organ Dysfunction Syndrome12
- Hemorrhage/Blood Loss/Bleeding6
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available5
- Sepsis3
- Stroke/CVA3
- Anxiety2
- Brain Injury2
- Dizziness1
- Hemorrhagic Stroke1
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-0126-2024 · Started Aug 1, 2023 · Posted by FDA Oct 20, 2023
Reason given: Artificial hearts contain a epoxy resin that has not been reviewed by the FDA for safety and efficacy.
Cause as FDA categorized it: Other
- SynCardia Total Artificial Heart System (70CC): 500101-001 SynCardia Total Artificial Heart System (50CC): 570500-000Open, Classified
Z-1022-2023 · Started Nov 2, 2022 · Posted by FDA Jan 27, 2023
Reason given: Due to tears reported tears in the cannula due to wear and tear stress.
Cause as FDA categorized it: Device Design
Z-0659-2017 · Started Oct 21, 2016
Reason given: The Main Printed Circuit Board Assembly (PCBA) of the affected Freedom Drivers may fail and cause the Freedom Driver to stop functioning without visual or audible alarms, resulting in the loss of life-sustaining function.
Cause as FDA categorized it: Process control
Z-2731-2015 · Started Aug 6, 2015 · Posted by FDA Sep 17, 2015
Reason given: SynCardia Systems, Inc. is recalling Freedom Drivers due to a specific component of the drive mechanism may fail and cause the drive mechanism to stop pumping; resulting in the loss of life-sustaining function.
Cause as FDA categorized it: Component design/selection
Z-2304-2008 · Started Jun 13, 2008 · Posted by FDA Sep 20, 2008
Reason given: Subassembly incorrectly aligned
Cause as FDA categorized it: Process control
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.