510(k) device
FEMTO LDV Z8 Femtosecond Surgical Laser
Sie Ag,Surgical Instrument Engineering · Ophthalmic Femtosecond Laser (OOE, Class II) · First FDA decision Oct 23, 2015, newest Apr 21, 2022Cataract
Last checked: Nov 20, 2026Source: FDA, CMS, ClinicalTrials.gov, PubMed
FDA decisions
2
510(k) pathway
Device trials
6
Naming "FEMTO LDV Z8"
MAUDE reports
11
Brand "FEMTO LDV Z8", since 2015
PubMed studies
30
Mentioning "FEMTO LDV Z8"
AI summary across all sources
Written by AI from the FDA, CMS, ClinicalTrials.gov, MAUDE and PubMed data on this page, plus web searches for payment facts.
Product review deck
A slide deck reviewing FEMTO LDV Z8 Femtosecond Surgical Laser
About ten slides on the technology, clinical evidence, regulatory history and commercialization, built with AI from the data on this page and web research, with sources. Download it as PowerPoint to edit or as PDF to share.
- Executive summary and key facts
- Technology, with pictures of the device
- Clinical evidence and trials
- Regulatory history and safety
- Reimbursement, market and competition
- Risks, outlook and sources
AI-written from public data and may contain mistakes. Check each fact against its source before relying on it. Not investment, legal, reimbursement or medical advice.
FDA decisions
The newest 2 of 2 decisions filed under this name.
K213559 · Decided Apr 21, 2022 · 168 days after receipt
Device type: Ophthalmic Femtosecond Laser (OOE) · Ophthalmology
1 more FDA decisions with a free account
Medicare coverage
National and local coverage documents whose titles match this device or its device type. Payment add-ons (NTAP, pass-through, TCET) and likely codes are covered in the summary above.
No Medicare coverage document has a title matching this device or device type. Many devices are paid under existing codes without one.
Clinical trials
Device studies on ClinicalTrials.gov that name "FEMTO LDV Z8", any status and sponsor, newest update first.
- Comparison of Femto-laser and Manually Created Clear Corneal Incisions During Cataract SurgeryCompleted
Ziemer Ophthalmic Systems AGNot on our public-company list
NCT04082273 · 80 participants · Start Aug 26, 2019 · Primary completion Jul 15, 2024
Device: Femtosecond Laser for Cataract Surgery · For: Cataract
Sites: Reinach, Switzerland
5 more trials with a free account
Adverse event reports (MAUDE)
Reports received since 2015 naming the brand "FEMTO LDV Z8", and for the whole product code for context.
This brand
11 reports: 0 malfunction, 11 injury, 0 death
Whole product code OOE
4,559 reports: 2,002 malfunction, 2,527 injury, 0 death
Newest reports for this brand
- FEMTO LDV Z8Injury
SIE AG, SURGICAL INSTRUMENT ENGINEERING
3005643720-2019-00001 · Received Feb 22, 2019 · Event dated Jan 25, 2019
Device type: Ophthalmic Femtosecond Laser (OOE)
Device problem as coded: Improper or Incorrect Procedure or Method
Patient problem as coded: Capsular Bag Tear
4 more reports with a free account
Recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
SIE AG, Surgical Instrument Engineering
Z-1272-2018 · Started Jan 21, 2018
Reason given: This Correction is being initiated based on a customer complaint received on November 16, 2017 which indicated a complication with a cataract procedure. An investigation completed by the manufacturer determined that although the complication was due to user error improvements could and should be made on our part to...
Cause as FDA categorized it: Device Design · Device type: Ophthalmic Femtosecond Laser (OOE)
Published research
The most relevant PubMed records for "FEMTO LDV Z8"[tiab]. The summary above synthesizes these.
- Keratorefractive Lenticule Extraction Using the Ziemer FEMTO LDV Z8 Platform (CLEAR): One-Year Results.
Bteich Y, Assaf JF, Gendy JE, et al. · J Refract Surg · 2024
- Efficacy of a new fragmentation pattern in femtosecond laser-assisted cataract surgery with the Ziemer FEMTO LDV Z8.
Boden KT, Julich-Haertel H, Rickmann A, et al. · Int Ophthalmol · 2023
- First Experience in Small Incision Lenticule Extraction with the Femto LDV Z8 and Lenticule Evaluation Using Scanning Electron Microscopy.
Wang M, Zhang F, Copruz CC, et al. · J Ophthalmol · 2020
- Novel method for preventing cyclorotation in Ziemer Femto LDV Z8 femtosecond laser-assisted cataract surgery with Verion image-guided system.
Lin HY, Chuang YJ, Lin PJ, et al. · Clin Ophthalmol · 2019
- [Results of corneal lenticule extraction for correction].
Doga AV, Kostenev SV, Mushkova IA, et al. · Vestn Oftalmol · 2020
- Intraoperative Optical Coherence Tomography (OCT)-Guided Femtosecond Laser-Assisted Descemet Membrane Endothelial Keratoplasty (iFAD).
Lim J, Abusayf MM, Liu YC, et al. · Bioengineering (Basel) · 2024
- Ex-vivo study on the surface quality of corneal lenticule and stroma after low energy femtosecond laser lenticule extraction.
Nanavaty MA, Naveed H, Ashena Z, et al. · Sci Rep · 2022Research Support, Non-U.S. Gov't
- Effectiveness of reducing corneal astigmatism after combined high-frequency LDV Z8 femtosecond laser-assisted phacoemulsification and arcuate keratotomy.
Lin HY, Chen S, Chuang YJ, et al. · Front Cell Dev Biol · 2022
- Safety and efficacy of cataract surgery performed with a low-energy femtosecond laser compared with conventional phacoemulsification in Chinese patients: a randomized clinical trial.
Zhou KJ, Huang Y, Wang Y, et al. · Eye Vis (Lond) · 2023
- Corneal lenticule extraction assisted by a low-energy femtosecond laser.
Izquierdo L, Sossa D, Ben-Shaul O, et al. · J Cataract Refract Surg · 2020
- Comparison of intraocular pressure during the application of a liquid patient interface (FEMTO LDV Z8) for femtosecond laser-assisted cataract surgery using two different vacuum levels.
Ebner M, Mariacher S, Januschowski K, et al. · Br J Ophthalmol · 2017Comparative Study, Research Support, Non-U.S. Gov't
- Impact of Patient Interface Diameter and Vacuum Level on Suction Stability Using a Flat Applanating Interface for Femtosecond Laser-Assisted LASIK.
Mariacher S, Mariacher M, Seuthe AM, et al. · Curr Eye Res · 2020Research Support, Non-U.S. Gov't
Other devices of this type
Also filed under product code OOE.
- LENSAR Laser System - fs 3D (LLS-fs 3D)Lensar, Inc. · 510(k) 2018
- VICTUS Femtosecond Laser PlatformTechnolas Perfect Vision GmbH · 510(k) 2020
- LenSx Laser System (8065000944)Alcon Laboratories, Inc. · 510(k) 2025
- Catalys Precision Laser SystemAmo Manufacturing USA, LLC · 510(k) 2023
- Catalys Precision Laser SystemOptimedica Corporation · 510(k) 2016
- TbdLensar, Inc. · 510(k) 2013
- Lensx 550 Laser System. Model 550Lensx Lasers, Inc. · 510(k) 2010
- ELITA Femtosecond Laser System, ELITA Patient InterfaceJohnson & Johnson Surgical Vision, Inc. · 510(k) 2023
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.