Device Trial Radar · NCT07949643
Operator-Assisted Oscilla Audiometry as a Basis for Hearing Aid Fitting (The Oscilla Validation Study)
Last checked: Nov 19, 2026Latest source record: Nov 19, 2026 (this record)Source: ClinicalTrials.govMethods & limitations
Status
Not yet recruiting
Enrollment
268
Planned (sponsor's estimate)
Start
Dec 1, 2026
Primary completion
Oct 31, 2027
Study completion Oct 31, 2027
Sponsor
- Lead sponsor
- Aalborg University HospitalAcademic, hospital or other
- Sponsor type
- Academic, hospital or other
- First posted
- Nov 19, 2026
Design
- Study type
- Interventional
- Phase
- Not applicable (typical for device studies)
- Allocation
- Randomized
- Model
- Parallel
- Masking
- Single (participant)
- Primary purpose
- Diagnostic
- FDA-regulated device
- No
- Who can join
- All sexes, from 18 years
- Healthy volunteers
- No
Summary
As written by the sponsor on the registry.
The purpose of this study is to find out whether hearing tests performed with the Oscilla A60 audiometer, where the patient largely tests themselves with minimal staff assistance, are accurate enough to be used for diagnosing hearing loss and as the sole basis for fitting hearing aids. Adults referred for their first hearing aids at Aalborg University Hospital will have their hearing tested both with the Oscilla audiometer and by a trained audiology assistant at the same visit. Participants are then randomly assigned to receive hearing aids fitted based on either the Oscilla hearing test or the audiology assistant's hearing test. The main questions are: (1) Do the Oscilla results agree with the audiology assistant's results for both air-conduction and bone-conduction hearing thresholds? (2) Do participants fitted based on the Oscilla test report a hearing benefit that is not worse than that of participants fitted based on the standard test, two months after fitting?
Conditions
- Hearing Loss, Sensorineural
- Presbyacusis
Interventions
Paired audiometry (Oscilla A60 and manual audiometry)Diagnostic test
At Visit 1, all participants undergo both operator-assisted Oscilla A60 audiometry (AudioConsole v4.6.0; AC and BC thresholds; operator places the BC headband only) in a quiet room (\<50 dB SPL) and manual pure-tone audiometry (AC and BC, ISO 8253-1) by a qualified audiology assistant in a sound-attenuating booth, in randomised order.
Oscilla-based hearing aid fittingDevice
Bilateral BTE/RIC hearing aid fitting to NAL-NL2 targets using the Oscilla A60 audiogram as the sole audiometric basis, verified with auto-REM (within ±5 dB of target). Other names: Oscilla A60 Diagnostic Audiometer.
Manual audiometry-based hearing aid fittingDevice
Bilateral BTE/RIC hearing aid fitting to NAL-NL2 targets using the audiology assistant audiogram, verified with auto-REM (within ±5 dB of target). Standard clinical practice.
Primary outcomes
What the sponsor says the trial is designed to measure.
- Agreement between Oscilla A60 and manual audiometry for air-conduction thresholdsTime frame: Visit 1 (Week 0)
- Agreement between Oscilla A60 and manual audiometry for bone-conduction thresholdsTime frame: Visit 1 (Week 0)
- Change in SSQ12 total scoreTime frame: Baseline (Week 0) and 2 months after hearing aid fitting
Locations
2 sites in 1 country. Site status is shown where the registry gives one.
Department of Audiology, Aalborg University Hospital
Aalborg, Denmark
Hearing Clinic, Aalborg University Hospital
Hobro, Denmark
Contact details for sites and study staff are not shown here. To ask about taking part, use the contact information on the registry record.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.